Measurement of mini-implant stability using resonance frequency analysis

被引:39
|
作者
Nienkemper, Manuel [1 ]
Wilmes, Benedict [1 ]
Panayotidis, Agamemnon [1 ]
Pauls, Alexander [1 ]
Golubovic, Vladimir [2 ]
Schwarz, Frank [2 ]
Drescher, Dieter [1 ]
机构
[1] Univ Dusseldorf, Dept Orthodont, D-40225 Dusseldorf, Germany
[2] Univ Dusseldorf, Dept Oral Surg & Admittance, D-40225 Dusseldorf, Germany
关键词
Orthodontics; Skeletal anchorage; Implant stability; Resonance frequency analysis; DAMPING CAPACITY ASSESSMENT; ORTHODONTIC ANCHORAGE; IN-VITRO; BONE THICKNESS; MICRO-IMPLANTS; INTERFACE; FAILURE; TISSUE;
D O I
10.2319/043012-354.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To investigate whether resonance frequency analysis (RFA) is suitable to measure orthodontic mini-implant stability. Implant size significantly affects the level of resonance frequency. Regarding the operating mode of RFA, it has to be proven whether the resonance frequency of mini-implants in bone fits the range of frequency emitted by the Osstell ISO device. Material and Methods: For this purpose the SmartPegs in the Osstell ISO device were modified to fit with the inner screw thread of orthodontic mini-implants, and 110 mini-implants were inserted into porcine pelvic bone. RFA was performed parallel and perpendicular to the run of superficial bone fibers. A suitability test, Periotest, was also performed in the same directions. Compacta thickness was measured using cone-beam computed tomography. Correlation tests and linear regression analysis were carried out between the three methods. Results: The RFA showed a mean Implant Stability Quotient value of 36.36 +/- 2.67, and the Periotest mean value was -2.10 +/- 1.17. The differences between the two directions of measurement were statistically significant (P > .001) for RFA and the Periotest. There was a high correlation between RFA and the Periotest (r = -0.90) and between RFA and compacta thickness (r = 0.71). The comparison between the Periotest and compacta thickness showed a correlation coefficient of r = -0.64. Conclusion: The present results suggest that RFA is feasible as a measurement method for orthodontic mini-implant stability. As a consequence, it could be used for clinical evaluation of current stability and allow stability-related loading of mini-implants to reduce the failure rate. (Angle Orthod. 2013;83:230-238.)
引用
收藏
页码:230 / 238
页数:9
相关论文
共 50 条
  • [21] Resonance frequency analysis of miniscrew implant stability
    Tseng, Yu-Chuan
    Pan, Chin-Yun
    Liu, Pao-Hsin
    Yang, Yi-Hsin
    Chang, Hong-Po
    Chen, Chun-Ming
    JOURNAL OF ORAL SCIENCE, 2018, 60 (01) : 64 - 69
  • [22] Mini-implant anchorage for the orthodontic practitioner
    Baumgaertel, Sebastian
    Razavi, Mohammad R.
    Hans, Mark G.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 133 (04) : 621 - 627
  • [23] Mini-implant Usage in Orthodontic Practice
    Acar, Yasemin Bahar
    Hergel, Civan Aved
    Ates, Mustafa
    Kucukkeles, Nazan
    TURKISH JOURNAL OF ORTHODONTICS, 2015, 28 (01) : 1 - 6
  • [24] In vitro comparative analysis of two resonance frequency measurement devices: Osstell implant stability coefficient and Penguin resonance frequency analysis
    Herrero-Climent, Mariano
    Falcao, Artur
    Lopez-Jarana, Paula
    Diaz-Castro, Carmen M.
    Rios-Carrasco, Blanca
    Vicente Rios-Santos, Jose
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (06) : 1124 - 1131
  • [25] Resonance frequency measurement for clinical testing and assessment of implant stability
    Di Raimondo, R
    Galasso, L
    5TH MEDITERRANEAN CONGRESS OF ORAL AND MAXILLOFACIAL SURGERY, 1999, : 103 - 107
  • [26] Mechanical stability assessment of novel orthodontic mini-implant designs: Part 2
    Hong, Christine
    Peter Truong
    Ha Na Song
    Wu, Benjamin M.
    Moon, Won
    ANGLE ORTHODONTIST, 2011, 81 (06) : 1001 - 1009
  • [27] The effect of varying healing times on orthodontic mini-implant stability: a microscopic computerized tomographic and biomechanical analysis
    Zhang, Qiang
    Zhao, Lixing
    Wu, Yeke
    Wang, Hui
    Zhao, Zhihe
    Xu, Zhenrui
    Wei, Xing
    Tang, Tian
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2011, 112 (04): : 423 - 429
  • [28] Assessment of mini-implant displacement using cone beam computed tomography
    Alves, Matheus, Jr.
    Baratieri, Carolina
    Nojima, Lincoln Issamu
    CLINICAL ORAL IMPLANTS RESEARCH, 2011, 22 (10) : 1151 - 1156
  • [29] Assessment of Implant Stability of Patients With and Without Radiotherapy Using Resonance Frequency Analysis
    Karayazgan-Saracoglu, Banu
    Atay, Arzu
    Zulfikar, Haluk
    Erpardo, Yuksel
    JOURNAL OF ORAL IMPLANTOLOGY, 2015, 41 (01) : 30 - 35
  • [30] Resonance Frequency Analysis in Assessing Implant Stability: A Retrospective Analysis
    Chen, Maggie Hsiao-Mei
    Lyons, Karl
    Tawse-Smith, Andrew
    Ma, Sunyoung
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2019, 32 (04) : 317 - 326