Significant reduction of radiation exposure to operator and staff during cardiac interventions by analysis of radiation leakage and improved lead shielding

被引:99
|
作者
Kuon, E
Schmitt, M
Dahm, JB
机构
[1] Klin Frank Schweiz, Dept Cardiol, D-91320 Ebermannstadt, Germany
[2] Univ Erlangen Nurnberg, Ctr Diagnost Radiol, Erlangen, Germany
[3] Ernst Moritz Arndt Univ Greifswald, Dept Internal Med, Greifswald, Germany
来源
AMERICAN JOURNAL OF CARDIOLOGY | 2002年 / 89卷 / 01期
关键词
D O I
10.1016/S0002-9149(01)02161-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The objectives of this study were to disclose and to reduce occupational radiation leakage in invasive cardiology. prospectively, we analyzed various dose parameters for 330 coronary procedures. We used a Rando phantom to measure scatter entrance skin air kerma to the operator (S-ESAK-O) during fluoroscopy for all standard tube angulations, and to plot isodose lines for 0degrees/0degrees-posterior anterior angulation. The patient's measured dose area product due to diagnostic catheterization and elective percutaneous transluminal coronary angioplasty was 6.2 and 10.4 Gycm(2), which represents 11% and 13% of currently typical values, respectively. With use of 0.5- and 1.0-mm overcouch and undercouch shielding, it was possible to reduce the mean of 4,686 nSv/Gycm(2) to 677 and 277 nSv/Gycm(2), respectively. Closure of radiation leakage up to 897 muSv/hour at the operator's gonadal height (80 to 105 cm), not heretofore described, was achieved by an additional 1.0-mm, lead-equivalent undercouch-top and overcouch-flop adjacent to the table, down to a S-ESAKO/dose area product level of 47.5 nSv/Gycm(2). With use of a 0.5-mm lead apron, collar, glasses, foot-switch shield and 1.0-mm lead cover around the patient's thighs, the operator received a mean S-ESAK-O of 8.5, while his forehead, eyes, thyroid, chest, gonads, and hands were exposed to 68.2, 1.2, 1.2, 1.2, 0.8, and 58.2 nSv/Gycm(2), respectively. In conclusion, radiation-attenuating intervention techniques and improved lead protection can effectively contribute to a new state of the art in invasive cardiology, with reduction of operator radiation exposure to 0.8% of typical S-ESAK-O levels in advanced catheterization laboratories. (C) 2002 by Excerpta Medica, Inc.
引用
收藏
页码:44 / 49
页数:6
相关论文
共 50 条
  • [21] Reduction of Exposure of Patients and Staff to Radiation During Fluoroscopically Guided Interventional Procedures
    Vano E.
    Current Radiology Reports, 2013, 1 (1) : 11 - 22
  • [22] Determinants of operator and patient radiation exposure during cardiac catheterization: Insights from the RadiCure (RADIation reduction during cardiac catheterization using real-timE monitoring) trial
    Christopoulos, Georgios
    Papayannis, Aristotelis C.
    Alomar, Mohammed
    Christakopoulos, Georgios E.
    Kotsia, Anna
    Michael, Tesfaldet T.
    Rangan, Bavana V.
    Roesle, Michele
    Shorrock, Deborah
    Makke, Lorenza
    Maragkoudakis, Spyros
    Mohammad, Atif
    Sarode, Karan
    Chambers, Charles E.
    Banerjee, Subhash
    Brilakis, Emmanouil S.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2016, 88 (07) : 1046 - 1055
  • [23] Predictors of High Operator Radiation Exposure During Cardiac Catheterization: Insights From the Radicure (Radiation Reduction During Cardiac Catheterization Using Real-time Monitoring) Study
    Christopoulos, Georgios
    Papayannis, Aristotelis C.
    Alomar, Mohammed
    Kotsia, Anna P.
    Michael, Tesfaldet T.
    Rangan, Bavana V.
    Roesle, Michele
    Shorrock, Deborah
    Makke, Lorenza
    Maragkoudakis, Spyros
    Mohammad, Atif
    Sarode, Karan
    Chambers, Charles E.
    Banerjee, Subhash
    Brilakis, Emmanouil S.
    CIRCULATION, 2014, 130
  • [24] THE EFFECT OF VARIOUS RADIATION SHIELDS ON OPERATOR EXPOSURE DURING CONGENITAL CARDIAC CATHETERIZATION
    Batlivala, Sarosh P.
    Magill, Dennise
    Felice, Marc A.
    Jones, Virginia
    Dori, Yoav
    Gillespie, Matthew
    Rome, Jonathan
    Glatz, Andrew
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 63 (12) : A522 - A522
  • [25] A new Design of the Lead Acrylic Glass Pane reduces the Radiation Exposure of the Staff during Heart Catheter Examinations and coronary Interventions
    Eder, H.
    Seidenbusch, M. C.
    Treitl, M.
    Gilligan, P.
    WIENER KLINISCHE WOCHENSCHRIFT, 2015, 127 (SUPPL 1) : S51 - S52
  • [26] Reduction of Operator Radiation Dose by a Pelvic Lead Shield During Cardiac Catheterization by Radial Access Comparison With Femoral Access
    Lange, Helmut W.
    von Boetticher, Heiner
    JACC-CARDIOVASCULAR INTERVENTIONS, 2012, 5 (04) : 445 - 449
  • [27] Radiation Exposure to Interventional Echocardiographers and Interventional Cardiologists during Structural Heart Interventions - Impact of Shielding
    Corrigan, Frank
    Iturbe, Jose Miguel
    Forcillo, Jessica
    Ramadan, Ronnie
    Condado, C. Jose
    Kamioka, Norihiko
    Thourani, Vinod
    Babaliaros, Vasilis
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 70 (18) : B130 - B130
  • [28] MONITORING AND REDUCTION OF PATIENT AND STAFF RADIATION EXPOSURE DURING INTERVENTIONAL AND THERAPEUTIC NEURORADIOLOGY PROCEDURES
    KUAN, HM
    MANZIONE, JV
    FERRETTI, JA
    RADIOLOGY, 1995, 197 : 359 - 359
  • [29] Estimating radiation exposure during paediatric cardiac catheterisation: a potential for radiation reduction with air gap technique
    Chamberlain, Reid C.
    Shindhelm, Alexis C.
    Wang, Chu
    Fleming, Gregory A.
    Hill, Kevin D.
    CARDIOLOGY IN THE YOUNG, 2019, 29 (12) : 1474 - 1480
  • [30] Radiation Exposure During Operative Fixation of Pediatric Supracondylar Humerus Fractures: Is Lead Shielding Necessary?
    Martus, Jeffrey E.
    Hilmes, Melissa A.
    Grice, Jared V.
    Stutz, Christopher M.
    Schoenecker, Jonathan G.
    Lovejoy, Steven A.
    Mencio, Gregory A.
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2018, 38 (05) : 249 - 253