Distal Femoral Fixation: A Biomechanical Comparison of Trigen Retrograde Intramedullary (IM) Nail, Dynamic Condylar Screw (DCS), and Locking Compression Plate (LCP) Condylar Plate

被引:84
|
作者
Heiney, Jake P. [1 ]
Barnett, Michael D. [4 ]
Vrabec, Gregory A. [2 ]
Schoenfeld, Andrew J. [2 ]
Baji, Avinash [3 ]
Njus, Glen O. [3 ]
机构
[1] Promed Hlth Syst, Dept Orthopaed Surg, Toledo, OH USA
[2] Akron Gen Med Ctr, Dept Orthopaed, Akron, OH USA
[3] Univ Akron, Dept Biomed Engn, Akron, OH 44325 USA
[4] Wright State Univ, Sch Med, Dept Orthopaed Surg, Orthopaed Surg Resideucy Program,Miami Valley Hos, Dayton, OH USA
关键词
Locking condylar plate; Distal femoral fracture; Biomechanical; Retrograde femoral nail; Dynamic condylar screw; INVASIVE STABILIZATION SYSTEM; FEMUR FRACTURE FIXATION; SUPRACONDYLAR FRACTURES; INTERNAL-FIXATION; ELDERLY-PATIENTS; BUTTRESS PLATE; SIDE PLATE; LISS;
D O I
10.1097/TA.0b013e31815edeb8
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background. The purpose of this study was to establish if there are biomechanical differences between implants in stiffness of construct, microdisplacement, and fatigue failure in a supracondylar femoral fracture model. Methods: A retrograde intramedullary (IM) nail, dynamic condylar screw (DCS), and locked condylar plate (LCP) were tested using 33-cm long synthetic femurs. A standardized supracondylar medial segmental defect was created in the distal femur bone models. A gap away from the distal joint axis and parallel to the knee axis was created for axial testing of the specimens (Arbeitsgemeinschaft fur Osteosynthesefragen [AO] type 33-A) and a T-fracture (33-C) was created for the fatigue testing of the specimens. Peak displacements were measured, and analysis was done to determine construct stiffness and gap micromotion in axial loading. Cyclic loading was performed for fatigue testing. Results: It was observed that there were statistically significant differences in micromotion across the fracture gap and overall stiffness of various implant constructs. The stiffness of the IM nail, DCS, and LCP were 1,106, 750, and 625 N/mm, respectively. The average total micromotion across the fracture gap for the IM nail, DCS, and LCP were 1.96, 10.55, and 17.74 mm, respectively. In fatigue testing, the IM nail distal screws failed at 9,000 cycles, the DCS did not fail (80,000 cycles completed), and the LCP failed at 19,000 and 23,500 cycles.. Conclusion: When considering micromotion and construct stiffness, the IM nail had statistically significant higher stiffness and significantly lower micromotion across the fracture gap with axial compression. Hence, the IM nail tested had the greatest stability for type 33-A fractures. However, the nail demonstrated the least amount of resistance to fatigue failure with type 33-C fractures, whereas the DCS did not fail with testing in any pattern.
引用
收藏
页码:443 / 449
页数:7
相关论文
共 50 条
  • [1] Biomechanical comparison of dynamic condylar screw and locking compression plate fixation in unstable distal femoral fractures: An in vitro study
    Singh, Ashutosh Kumar
    Rastogi, Amit
    Singh, Vakil
    INDIAN JOURNAL OF ORTHOPAEDICS, 2013, 47 (06) : 615 - 620
  • [2] Biomechanical comparison of dynamic condylar screw and locking compression plate fixation in unstable distal femoral fractures: An in vitro study
    Ashutosh Kumar Singh
    Amit Rastogi
    Vakil Singh
    Indian Journal of Orthopaedics, 2013, 47 : 615 - 620
  • [3] Biomechanical analysis of distal femoral fracture fixation: dynamic condylar screw versus locked compression plate
    Narsaria, Nidhi
    Singh, Ashutosh K.
    Rastogi, Amit
    Singh, Vakil
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2014, 19 (05) : 770 - 775
  • [4] Biomechanical evaluation of intramedullary retrograde nail and dynamic condylar screw used in unstable distal femoral fractures
    Dizdar, Ayberk
    Celik, Talip
    Ozkan, Arif
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (04) : 1755 - 1764
  • [5] Biomechanical evaluation of intramedullary retrograde nail and dynamic condylar screw used in unstable distal femoral fractures
    Ayberk Dizdar
    Talip Celik
    Arif Ozkan
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1755 - 1764
  • [6] Distal femoral fixation: A biomechanical comparison of retrograde nail, retrograde intramedullary nail, and prototype locking retrograde nail
    Heiney, Jake P.
    Battula, Suneel
    O'Connor, Jill A.
    Ebraheim, Nabil
    Schoenfeld, Andrew J.
    Vrabec, Gregory
    CLINICAL BIOMECHANICS, 2012, 27 (07) : 692 - 696
  • [7] The mechanics of internal fixation of fractures of the distal femur: a comparison of the condylar screw (DCS) with the condylar plate (CP)
    Harder, Y
    Martinet, O
    Barraud, GE
    Cordey, J
    Regazzoni, P
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1999, 30 : S31 - S39
  • [8] Bridge plate osteosynthesis using dynamic condylar screw (DCS) or retrograde intramedullary supracondylar nail (RIMSN) in the treatment of distal femoral fractures: comparison of two methods in a prospective randomized study
    Dar, Gh Nabi
    Tak, Shafaat Rashid
    Kangoo, Khursheed Ahmed
    Halwai, Manzoor Ahmed
    ULUSAL TRAVMA VE ACIL CERRAHI DERGISI-TURKISH JOURNAL OF TRAUMA & EMERGENCY SURGERY, 2009, 15 (02): : 148 - 153
  • [9] Dynamic Condylar Screw (DCS) versus Condylar Blade Plate (CBP) in Complex Supracondylar Femoral Fractures - A Biomechanical Study
    Sirbu, P. D.
    Carata, E.
    Petreus, T.
    Munteanu, F.
    Popescu, C.
    Asaftei, R.
    Botez, P.
    INTERNATIONAL CONFERENCE ON ADVANCEMENTS OF MEDICINE AND HEALTH CARE THROUGH TECHNOLOGY, 2009, 26 : 409 - +
  • [10] Comparison of Retrograde Femoral Interlocking Nail Versus Dynamic Condylar Screw in Distal Femoral Shaft Fractures
    Shafi, Muhammad Kamran
    Noor, Fraz
    Zeb, Syed Alam
    Ishfaq, Muhammad
    Bin Tahir, Yousaf
    Khan, Tauqeer Nawaz
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2021, 15 (06): : 2145 - 2148