Capacitive Heart-Rate Sensing on Touch Screen Panel with Laterally Interspaced Electrodes

被引:4
|
作者
Kim, Junhyung [1 ]
Song, Wonho [2 ]
Jung, Sungchul [3 ]
Kim, Yuna [4 ]
Park, Wonsang [4 ]
You, Bonghyun [4 ]
Park, Kibog [1 ,2 ]
机构
[1] Ulsan Natl Inst Sci & Technol UNIST, Sch Elect & Comp Engn, Ulsan 44919, South Korea
[2] Ulsan Natl Inst Sci & Technol UNIST, Dept Phys, Ulsan 44919, South Korea
[3] SK Hynix, Icheon Si 17336, Gyeonggi Do, South Korea
[4] Samsung Display Giheung Campus, Yongin 17113, Gyeonggi Do, South Korea
基金
新加坡国家研究基金会;
关键词
heart-rate sensing; capacitive touch screen panel; laterally-interspaced electrodes; ulnar artery; effective dielectric constant; fast fourier transformation; biomedical monitoring; multi-functional sensor;
D O I
10.3390/s20143986
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
It is demonstrated that the heart-rate can be sensed capacitively on a touch screen panel (TSP) together with touch signals. The existing heart-rate sensing systems measure blood pulses by tracing the amount of light reflected from blood vessels during a number of cardiac cycles. This type of sensing system requires a considerable amount of power and space to be implemented in multi-functional mobile devices such as smart phones. It is found that the variation of the effective dielectric constant of finger stemming from the difference of systolic and diastolic blood flows can be measured with laterally interspaced top electrodes of TSP. The spacing between a pair of non-adjacent top electrodes turns out to be wide enough to distinguish heart-rate signals from noises. With the aid of fast Fourier transform, the heart-rate can be extracted reliably, which matches with the one obtained by actually counting heart beats on the wrist.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 50 条
  • [41] Self Capacitance Mismatch Calibration Technique for Fully-Differential Touch Screen Panel Self Capacitance Sensing System
    Seong, Siheon
    Lee, Sewon
    Bae, Sunghyun
    Lee, Minjae
    SENSORS, 2023, 23 (07)
  • [42] A 50.7-dB-DR Finger-Resistance Extracting Multi-Touch Sensor IC for Soft Classification of Fingers Contacted on 6.7-in Capacitive Touch Screen Panel
    Song, Tae-Gyun
    Kim, Dong-Kyu
    Cho, Jeong-Hyun
    Lee, Ji-Hun
    Kim, Hyun-Sik
    IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2021, 56 (11) : 3470 - 3485
  • [43] Heart-rate Data Transmission Control for Vital Sensing from Densely-located Exercisers
    Okita, Kazuki
    Yomo, Hiroyuki
    Kawabata, Takashi
    2017 11TH INTERNATIONAL SYMPOSIUM ON MEDICAL INFORMATION AND COMMUNICATION TECHNOLOGY (ISMICT), 2017, : 69 - 72
  • [44] RECORDING SKIN RESISTANCE AND BEAT-BY-BEAT HEART-RATE FROM SAME PAIR OF DRY ELECTRODES
    GEDDES, LA
    BOURLAND, JD
    SMALLING, RW
    STEINBERG, RB
    PSYCHOPHYSIOLOGY, 1974, 11 (03) : 394 - 397
  • [45] USE OF SAME PAIR OF DRY ELECTRODES TO RECORD SKIN RESISTANCE AND BEAT-BY-BEAT HEART-RATE
    GEDDES, LA
    BOURLAND, JD
    SMALLING, RW
    STEINBERG, RB
    MEDICAL & BIOLOGICAL ENGINEERING, 1975, 13 (01): : 89 - 96
  • [46] Ultra-high transparent sandwich structure with a silicon dioxide passivation layer prepared on a colorless polyimide substrate for a flexible capacitive touch screen panel
    Wu, Chia-Ching
    SOLAR ENERGY MATERIALS AND SOLAR CELLS, 2020, 207
  • [47] Healable Capacitive Touch Screen Sensors Based on Transparent Composite Electrodes Comprising Silver Nanowires and a Furan/Maleimide Diels-Alder Cycloaddition Polymer
    Li, Junpeng
    Liang, Jiajie
    Li, Lu
    Ren, Fengbo
    Hu, Wei
    Li, Juan
    Qi, Shuhua
    Pei, Qibing
    ACS NANO, 2014, 8 (12) : 12874 - 12882
  • [48] A Novel Noise Suppression Solution for Low Cost Microcontroller-based Touch Panel: Using Inverse Charge Transfer Capacitive Sensing Method
    Yu, Yu-Hsiang
    Sun, Tsung-Ying
    2014 IEEE 3RD GLOBAL CONFERENCE ON CONSUMER ELECTRONICS (GCCE), 2014, : 500 - 504
  • [49] A Supervised Machine Learning Algorithm for Heart-Rate Detection Using Doppler Motion-Sensing Radar
    Saluja, Justin
    Casanova, Joaquin
    Lin, Jenshan
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2020, 4 (01): : 45 - 51
  • [50] A low-power fast readout circuit using a dual-mode sensing algorithm for medium-size capacitive touch screen panels
    Kim, Hyeon-June
    Jo, Dong-Shin
    Yang, Jun-Hyeok
    Ryu, Seung-Tak
    Digest of Technical Papers - SID International Symposium, 2014, 45 (01): : 540 - 543