Commissioning of modulator-based IMRT with XiO treatment planning system

被引:8
|
作者
Oguchi, Hiroshi [1 ,2 ]
Obata, Yasunori [3 ]
机构
[1] Nagoya Univ, Grad Sch Med, Program Radiol & Med Lab Sci, Higashi Ku, Aichi 4618673, Japan
[2] Shinshu Univ Hosp, Dept Radiol, Nagano 3908621, Japan
[3] Nagoya Univ, Sch Hlth Sci, Dept Radiol Technol, Higashi Ku, Aichi 4618673, Japan
关键词
radiation therapy; intensity modulation; physical modulator; dose verification; ATTENUATION COEFFICIENTS; CONFORMAL RADIOTHERAPY; INTENSITY; DELIVERY; COMPENSATORS; FIELDS; VERIFICATION; GENERATION; SCATTER; BEAMS;
D O I
10.1118/1.2996285
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This article describes the procedures for correction of the modulator thickness and commissioning of the XiO treatment planning system (TPS) for modulator-based intensity modulated radiation therapy (M-IMRT). This modulator manufacturing system adopts a method in which the modulator is milled using a floor-type computer-aided numerical control milling machine (CNC-mill) with modulator data calculated by XiO TPS. XiO TPS uses only effective attenuation coefficients (EAC) for modulator thickness calculation. This article describes a modified method for assessing modulator thickness. A two-dimensional linear attenuation array was used to correct the modulator thickness calculated by XiO. Narrow-beam geometry was used for measuring the linear attenuation coefficient (LAC) at off-axis positions (OAP) for varying brass thicknesses. An equation for the two-dimensional LAC ratio (2D-LACR) can be used to calculate the corrected modulator thickness. It is assumed that the broad beam EAC of a small field varies with the brass thickness and the OAP distance in the same way as that of LACR, so the two-dimensional EAC (2D-EAC) is equal to the EAC corrected using the LACR. The dose distribution was evaluated for three geometric patterns and one clinical case on low energy x ray (4 MV) with a large field size (20 x 20 cm(2)). The results using the proposed correction method of modulator thickness showed a good agreement between the measured dose distributions and the dose distributions calculated by TPS with the correction. Hence, the method is effective to improve the accuracy of M-IMRT in XiO TPS. An important problem for the brass modulator is the milling condition, such as the drill diameter and the cutting pitch size. It is necessary to improve the accuracy of M-IMRT for the "softening" and "hardening" effects of the beam to be considered in dose calculation in patients and the modulator profile design. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.2996285]
引用
收藏
页码:261 / 269
页数:9
相关论文
共 50 条
  • [1] Commissioning Compensator-Based IMRT in the Pinnacle Treatment Planning System
    Opp, D.
    Feygelman, V.
    Forster, K.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [2] Commissioning compensator-based IMRT on the Pinnacle treatment planning system
    Opp, Daniel
    Forster, Kenneth
    Feygelman, Vladimir
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 310 - 325
  • [3] Comparison of the KonRad IMRT and XiO treatment planning systems
    Reitz, Bodo
    Miften, Moyed
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (03): : 122 - 135
  • [4] Commissioning of Monaco Monte Carlo IMRT Treatment Planning System
    Li, J.
    Doemer, A.
    Cao, J.
    Podder, T.
    Harrison, A.
    Yu, Y.
    Galvin, J.
    Xiao, Y.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [5] Quantitative Tools to Assess the Fidelity of Treatment Planning System Commissioning for IMRT
    Yan, G.
    Liu, C.
    Palta, J.
    Li, J.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [6] Commissioning of a commercial IMRT treatment planning system for standard elekta MLC
    Kim, H
    Chen, Y
    Xiao, Y
    Anne, P
    Galvin, J
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1348 - 1348
  • [7] Independent verification for XiO treatment planning system
    Oliveira, A
    Germano, S
    Pereira, P
    Batel, V
    Marques, F
    Almeida, P
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S195 - S196
  • [8] Commissioning and quality assurance of a commercial stereotactic treatment-planning system for extracranial IMRT
    Wang, Lu
    Li, Jinsheng
    Paskalev, Kamen
    Hoban, Peter
    Luo, Wei
    Chen, Lili
    McNeeley, Shawn
    Price, Robert
    Ma, Charlie
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (01): : 21 - 34
  • [9] Implementation of enhanced dynamic wedges into XIO treatment planning system
    Rodrigues, C.
    Batel, V.
    Germano, S.
    Monteiro Grillo, I.
    Pinto, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S512 - S513
  • [10] A New Precision Method to Do IMRT Verification Work in Tomotherapy Treatment Planning System Commissioning
    Pang, T.
    Yu, L.
    Yang, B.
    Qiu, J.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E487 - E487