Improved quantification of mass fraction of colorants in textile by high-performance liquid chromatography coupled with tandem mass spectrometric detector

被引:3
|
作者
Tolgyesi, Adam [1 ]
Sharma, Virender K. [2 ]
Kulati, Anita [1 ]
机构
[1] EMI TUV SUD Ltd, Dugonics Utca 11, H-1043 Budapest, Hungary
[2] Texas A&M Univ, Program Environm & Sustainabil, Sch Publ Hlth, Dept Environm & Occupat Hlth, 212 Adriance Lab Rd,1266 TAMU, College Stn, TX 77843 USA
关键词
Colorants; Textiles; LC-MS; MS; HPLC-DAD; Standard method; Exact-matching double isotope dilution mass spectrometry; MILD EXTRACTION; DYES; IDENTIFICATION;
D O I
10.1007/s00769-020-01438-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The DIN 54231:2005 and ISO 16373-2:2014 standard methods recommend the application of HPLC separations and detection by either diode array (HPLC-DAD) or mass spectrometric detectors (LC-MS or LC-MS/MS) for the evaluation of mass concentration of colorants (y >= 1.0 mu g/mL) in textile sample extracts. The present investigation reports for the first time that the LC-MS/MS method has low accuracy for most of textile colorants if the recommended calibration range of the standard methods (1.0 mu g/mL to 20 mu g/mL or 5.0 mu g/mL to 50 mu g/mL) is used. The present paper reports that the quantification of mass concentration of colorants in textile extract by the LC-MS/MS method does not match with the quantification by the HPLC-DAD method, which has fit-for-purpose linearity and accuracy in the ranges mentioned above. A systematic study of the MS/MS method was performed to prepare real and spiked samples for quantifying mass concentration of colorants (y <= 1.0 mu g/mL) in textile extracts for compounds having LOQ <= 1.0 mg/kg. The results obtained could match with the quantification by the HPLC-DAD method. A lower calibration range was needed to quantify the dyes in the diluted sample extracts in which the background matrix constituents had lower influence on the quantification. Significantly, the LC-MS/MS method utilizing sample dilution and lower calibration range was validated and then used to proficiency test (PT) samples with success. Furthermore, the evaluation of mass fraction of disperse yellow 3 in PT sample by exact-matching double isotope dilution mass spectrometry was also performed. The strategy presented herein to quantify a wide range of colorants using the LC-MS/MS method could improve the reproducibility of analyzing colorants using protocols of standard recommendations.
引用
收藏
页码:259 / 272
页数:14
相关论文
共 50 条
  • [1] Improved quantification of mass fraction of colorants in textile by high-performance liquid chromatography coupled with tandem mass spectrometric detector
    Ádám Tölgyesi
    Virender K. Sharma
    Anita Kulati
    Accreditation and Quality Assurance, 2020, 25 : 259 - 272
  • [2] Felodipine quantification in human plasma by high-performance liquid chromatography coupled to tandem mass spectrometry
    Migliorança, LH
    Barrientos-Astigarraga, RE
    Schug, BS
    Blume, HH
    Pereira, AS
    De Nucci, G
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 814 (02): : 217 - 223
  • [3] Quantification of Tamsulosin in Human Plasma by High-Performance Liquid Chromatography Coupled with Electrospray Tandem Mass Spectrometry
    Shakya, A. K.
    Hakooz, N. M.
    Arafat, T. A.
    Abu-Awwad, A. N. A.
    El-Ghani, J. A.
    Melhim, M.
    ACTA CHROMATOGRAPHICA, 2010, 22 (01) : 99 - 116
  • [4] Development and validation of a high-performance liquid chromatography-tandem mass spectrometric method for quantification of daunorubicin in rat plasma
    Yang, Yuhui
    TALANTA, 2007, 71 (02) : 596 - 604
  • [5] A bioequivalence study of citalopram based on quantification by high-performance liquid chromatography coupled to electrospray tandem mass spectrometry
    Mendes, GD
    Borges, NCD
    Pereira, AD
    Mendes, FD
    Barrientos-Astigarraga, RE
    De Nucci, G
    INTERNATIONAL JOURNAL OF CLINICAL PHARMACOLOGY AND THERAPEUTICS, 2005, 43 (08) : 389 - 398
  • [6] Quantification of 5-azacytidine in plasma by electrospray tandem mass spectrometry coupled with high-performance liquid chromatography
    Zhao, M
    Rudek, MA
    He, P
    Hartke, C
    Gore, S
    Carducci, MA
    Baker, SD
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 813 (1-2): : 81 - 88
  • [7] Chemical constituents of Meconopsis horridula and their simultaneous quantification by high-performance liquid chromatography coupled with tandem mass spectrometry
    Liu, Jiajia
    Wu, Haimei
    Zheng, Feng
    Liu, Wenyuan
    Feng, Feng
    Xie, Ning
    JOURNAL OF SEPARATION SCIENCE, 2014, 37 (18) : 2513 - 2522
  • [8] A bioequivalence study of gliclazide based on quantification by high-performance liquid chromatography coupled with electrospray tandem mass spectrometry
    Mendes, G. D.
    Moreira, L. D.
    Pereira, A. dos S.
    Borges, A.
    Yui, F.
    Mendes, F. D.
    de Nucci, G.
    INTERNATIONAL JOURNAL OF CLINICAL PHARMACOLOGY AND THERAPEUTICS, 2007, 45 (03) : 175 - 185
  • [9] Determination of nifedipine in dog plasma by high-performance liquid chromatography with tandem mass spectrometric detection
    Pan, Xigui
    Zhou, Shunchang
    Fu, Qinqin
    Hu, Xianming
    Wu, Jianhong
    BIOMEDICAL CHROMATOGRAPHY, 2014, 28 (07) : 1036 - 1040
  • [10] Development and validation of a capillary high-performance liquid chromatography/electrospray tandem mass spectrometric method for the quantification of bisphenol A in air samples
    Sabatini, L
    Barbieri, A
    Violante, FS
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (23) : 3468 - 3472