Dose reduction and adequate image quality in digital radiography: a contradiction?

被引:0
|
作者
Poetter-Lang, S. [1 ]
Duenkelmeyer, M. [1 ]
Uffmann, M. [2 ]
机构
[1] Med Univ Wien, Univ Klin Radiodiagnost, A-1090 Vienna, Austria
[2] Krankenhaus Neunkirchen, Abt Radiodiagnost, Neunkirchen, Germany
来源
RADIOLOGE | 2012年 / 52卷 / 10期
关键词
Digital detector systems; Dose efficiency; Improved resolution; Quality control; ALARA (as low as reasonably achievable); FLAT-PANEL-DETECTOR; COMPUTED RADIOGRAPHY; CHEST RADIOGRAPHY; AMORPHOUS-SILICON; SCREEN-FILM; PRINCIPLES; PARAMETERS; SYSTEM;
D O I
10.1007/s00117-012-2337-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dose reduction and adequate image quality in digital radiography - a contradiction? Digital radiography has already replaced traditional screen-film systems. Substantial improvements in both dose efficiency and spatial resolution demonstrate the rapid developments in digital radiography. Needle-detector systems have shown up to a 50% dose reduction compared to traditional screen-film systems. There is also a dose reduction capability of up to 50% comparing direct radiography (DR) systems to computed radiography (CR) systems for chest X-rays. However, despite the most recent achievements of CR technology, the dose efficiency of DR systems (caesium iodide flat-panel detector) is unparalleled. The progress in detector technology has contributed to dose reduction and improved image quality, while saving time and providing a higher examination rate. The use of dose indicators and longitudinal dose control are important to avoid substantial accidental dose increase. The dose applied to patients should fall markedly below the defined diagnostic reference levels within the European Union. Regular quality control, as well as continuous education and training of medical and technical personnel, contribute to ensure that the ALARA (as low as reasonably achievable) principle is consistently followed.
引用
收藏
页码:898 / 904
页数:7
相关论文
共 50 条
  • [1] Dosisreduktion und adäquate Bildqualität in der digitalen Radiographie: ein Widerspruch?Dose reduction and adequate image quality in digital radiography: a contradiction?
    S. Pötter-Lang
    M. Dünkelmeyer
    M. Uffmann
    [J]. Der Radiologe, 2012, 52 (10): : 898 - 904
  • [2] Dose reduction in digital chest radiography and perceived image quality
    Kroft, L. J. M.
    Veldkamp, W. J. H.
    Mertens, B. J. A.
    Van Delft, J-P A.
    Geleijns, J.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (960): : 984 - 988
  • [3] Digital chest radiography image quality assessment with dose reduction
    Grewal, R. K.
    Young, N.
    Collins, L.
    Karunaratne, N.
    Sabharwal, R.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 71 - 80
  • [4] Digital chest radiography image quality assessment with dose reduction
    R. K. Grewal
    N. Young
    L. Collins
    N. Karunaratne
    R. Sabharwal
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 71 - 80
  • [5] New image quality and dose reduction technique for pediatric digital radiography
    Feghali, Joelle A.
    Chambers, Greg
    Delepierre, Julie
    Chapeliere, Sophie
    Mannes, Ines
    Adamsbaum, Catherine
    [J]. DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2021, 102 (7-8) : 463 - 470
  • [6] Effect of dose reduction in digital dental panoramic radiography on image quality
    Dannewitz, B
    Hassfeld, S
    Eickholz, P
    Mühling, J
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2002, 31 (01) : 50 - 55
  • [7] DIGITAL RADIOGRAPHY: IMAGE QUALITY AND RADIATION DOSE
    Seibert, J. Anthony
    [J]. HEALTH PHYSICS, 2008, 95 (05): : 586 - 598
  • [8] A Technique for Simulating the Effect of Dose Reduction on Image Quality in Digital Chest Radiography
    Veldkamp, Wouter J. H.
    Kroft, Lucia J. M.
    van Delft, Jan Pieter A.
    Geleijns, Jacob
    [J]. JOURNAL OF DIGITAL IMAGING, 2009, 22 (02) : 114 - 125
  • [9] Optimization of Dose and Image Quality for Computed Radiography and Digital Radiography
    John E. Aldrich
    Emerenciana Duran
    Pat Dunlop
    John R. Mayo
    [J]. Journal of Digital Imaging, 2006, 19 : 126 - 131
  • [10] A Technique for Simulating the Effect of Dose Reduction on Image Quality in Digital Chest Radiography
    Wouter J.H. Veldkamp
    Lucia J.M. Kroft
    Jan Pieter A. van Delft
    Jacob Geleijns
    [J]. Journal of Digital Imaging, 2009, 22 : 114 - 125