Testing of a sCVD diamond detection system in the CROCUS reactor

被引:8
|
作者
Hursin, M. [1 ,2 ]
Weiss, C. [3 ]
Frajtag, P. [2 ]
Lamirand, V. [1 ,2 ]
Perret, G. [1 ]
Kavrigin, P. [3 ,4 ]
Pautz, A. [1 ,2 ]
Griesmayer, E. [3 ,4 ]
机构
[1] Paul Scherrer Inst, NES LRS, CH-5232 Villigen, Switzerland
[2] Ecole Polytech Fed Lausanne, LRS, CH-1015 Lausanne, Switzerland
[3] CIVIDEC Instrumentat, Vienna, Austria
[4] Vienna Univ Technol, Vienna, Austria
来源
EUROPEAN PHYSICAL JOURNAL A | 2018年 / 54卷 / 05期
关键词
D O I
10.1140/epja/i2018-12519-1
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
The paper describes the testing of the NEUTON detection system into CROCUS, the zero-power reactor of the Ecole Polytechnique Federale de Lausanne (EPFL). NEUTON is composed of a 4 mm x 4 mm sCVD diamond detector with a Li-6 converter and the associated acquisition electronics. It is developed by CIVIDEC Instrumentation GmbH. The use of a diamond detector with converter in the mixed radiation field of a nuclear reactor is challenging because these detectors are sensitive to gamma-rays, fast neutrons and thermal neutrons through conversion in Li-6. In NEUTON, the rejection of gamma-rays is achieved in real time, via the analysis of the signal pulse shape from the detector. To do so, a few signal characteristics (amplitude, area and FWHM) are recorded in the integrated Field Programmable Gate Arrays (FPGA) of the system. This treatment does not induce any dead time. Measurements in CROCUS demonstrated for the first time the capability of a system like NEUTON to detect and separate fast neutrons, thermal neutrons, and gamma-rays. The system response was shown to be linear with respect to the reactor power (up to 35 W) and its thermal sensitivity was found to be (3.5 +/- 0.2) x 10(-5) cps/nv.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] THERMOLUMINESCENT DOSIMETRY MONITORING SYSTEM FOR NATIONAL REACTOR TESTING STATION
    CIPPERLEY, FV
    HEALTH PHYSICS, 1966, 12 (11): : 1637 - +
  • [42] STRENGTH TESTING OF DIAMOND
    FIELD, JE
    HAUSER, HM
    HUTCHING.IM
    WOODWARD, AC
    INDUSTRIAL DIAMOND REVIEW, 1974, (JUL): : 255 - 259
  • [43] MALFUNCTION DETECTION OF NUCLEAR-REACTOR SYSTEM BY DYNAMIC DATA SYSTEM
    WU, SM
    CHOW, MC
    CAIN, D
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1976, 23 (JUN18): : 221 - 222
  • [44] REACTOR COOLANT PRESSURE BOUNDARY LEAKAGE DETECTION SYSTEM
    DISSING, E
    SVANSSON, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1980, 27 (01) : 769 - 775
  • [45] Contribution to the verification of the MORET 5 code using the CROCUS reactor physics benchmark
    Leclaire, N.
    Cochet, B.
    ANNALS OF NUCLEAR ENERGY, 2021, 155
  • [46] Laser plasma CVD diamond reactor
    Bolshakov, AP
    Konov, VI
    Prokhorov, AM
    Uglov, SA
    Dausinger, F
    DIAMOND AND RELATED MATERIALS, 2001, 10 (9-10) : 1559 - 1564
  • [47] Design of an Open-Loop Pile-Oscillation Program in the CROCUS Reactor
    Ligonnet, Thomas
    Laureau, Axel
    Simbsler, Philippe
    Pautz, Andreas
    Lamirand, Vincent
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2024, 71 (05) : 1001 - 1009
  • [48] A benchmark on the calculation of kinetic parameters based on reactivity effect experiments in the CROCUS reactor
    Paratte, J. M.
    Frueh, R.
    Kasemeyer, U.
    Kalugin, M. A.
    Timm, W.
    Chawla, R.
    ANNALS OF NUCLEAR ENERGY, 2006, 33 (08) : 739 - 748
  • [49] USE OF CROCUS REACTOR IN LABORATOIRE-DE-GENIE-ATOMIQUE OF EPF, LAUSANNE
    TAI, AS
    ZEITSCHRIFT FUR ANGEWANDTE MATHEMATIK UND PHYSIK, 1971, 22 (04): : 792 - &
  • [50] PISTIL, a reactivity modulation device to probe the transfer function of the nuclear reactor CROCUS
    Jiang, Yifeng
    Geslot, Benoit
    Lamirand, Vincent
    Leconte, Pierre
    Godat, Daniel
    Braun, Laurent
    Frajtag, Pavel
    Coquelet-Pascal, Christine
    Pautz, Andreas
    ANIMMA 2021 - ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS, 2021, 253