Performance of the grating-crystal monochromator of the ALOISA beamline at the Elettra Synchrotron

被引:181
|
作者
Floreano, L
Naletto, G
Cvetko, D
Gotter, R
Malvezzi, M
Marassi, L
Morgante, A
Santaniello, A
Verdini, A
Tommasini, F
Tondello, G
机构
[1] Ist Nazl Fis Mat, Lab TASC, I-34012 Trieste, Italy
[2] Univ Padua, Dept Elect & Informat, INFM, Unit Padova, I-35131 Padua, Italy
[3] Sincrotrone Trieste, I-34012 Trieste, Italy
[4] Univ Pavia, Dept Elect, I-27100 Pavia, Italy
[5] Univ Modena, Dept Phys, I-41100 Modena, Italy
[6] Univ Ljubljana, Dept Phys, Ljubljana 61000, Slovenia
[7] Univ Trieste, Dept Phys, I-34100 Trieste, Italy
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1999年 / 70卷 / 10期
关键词
D O I
10.1063/1.1150001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The new beamline ALOISA, now operational at the Elettra Synchrotron, is designed for surface studies by means of several experimental techniques: surface x-ray diffraction and reflectivity, photoemission spectroscopy, photoelectron diffraction, e(-)-Auger coincidence spectroscopy. A new monochromator has been specifically designed and realized for this multipurpose beamline: it makes use of a channel-cut Si crystal dispersive element for the 3-8 keV range and of a plane mirror-plane grating element for the 200-2000 eV range. Both dispersive elements share the same optical system. In the low energy range (200-900 eV) the spectral resolving power exceeds 5000 while maintaining a throughput higher than 10(10) photons/s/200 mA/0.02% BW. In the case of the N-2 1s -->pi* and Ne 1s --> 3p transitions, the extremely high signal-to-noise ratio of the absorption spectra allowed a very accurate determination of the corresponding natural linewidth (116 +/- 2 and 250 +/- 10 eV, respectively). Moreover, the vibrational structure of the CO-oxygen 1s -->pi(*) transition has been fully resolved. In the high energy range, the measured flux exceeds 10(10) photons/s/200 mA up to 6.5 keV with a resolving power of similar to 7500. (C) 1999 American Institute of Physics. [S0034-6748(99)01410-0].
引用
收藏
页码:3855 / 3864
页数:10
相关论文
共 50 条
  • [31] High efficiency diffraction grating for EUV lithography beamline monochromator
    Voronov, D. L.
    Warwick, T.
    Gullikson, E. M.
    Salmassi, F.
    Naulleau, P.
    Artemiev, N. A.
    Lum, P.
    Padmore, H. A.
    ADVANCES IN X-RAY/EUV OPTICS AND COMPONENTS XI, 2016, 9963
  • [32] Commissioning and performance of the variable line spacing plane grating monochromator beamline at the Canadian Light Source
    Hu, Y. F.
    Zuin, L.
    Wright, G.
    Igarashi, R.
    McKibben, M.
    Wilson, T.
    Chen, S. Y.
    Johnson, T.
    Maxwell, D.
    Yates, B. W.
    Sham, T. K.
    Reininger, R.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (08):
  • [33] Thin film stress and texture analysis at the MCX synchrotron radiation beamline at ELETTRA
    Ortolani, Matteo
    Ricardo, Cristy Leonor Azanza
    Lausi, Andrea
    Scardi, Paolo
    RESIDUAL STRESSES VIII, 2011, 681 : 115 - +
  • [34] Performance of SISSI, the infrared beamline of the ELETTRA storage ring
    Lupi, Stefano
    Nucara, Alessandro
    Perucchi, Andrea
    Calvani, Paolo
    Ortolani, Michele
    Quaroni, Luca
    Kiskinova, Maya
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS, 2007, 24 (04) : 959 - 964
  • [35] SIMPLE PLANE GRATING MONOCHROMATOR FOR SYNCHROTRON-RADIATION
    LU, LJ
    COCCO, D
    JARK, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 339 (03): : 604 - 609
  • [36] COMPACT MONOCHROMATOR FOR THE MULTIPURPOSE X3 SYNCHROTRON BEAMLINE AT NSLS
    DAROVSKY, A
    MESHKOVSKY, I
    COPPENS, P
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02): : 2085 - 2087
  • [37] A double crystal X-ray monochromator for the SpLine diffraction and absorption synchrotron bending magnet beamline at the ESRF
    Rubio-Zuazo, J.
    Collado-Negro, Victor
    Heyman, C.
    Ferrer, Pilar
    da Silva, Ivan
    Gallastegui, J. A.
    Gutierrez-Leon, A.
    Castro, G. R.
    11TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI 2012), 2013, 425
  • [38] A DOUBLE CRYSTAL MONOCHROMATOR FOR THE PHOTON FACTORY WIGGLER BEAMLINE
    KAWATA, H
    ANDO, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3): : 368 - 372
  • [39] A compact double crystal monochromator for electrochemistry beamline at PLS
    Rah, S
    Kim, SN
    Kim, GH
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 : 388 - 391
  • [40] Thermal bilayer monochromator crystal for a high resolution multiwavelength anomalous diffraction beamline at the European synchrotron radiation facility
    Ferrer, Jean-Luc
    Roth, Michel
    Fanchon, Eric
    Poupon, Gilles
    Saint-Lager, Marie-Claire
    Berar, J.-F.
    Review of Scientific Instruments, 1995, 66 (2 pt 2): : 2089 - 2091