Commissioning and quality assurance of a commercial stereotactic treatment-planning system for extracranial IMRT

被引:11
|
作者
Wang, Lu [1 ]
Li, Jinsheng [1 ]
Paskalev, Kamen [1 ]
Hoban, Peter [1 ]
Luo, Wei [1 ]
Chen, Lili [1 ]
McNeeley, Shawn [1 ]
Price, Robert [1 ]
Ma, Charlie [1 ]
机构
[1] Fox Chase Canc Ctr, Dept Radiat Oncol, Philadelphia, PA 19111 USA
来源
关键词
stereotactic IMRT; Monte Carlo; treatment-planning system;
D O I
10.1120/jacmp.2027.25368
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A 3D treatment-planning system (TPS) for stereotactic intensity-modulated radiotherapy (IMRT) using a micro-multileaf collimator has been made available by Radionics. In this work, we report our comprehensive quality assurance (QA) procedure for commissioning this TPS. First, the accuracy of stereotaxy established with a body frame was checked to ensure accurate determination of a target position within the planning system. Second, the CT-to-electron density conversion curve used in the TPS was fitted to our site-specific measurement data to ensure the accuracy of dose calculation and measurement verification in a QA phantom. Using the QA phantom, the radiological path lengths were verified against known geometrical depths to ensure the accuracy of the ray-tracing algorithm. We also checked inter-and intraleaf leakage/transmission for adequate jaw settings. Measurements for dose verification were performed in various head/neck and prostate IMRT treatment plans using the patient-specific optimized fluence maps. Both ion chamber and film were used for point dose and isodose distribution verifications. To ensure that adjacent organs at risk receive dose within the expectation, we used the Monte Carlo method to calculate dose distributions and dose-volume histograms (DVHs) for these organs at risk. The dosimetric accuracy satisfied the published acceptability criteria. The Monte Carlo calculations confirmed the measured dose distributions for target volumes. For organs located on the beam boundary or outside the beam, some differences in the DVHs were noticed. However, the plans calculated by both methods met our clinical criteria. We conclude that the accuracy of the XKnife (TM) RT2 treatment-planning system is adequate for the clinical implementation of stereotactic IMRT.
引用
收藏
页码:21 / 34
页数:14
相关论文
共 50 条
  • [1] Validation of physics improvements for IMRT with a commercial treatment-planning system
    Cadman, Patrick
    McNutt, Todd
    Bzdusek, Karl
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2005, 6 (02): : 74 - 86
  • [2] A comparison of two commercial treatment-planning systems for IMRT
    Petric, M. Peter
    Clark, Brenda G.
    Robar, James L.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2005, 6 (03): : 63 - 80
  • [3] Commissioning and Quality Assurance of the First Commercial Hybrid MRI-IMRT System
    Green, O. Pechenaya
    Goddu, S.
    Mutic, S.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3785 - 3785
  • [4] Commercial IMRT treatment-planning dose-calculation algorithm
    Low, DA
    Mutic, S
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (04): : 933 - 937
  • [5] Commissioning of a commercial IMRT treatment planning system for standard elekta MLC
    Kim, H
    Chen, Y
    Xiao, Y
    Anne, P
    Galvin, J
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1348 - 1348
  • [6] IMRT QUALITY ASSURANCE USING A SECOND TREATMENT PLANNING SYSTEM
    Anjum, Muhammad Naeem
    Parker, William
    Ruo, Russell
    Aldahlawi, Ismail
    Afzal, Muhammad
    [J]. MEDICAL DOSIMETRY, 2010, 35 (04) : 274 - 279
  • [7] COMMISSIONING AND QUALITY ASSURANCE OF TREATMENT PLANNING COMPUTERS
    VANDYK, J
    BARNETT, RB
    CYGLER, JE
    SHRAGGE, PC
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (02): : 261 - 273
  • [8] COMMISSIONING AND QUALITY ASSURANCE OF TREATMENT PLANNING COMPUTERS - COMMENTS
    MCSHAN, DL
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (02): : 371 - 372
  • [9] A comprehensive commissioning procedure for stereotactic extracranial IMRT using a micro multileaf collimator
    Wang, L
    Paskalev, K
    Li, J
    Price, R
    Chen, L
    Fourkal, E
    Ding, M
    Yang, J
    Xiong, W
    Ma, C
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1499 - 1500
  • [10] Commissioning of Monaco Monte Carlo IMRT Treatment Planning System
    Li, J.
    Doemer, A.
    Cao, J.
    Podder, T.
    Harrison, A.
    Yu, Y.
    Galvin, J.
    Xiao, Y.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)