Effect of particle size of Fully Porous-Coated (FPC) implants on osseointegration

被引:1
|
作者
Rungsiyakull, Chaiy [1 ]
Li, Qing [1 ]
Li, Wei [1 ]
Appleyard, Richard [2 ]
Swain, Michael [3 ]
机构
[1] Univ Sydney, Sch Aerosp Mech & Mechatron Engn, Sydney, NSW 2006, Australia
[2] Univ Sydney, Royal N Shore Hosp, Murray Maxwell Biomechan Lab, Sydney, NSW 2006, Australia
[3] Univ Sydney, Fac Dent, Sydney, NSW 2006, Australia
基金
澳大利亚研究理事会;
关键词
dental implant; coating; osseointegration; finite element analysis; multiscale modelling;
D O I
10.4028/www.scientific.net/AMR.47-50.916
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper aims at providing a preliminary understanding in biomechanics with respect to the effect of the particle size of Fully Porous-Coated (FPC) dental implant on osseointegration. 2D multiscale finite element models are created for a typical dental implantation setting. Under a certain mastication force (<200N), a global response is first obtained from a macro-scale model (without considering morphological details on the coated surface), and then it is transferred to micro-scale models (with coated surface morphology details in three different particle sizes). An equivalent strain is analyzed to investigate the effect of particle size of the FPC materials on osseointegration and initiation of bone remodelling. The result reveals that increasing particle sizes has a significant effect on biomechanical and bone remodelling responses.
引用
收藏
页码:916 / +
页数:2
相关论文
共 50 条
  • [41] The effect of adjuvant calcium phosphate coating on a porous-coated femoral stem
    Tanzer, M
    Gollish, J
    Leighton, R
    Orrell, K
    Giacchino, A
    Welsh, P
    Shea, B
    Wells, G
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (424) : 153 - 160
  • [42] Improved fixation of porous-coated versus grit-blasted surface texture of hydroxyapatite-coated implants in dogs
    Overgaard, S
    Lind, M
    Rahbek, O
    Bunger, C
    Soballe, K
    ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 (04): : 337 - 343
  • [43] EFFECT OF THE SURFACE GEOMETRY OF SMOOTH AND POROUS-COATED TITANIUM-ALLOY
    INOUE, T
    YAMAMURA, T
    MELCHER, AH
    JOURNAL OF DENTAL RESEARCH, 1986, 65 (04) : 583 - 583
  • [44] Initial healing in the dog of submerged versus non-submerged porous-coated endosseous dental implants
    Levy, D
    Deporter, DA
    Pilliar, RM
    Watson, PA
    Valiquette, N
    CLINICAL ORAL IMPLANTS RESEARCH, 1996, 7 (02) : 101 - 110
  • [45] Periodontal parameters around porous-coated dental implants after 3 to 4 years supporting overdentures
    Levy, D
    Deporter, DA
    Watson, PA
    Pilliar, RM
    JOURNAL OF CLINICAL PERIODONTOLOGY, 1996, 23 (06) : 517 - 522
  • [46] EXPERIMENTAL-DETERMINATION OF FRICTION CHARACTERISTICS AT THE TRABECULAR BONE POROUS-COATED METAL INTERFACE IN CEMENTLESS IMPLANTS
    SHIRAZIADL, A
    DAMMAK, M
    PAIEMENT, G
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (02): : 167 - 175
  • [47] Demineralized bone matrix and human cancellous bone enhance fixation of porous-coated titanium implants in sheep
    Babiker, Hassan
    Ding, Ming
    Overgaard, Soren
    JOURNAL OF TISSUE ENGINEERING AND REGENERATIVE MEDICINE, 2016, 10 (03) : 245 - 251
  • [48] A comparison of radiographic bone height and probing attachment level measurements adjacent to porous-coated dental implants in humans
    Levy, D
    Deporter, DA
    Pharoah, M
    Tomlinson, G
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1997, 12 (04) : 541 - 546
  • [49] EFFECT OF LENGTH-TO-DIAMETER RATIO ON CRITICAL HEAT FLUX IN POROUS-COATED TUBES
    Yildiz, Sebiha
    THERMAL SCIENCE, 2021, 25 (01): : 613 - 623
  • [50] THE EFFECT OF LOW-DOSE METHOTREXATE UPON BONE INGROWTH IN A POROUS-COATED IMPLANT
    POPP, JE
    STERN, LS
    WASIELEWSKI, RC
    JOURNAL OF BONE AND MINERAL RESEARCH, 1995, 10 : S437 - S437