Intrafraction Verification of Gated RapidArc by Using Beam-Level Kilovoltage X-Ray Images

被引:26
|
作者
Li, Ruijiang [1 ]
Mok, Edward [1 ]
Chang, Daniel T. [1 ]
Daly, Megan [1 ]
Loo, Billy W., Jr. [1 ]
Diehn, Maximilian [1 ]
Quynh-Thu Le [1 ]
Koong, Albert [1 ]
Xing, Lei [1 ]
机构
[1] Stanford Univ, Sch Med, Dept Radiat Oncol, Stanford, CA 94305 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
BODY RADIATION-THERAPY; COMPUTED-TOMOGRAPHY; RADIOTHERAPY; LOCALIZATION; FEASIBILITY; ACCURACY; TRACKING; CANCER;
D O I
10.1016/j.ijrobp.2012.03.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To verify the geometric accuracy of gated RapidArc treatment using kV images acquired during dose delivery. Methods and Materials: Twenty patients were treated using the gated RapidArc technique with a Varian TrueBeam STx linear accelerator. One to 7 metallic fiducial markers were implanted inside or near the tumor target before treatment simulation. For patient setup and treatment verification purposes, the internal target volume (ITV) was created, corresponding to each implanted marker. The gating signal was generated from the Real-time Position Management (RPM) system. At the beginning of each fraction, individualized respiratory gating amplitude thresholds were set based on fluoroscopic image guidance. During the treatment, we acquired kV images immediately before MV beam-on at every breathing cycle, using the on-board imaging system. After the treatment, all implanted markers were detected, and their 3-dimensional (3D) positions in the patient were estimated using software developed in-house. The distance from the marker to the corresponding ITV was calculated for each patient by averaging over all markers and all fractions. Results: The average 3D distance between the markers and their ITVs was 0.8 +/- 0.5 mm (range, 0-1.7 mm) and was 2.1 +/- 1.2 mm at the 95th percentile (range, 0-3.8 mm). On average, a left-right margin of 0.6 mm, an anterior-posterior margin of 0.8 mm, and a superior-inferior margin of 1.5 mm is required to account for 95% of the intrafraction uncertainty in RPM-based RapidArc gating. Conclusion: To our knowledge, this is the first clinical report of intrafraction verification of respiration-gated RapidArc treatment in stereotactic ablative radiation therapy. For some patients, the markers deviated significantly from the ITV by more than 2 mm at the beginning of the MV beam-on. This emphasizes the need for gating techniques with beam-on/-off controlled directly by the actual position of the tumor target instead of external surrogates such as RPM. (C) 2012 Elsevier Inc.
引用
收藏
页码:E709 / E715
页数:7
相关论文
共 50 条
  • [1] Intrafraction Verification using Beam Level kV Images for Gated RapidArc
    Li, R.
    Mok, E.
    Chang, D.
    Koong, A.
    Daly, M. E.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S152 - S153
  • [2] Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images
    Li, Ruijiang
    Mok, Edward
    Han, Bin
    Koong, Albert
    Xing, Lei
    MEDICAL PHYSICS, 2012, 39 (05) : 2686 - 2693
  • [3] Advances in kilovoltage x-ray beam dosimetry
    Hill, Robin
    Healy, Brendan
    Holloway, Lois
    Kuncic, Zdenka
    Thwaites, David
    Baldock, Clive
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (06): : R183 - R231
  • [4] In-Phantom Detector Response in a Kilovoltage X-Ray Beam
    Lawless, M.
    DeWerd, L.
    MEDICAL PHYSICS, 2014, 41 (06) : 426 - 426
  • [5] Evaluation of radiation dosimeters for kilovoltage x-ray beam dosimetry
    Hill, R.
    Me, Z.
    Haque, M.
    Baldock, C.
    MEDICAL PHYSICS, 2007, 34 (06) : 2336 - 2337
  • [6] Geometric Accuracy Verification for Gated RapidArc Using Gated CBCT Reconstructed From Continuously Triggered KV Images
    Cheung, Michael L. M.
    Lee, Louis K. Y.
    Kam, Michael K. M.
    Yu, Brian K. H.
    Chan, Anthony T. C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [7] A novel attachment system for cutouts in kilovoltage x-ray beam therapy
    Baumgartl, M.
    Kohler, G.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S813 - S813
  • [8] A Novel attachment system for cutouts in kilovoltage X-ray beam therapy
    Baumgartl, M.
    Kohler, G.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 872 - 872
  • [9] Broad beam attenuation measurements in lead in kilovoltage X-ray beams
    Akber, Syed F.
    Das, Indra J.
    Kehwar, Than S.
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2008, 18 (03): : 197 - 202
  • [10] Verification of nonlinearity in digital X-ray images using surrogate method
    Sugiura, A.
    Yokoyama, K.
    Takada, H.
    Yasuda, N.
    Horii, A.
    Kida, K.
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961