Radiation dose reduction for chest CT with non-linear adaptive filters

被引:9
|
作者
Singh, Sarabjeet [1 ]
Digumarthy, Subba R. [1 ]
Back, Anni [2 ]
Shepard, Jo-anne O. [1 ]
Kalra, Mannudeep K. [1 ]
机构
[1] Massachusetts Gen Hosp Imaging, Boston, MA USA
[2] ContextVis AB, Linkoping, Sweden
关键词
CT radiation dose reduction; non-linear adaptive filters; postprocessing software; chest CT image quality; low dose chest CT; image noise reduction; IMAGE QUALITY; COMPUTED-TOMOGRAPHY; ABDOMEN; ADULTS;
D O I
10.1258/ar.2012.120045
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: CT radiation dose reduction results in increased noise or graininess of images which affects the diagnostic information. One of the approaches to lower radiation exposure to patients is to reduce image noise with the use of image processing software in low radiation dose images. Purpose: To assess image quality and accuracy of non-linear adaptive filters (NLAF) at low dose chest CT. Material and Methods: In an IRB approved prospective study, 24 patients (mean age, 63 +/- 7.3 years; M: F ratio, 11: 13) gave informed consent for acquisition of four additional chest CT image series at 150, 110, 75, and 40 mAs (baseline image series) on a 64-slice MDCT over an identical 10-cm length. NLAF was used to process three low dose (110, 75, and 40 mAs) image series (postprocessed image series). Two radiologists reviewed baseline and postprocessed images in a blinded manner for image quality. Objective noise, CT attenuation values, patient weight, transverse diameters, CTDIvol, and DLP were recorded. Statistical analysis was performed using parametric and non-parametric tests for comparing postprocessed and baseline images. Results: No lesions were missed on baseline or postprocessed CT images (n = 80 lesions, 73 lesions <1 cm). At 40 mAs, subjective noise in mediastinal window settings were graded as unacceptable in baseline images and acceptable in postprocessed images. Visibility of smaller structures improved from suboptimal visibility in baseline images at 40 mAs to excellent in postprocessed images at 40 mAs. No major artifacts were seen due to NLAF postprocessing, except for minor beam hardening artifacts not affecting diagnostic decision-making (14/22) in both baseline and postprocessed image series. Diagnostic confidence for chest CT was improved to fully confident in postprocessed images at 40 mAs. Compared to baseline images, postprocessing reduced objective noise by 26% (14.2 +/- 4.7/19.2 +/- 6.4), 31.5% (15.2 +/- 4.7/22.2 +/- 5.7), and 41.5% (16.9 +/- 6/28.9 +/- 10.2) at 110 mAs, 75 mAs, and 40 mAs tube current-time product levels. Conclusion: Applications of NLAF can help reduce tube current down to 40 mAs for chest CT while maintaining lesion conspicuity and image quality.
引用
收藏
页码:169 / 174
页数:6
相关论文
共 50 条
  • [1] Radiation dose reduction with application of non-linear adaptive filters for abdominal CT
    Singh, Sarabjeet
    Kalra, Mannudeep K.
    Sung, Mi Kim
    Back, Anni
    Blake, Michael A.
    [J]. WORLD JOURNAL OF RADIOLOGY, 2012, 4 (01): : 21 - 28
  • [2] Radiation dose reduction in chest CT: A review
    Kubo, Takeshi
    Lin, Pei-Jan Paul
    Stiller, Wolfram
    Takahashi, Masaya
    Kauczor, Hans-Ulrich
    Ohno, Yoshiharu
    Hatabu, Hiroto
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (02) : 335 - 343
  • [3] Non-linear effects in LMS adaptive filters
    Reuter, M
    Quirk, K
    Zeidler, J
    Milstein, L
    [J]. IEEE 2000 ADAPTIVE SYSTEMS FOR SIGNAL PROCESSING, COMMUNICATIONS, AND CONTROL SYMPOSIUM - PROCEEDINGS, 2000, : 141 - 146
  • [4] Data structure and non-linear effects in adaptive filters
    Beex, AA
    Zeidler, JR
    [J]. DSP 2002: 14TH INTERNATIONAL CONFERENCE ON DIGITAL SIGNAL PROCESSING PROCEEDINGS, VOLS 1 AND 2, 2002, : 659 - 662
  • [5] Application of adaptive non-linear 2D and 3D postprocessing filters for reduced dose abdominal CT
    Borgen, Lars
    Kalra, Mannudeep K.
    Laerum, Frode
    Hachette, Isabelle W.
    Fredriksson, Carina H.
    Sandborg, Michael
    Smedby, Orjan
    [J]. ACTA RADIOLOGICA, 2012, 53 (03) : 335 - 342
  • [6] Adaptive Statistical Iterative Reconstruction Technique for Radiation Dose Reduction in Chest CT: A Pilot Study
    Singh, Sarabjeet
    Kalra, Mannudeep K.
    Gilman, Matthew D.
    Hsieh, Jiang
    Pien, Homer H.
    Digumarthy, Subba R.
    Shepard, Jo-Anne O.
    [J]. RADIOLOGY, 2011, 259 (02) : 565 - 573
  • [7] Evaluation of a Radiation Dose Reduction Strategy for Pediatric Chest CT
    Kim, Jee-Eun
    Newman, Beverley
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05) : 1188 - 1193
  • [8] Prospective Evaluation of Effect of 2D Adaptive Filters on Low Radiation Dose Chest and Abdominal CT
    Singh, S.
    Kalra, M.
    Sharma, A.
    Forsberg, A.
    Shepard, J.
    Digumarthy, S.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05)
  • [9] Can noise reduction filters improve low-radiation-dose chest CT images? Pilot study
    Kalra, MK
    Wittram, C
    Maher, MM
    Sharma, A
    Avinash, GB
    Karau, K
    Toth, TL
    Halpern, E
    Saini, S
    Shepard, JA
    [J]. RADIOLOGY, 2003, 228 (01) : 257 - 264
  • [10] Estimated Radiation Dose Reduction Using Non-Linear Diffusion Method in Computed Radiography
    Guadalupe Sanchez, Ma.
    Juste, Belen
    Vidal, Vicente
    Verdu, Gumersindo
    Mayo, Patricia
    Rodenas, Francisco
    [J]. 2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 1502 - 1505