A new scanning technique for X-ray computed tomography

被引:2
|
作者
Nukaga, Jun [1 ]
Kitazawa, So [1 ]
Kamimura, Hiroshi [1 ]
机构
[1] Hitachi Ltd, Power Syst Power & Ind Syst R&D Lab, Hitachi, Ibaraki 3191221, Japan
关键词
X-ray; CT; third generation scanning; partial volume; artifact;
D O I
10.1016/j.ndteint.2008.04.003
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
A new scanning technique for reducing artifacts due to the partial volume effect is presented which maintains the spatial resolution of the conventional method. These artifacts are stripes and they have a ring shape between the source and the rotation center. Experimental and simulation results are in good agreement. The scanning technique has two data acquisitions per one cross-section, "+" quarter-sensor offset and "-" quarter-sensor offset. The partial volume effect is reduced by mathematical averaging the corresponding data of two data sets and doing image reconstruction with the average data. (c) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:564 / 569
页数:6
相关论文
共 50 条
  • [41] A new method of performance verification for x-ray computed tomography measurements
    Leonard, F.
    Brown, S. B.
    Withers, P. J.
    Mummery, P. M.
    McCarthy, M. B.
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2014, 25 (06)
  • [42] Microscopic x-ray luminescence computed tomography
    Zhang, Wei
    Zhu, Dianwen
    Zhang, Kun
    Li, Changqing
    [J]. MULTIMODAL BIOMEDICAL IMAGING X, 2015, 9316
  • [43] Contemporary developments in computed X-ray tomography
    Seibert, JA
    [J]. RADIOLOGY, 1998, 209P : 207 - 208
  • [44] X-ray computed tomography in life sciences
    Shelley D. Rawson
    Jekaterina Maksimcuka
    Philip J. Withers
    Sarah H. Cartmell
    [J]. BMC Biology, 18
  • [45] X-ray computed tomography for medical imaging
    Hiriyannaiah, HP
    [J]. IEEE SIGNAL PROCESSING MAGAZINE, 1997, 14 (02) : 42 - 59
  • [46] X-ray luminescence computed tomography using a focused x-ray beam
    Zhang, Wei
    Lun, Michael C.
    Nguyen, Alex Anh-Tu
    Li, Changqing
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2017, 22 (11)
  • [47] X-ray computed tomography of ultralightweight metals
    Winter, JM
    Green, RE
    Waters, AM
    Green, WH
    [J]. RESEARCH IN NONDESTRUCTIVE EVALUATION, 1999, 11 (04) : 199 - 211
  • [48] STATISTICAL ASPECTS OF COMPUTED X-RAY TOMOGRAPHY
    CHESLER, DA
    PELC, NJ
    RIEDERER, SJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (01): : 130 - 131
  • [49] NEW X-RAY DIGITAL RADIOGRAPHY AND COMPUTED TOMOGRAPHY FOR CULTURAL HERITAGE
    Casali, F.
    Bettuzzi, M.
    Brancaccio, R.
    Morigi, M. P.
    [J]. SCIENCE FOR CULTURAL HERITAGE: TECHNOLOGICAL INNOVATION AND CASE STUDIES IN MARINE AND LAND ARCHAEOLOGY IN THE ADRIATIC REGION AND INLAND, 2010, : 85 - +
  • [50] New x-ray computed tomography system for laboratory mouse imaging
    Paulus, M.J.
    Sari-Sarraf, H.
    Gleason, S.S.
    Bobrek, M.
    Hicks, J.S.
    Johnson, D.K.
    Behel, J.K.
    Thompson, L.H.
    Allen, W.C.
    [J]. IEEE Transactions on Nuclear Science, 1999, 46 (3 II): : 558 - 564