Testing of a sCVD diamond detection system in the CROCUS reactor

被引:8
|
作者
Hursin, M. [1 ,2 ]
Weiss, C. [3 ]
Frajtag, P. [2 ]
Lamirand, V. [1 ,2 ]
Perret, G. [1 ]
Kavrigin, P. [3 ,4 ]
Pautz, A. [1 ,2 ]
Griesmayer, E. [3 ,4 ]
机构
[1] Paul Scherrer Inst, NES LRS, CH-5232 Villigen, Switzerland
[2] Ecole Polytech Fed Lausanne, LRS, CH-1015 Lausanne, Switzerland
[3] CIVIDEC Instrumentat, Vienna, Austria
[4] Vienna Univ Technol, Vienna, Austria
来源
EUROPEAN PHYSICAL JOURNAL A | 2018年 / 54卷 / 05期
关键词
D O I
10.1140/epja/i2018-12519-1
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
The paper describes the testing of the NEUTON detection system into CROCUS, the zero-power reactor of the Ecole Polytechnique Federale de Lausanne (EPFL). NEUTON is composed of a 4 mm x 4 mm sCVD diamond detector with a Li-6 converter and the associated acquisition electronics. It is developed by CIVIDEC Instrumentation GmbH. The use of a diamond detector with converter in the mixed radiation field of a nuclear reactor is challenging because these detectors are sensitive to gamma-rays, fast neutrons and thermal neutrons through conversion in Li-6. In NEUTON, the rejection of gamma-rays is achieved in real time, via the analysis of the signal pulse shape from the detector. To do so, a few signal characteristics (amplitude, area and FWHM) are recorded in the integrated Field Programmable Gate Arrays (FPGA) of the system. This treatment does not induce any dead time. Measurements in CROCUS demonstrated for the first time the capability of a system like NEUTON to detect and separate fast neutrons, thermal neutrons, and gamma-rays. The system response was shown to be linear with respect to the reactor power (up to 35 W) and its thermal sensitivity was found to be (3.5 +/- 0.2) x 10(-5) cps/nv.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] The CAP-reactor, a novel microwave CVD system for diamond deposition
    Pleuler, E
    Wild, C
    Füner, M
    Koidl, P
    DIAMOND AND RELATED MATERIALS, 2002, 11 (3-6) : 467 - 471
  • [22] FFTF REACTOR SYSTEM TESTING PROGRAM RESULTS AND OVERVIEW
    PECKINPAUGH, CL
    FARABEE, OA
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1980, 34 (JUN): : 707 - 709
  • [23] Delayed gamma fraction determination in the zero power reactor CROCUS
    Pakari, Oskari
    Mager, Tom
    Lamirand, Vincent
    Frajtag, Pavel
    Pautz, Andreas
    EPJ NUCLEAR SCIENCES & TECHNOLOGIES, 2021, 7
  • [24] In-core dosimetry for the validation of neutron spectra in the CROCUS reactor
    Jiang, Y.
    Laureau, A.
    Lamirand, V
    Frajtag, P.
    Pautz, A.
    ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA 2019), 2020, 225
  • [25] REACTOR SAFETY CHANNEL SYSTEM WITH ONLINE TESTING CAPABILITY
    SPRACKLEN, HP
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (01) : 662 - 654
  • [26] DELPHIN - a new system for testing reactor pressure vessels
    Dressler, K
    ATW-INTERNATIONALE ZEITSCHRIFT FUR KERNENERGIE, 1999, 44 (06): : 369 - +
  • [27] Testing of a reactor control system in closed-loop
    Petruzela, Ivan
    2007 IEEE INTERNATIONAL SYMPOSIUM ON INDUSTRIAL ELECTRONICS, PROCEEDINGS, VOLS 1-8, 2007, : 314 - 317
  • [28] Design of a 150-miniature detectors 3D core-mapping system for the CROCUS reactor
    Vitullo, Fanny
    Lamirand, Vincent
    Ambrozic, Klemen
    Braun, Laurent
    Godat, Daniel
    Frajtag, Pavel
    Pautz, Andreas
    ANIMMA 2021 - ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS, 2021, 253
  • [29] EXPERIMENTS OF ANOMALY DETECTION SYSTEM OF A REACTOR CORE
    OHSAWA, Y
    KATO, K
    OYAMADA, R
    NUCLEAR TECHNOLOGY, 1974, 23 (01) : 5 - 13
  • [30] Diamond UV sensor used in fire detection system
    不详
    INDUSTRIAL DIAMOND REVIEW, 2006, (02): : 8 - 8