An ultra-high field strength MR image-guided robotic needle delivery system for in-bore small animal interventions

被引:2
|
作者
Gravett, Matthew [1 ,2 ]
Cepek, Jeremy [1 ,3 ]
Fenster, Aaron [1 ,2 ]
机构
[1] Robarts Res Inst, London, ON N6A 5B7, Canada
[2] Western Univ, Biomed Engn, London, ON N6A 5B9, Canada
[3] Western Univ, Schulich Sch Med & Dent, London, ON N6A 5C1, Canada
关键词
image-guided intervention; medical robotics; MR-guided preclinical intervention; needle delivery system; STEREOTAXIC SURGERY; MOUSE BRAINS; INJECTIONS; GUIDANCE; SCANNER; MODELS; MICE;
D O I
10.1002/mp.12534
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe purpose of this study was to develop and validate an image-guided robotic needle delivery system for accurate and repeatable needle targeting procedures in mouse brains inside the 12 cm inner diameter gradient coil insert of a 9.4 T MR scanner. Many preclinical research techniques require the use of accurate needle deliveries to soft tissues, including brain tissue. Soft tissues are optimally visualized in MR images, which offer high-soft tissue contrast, as well as a range of unique imaging techniques, including functional, spectroscopy and thermal imaging, however, there are currently no solutions for delivering needles to small animal brains inside the bore of an ultra-high field MR scanner. This paper describes the mechatronic design, evaluation of MR compatibility, registration technique, mechanical calibration, the quantitative validation of the in-bore image-guided needle targeting accuracy and repeatability, and demonstrated the system's ability to deliver needles in situ. MethodsOur six degree-of-freedom, MR compatible, mechatronic system was designed to fit inside the bore of a 9.4 T MR scanner and is actuated using a combination of piezoelectric and hydraulic mechanisms. The MR compatibility and targeting accuracy of the needle delivery system are evaluated to ensure that the system is precisely calibrated to perform the needle targeting procedures. A semi-automated image registration is performed to link the robot coordinates to the MR coordinate system. Soft tissue targets can be accurately localized in MR images, followed by automatic alignment of the needle trajectory to the target. Intra-procedure visualization of the needle target location and the needle were confirmed through MR images after needle insertion. ResultsThe effects of geometric distortions and signal noise were found to be below threshold that would have an impact on the accuracy of the system. The system was found to have negligible effect on the MR image signal noise and geometric distortion. The system was mechanically calibrated and the mean image-guided needle targeting and needle trajectory accuracies were quantified in an image-guided tissue mimicking phantom experiment to be 178 54 m and 0.27 +/- 0.65 degrees, respectively. ConclusionsAn MR image-guided system for in-bore needle deliveries to soft tissue targets in small animal models has been developed. The results of the needle targeting accuracy experiments in phantoms indicate that this system has the potential to deliver needles to the smallest soft tissue structures relevant in preclinical studies, at a wide variety of needle trajectories. Future work in the form of a fully-automated needle driver with precise depth control would benefit this system in terms of its applicability to a wider range of animal models and organ targets.
引用
收藏
页码:5544 / 5555
页数:12
相关论文
共 10 条
  • [1] 3D image-guided robotic needle positioning system for small animal interventions
    Bax, Jeffrey S.
    Waring, Christopher S. R.
    Sherebrin, Shi
    Stapleton, Shawn
    Hudson, Thomas J.
    Jaffray, David A.
    Lacefield, James C.
    Fenster, Aaron
    [J]. MEDICAL PHYSICS, 2013, 40 (01)
  • [2] Design and testing of a robotic system for MR image-guided interventions
    Christoforou, Eftychios G.
    Tsekos, Nikolaos V.
    Oezcan, Alpay
    [J]. JOURNAL OF INTELLIGENT & ROBOTIC SYSTEMS, 2006, 47 (02) : 175 - 196
  • [3] Design and Testing of a Robotic System for mr Image-guided Interventions
    Eftychios G. Christoforou
    Nikolaos V. Tsekos
    Alpay Özcan
    [J]. Journal of Intelligent and Robotic Systems, 2006, 47 : 175 - 196
  • [4] Characterization of the Precision and Accuracy of a New, Preclinical, MRI-Guided Focused Ultrasound System for Image-Guided Interventions in Small-Bore, High-Field Magnets
    Ellens, N.
    Farahani, K.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3263 - 3264
  • [5] MR-GUIDED NEEDLE-BIOPSY WITH A HIGH-FIELD-STRENGTH MR SYSTEM
    LUFKIN, RB
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 1991, 12 (06) : 1268 - 1268
  • [6] Towards a Modular, Customizable Robotic System for Needle-Based Image-Guided Interventions: Preliminary Designs, Implementation, and Testing
    Navkar, Nikhil V.
    Barbosa, Leonardo
    Balakrishnan, Shidin
    Abinahed, Julien
    El Ansari, Walid
    Al-Rumaihi, Khalid
    Darweesh, Adham
    Al-Ansari, Abdulla
    Tsekos, Nikolaos V.
    Karkoub, Mansour
    Gharib, Mohamed
    [J]. 2017 IEEE 17TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING (BIBE), 2017, : 444 - 449
  • [7] Image-guided cardiac cell delivery using high-resolution small-animal ultrasound
    Rodriguez-Porcel, M
    Gheysens, O
    Chen, IY
    Wu, JC
    Gambhir, SS
    [J]. MOLECULAR THERAPY, 2005, 12 (06) : 1142 - 1147
  • [8] Registration of a needle-positioning robot to high-resolution 3-D ultrasound and computed tomography for image-guided interventions in small animals
    Waspe, Adam C.
    Lacefield, James C.
    Holdsworth, David W.
    Fenster, Aaron
    [J]. MEDICAL IMAGING 2008: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, PTS 1 AND 2, 2008, 6918
  • [9] An Image-Guided Micro-Radiotherapy System Based On Carbon Nanotube Field Emission Technology for Small Animal Irradiation for Cancer Research
    Wang, S.
    Schreiber, E.
    Calderon-Colon, X.
    Peng, R.
    Sultana, S.
    Zhang, J.
    Zhou, O.
    Chang, S.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [10] Gd2O3/b-TiO2 composite nanoprobes with ultra-high photoconversion efficiency for MR image-guided NIR-II photothermal therapy
    Chen, Jia
    Chen, Tianxiang
    Fang, Qianlan
    Pan, Chunshu
    Akakuru, Ozioma Udochukwu
    Ren, Wenzhi
    Lin, Jie
    Sheng, Aizhu
    Ma, Xuehua
    Wu, Aiguo
    [J]. EXPLORATION, 2022, 2 (06):