Study of Head Scatter Factor in 4MV Photon Beam Used in Radiotherapy

被引:1
|
作者
Appasamy, Murugan [1 ,2 ]
Xavier, Sidonia Valas [1 ]
Kuppusamy, Thayalan [1 ]
Velayudham, Ramasubramanian [2 ]
机构
[1] Dr Kamakshi Mem Hosp, Med Phys Div, Chennai 600100, Tamil Nadu, India
[2] VIT Univ, Nucl & Med Phys Div, Sch Adv Sci, Vellore 632014, Tamil Nadu, India
关键词
Head scatter factor (S-c); Phantom scatter factor (S-p); 4; MV; Miniphantom; OUTPUT FACTOR; ELECTRON CONTAMINATION; LINEAR ACCELERATORS; MINI-PHANTOM; RATIO; AIR; CO-60; FIELD;
D O I
10.7785/tcrt.2012.500316
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The 4 MV photon beam offers equal build-up region behavior like Co-60 beam and it plays a major role in head and neck and pediatric radiotherapy. In this study an attempt is made to study the head scatter factor (S-c) for 4 MV photon beam using locally designed PMMA and Brass miniphantoms. The S-c is measured in combination of PMMA miniphantom with 0.6 cc chamber and Brass miniphantom with 0.6 cc and 0.13 cc chambers. The measured S-c is compared with the literature data and it agrees within +/- 1.98%. The study reveals that either 0.13 cc or 0.6 cc chamber with PMMA or Brass phantom materials can be used for S-c measurements in a 4 MV photon beam. The variation of SSD does not alter the head scatter factor: The collimator exchange effect is found to be within 1, and it is less than that of other linear accelerators. It is also found that the presence of internal wedge has significant contribution to head scatter factor. The Phantom scatter factor is also calculated and it agrees within +/- 1% with published data.
引用
收藏
页码:207 / 216
页数:10
相关论文
共 50 条
  • [21] CONTAMINANT ELECTRONS IN THE BUILDUP REGION OF A 4 MV PHOTON-BEAM
    RUSTGI, SN
    GROMADZKI, ZC
    LING, CC
    YORKE, ED
    PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (06): : 659 - 665
  • [22] Head scatter factors for small MV photon fields. Part II: The effects of source size and detector
    McKerracher, Carotyn
    Thwaites, David I.
    RADIOTHERAPY AND ONCOLOGY, 2007, 85 (02) : 286 - 291
  • [23] Head scatter factors for small MV photon fields. Part I: A comparison of phantom types and methodologies
    McKerracher, Carolyn
    Thwaites, David I.
    RADIOTHERAPY AND ONCOLOGY, 2007, 85 (02) : 277 - 285
  • [24] Tissue composition effect on dose distribution in radiotherapy with a 6 MV photon beam of a medical linac
    Ghorbani, Mahdi
    Noghreiyan, Atefeh Vejdani
    Tabatabaei, Zahra Sadat
    Pakravan, Delaram
    Davenport, David
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (01) : 237 - 244
  • [25] Estimation of photoneutron intensities around radiotherapy linear accelerator 23-MV photon beam
    Shweikani, R.
    Anjak, O.
    APPLIED RADIATION AND ISOTOPES, 2015, 99 : 168 - 171
  • [26] Commissioning and Potential Clinical Implementation and Advantages of a 2.5 MV Photon Beam for Selected Radiotherapy Cases
    Chen, H.
    Guo, F.
    Carlson, D.
    Deng, J.
    Nath, R.
    Chen, Z.
    MEDICAL PHYSICS, 2017, 44 (06) : 3206 - 3206
  • [27] EFFECT OF FIELD IRREGULARITIES ON THE DOSE DISTRIBUTION OF 4 MV PHOTON-BEAM
    NIROOMANDRAD, A
    PALIWAL, BR
    RADIOTHERAPY AND ONCOLOGY, 1987, 8 (01) : 43 - 48
  • [28] EFFECT OF FIELD IRREGULARITIES ON THE DOSE DISTRIBUTION OF 4 MV PHOTON-BEAM
    NIROOMANDRAD, A
    PALIWAL, B
    MEDICAL PHYSICS, 1987, 14 (03) : 482 - 482
  • [29] Evaluation of Tissue Composition Effect On Dose Distribution in Radiotherapy with 6 MV Photon Beam of a Medical Linac
    Ghorbani, M.
    Tabatabaei, Z.
    Noghreiyan, A. Vejdani
    Meigooni, A. Soleimani
    MEDICAL PHYSICS, 2015, 42 (06) : 3428 - 3428
  • [30] Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments
    Decker, Savannah M.
    Bruza, Petr
    Zhang, Rongxiao
    Williams, Benjamin B.
    Jarvis, Lesley A.
    Pogue, Brian W.
    Gladstone, David J.
    MEDICAL PHYSICS, 2024, 51 (08) : 5754 - 5763