摘要
We have developed and fabricated SQUID gradiometers in a Nb/Al-AlOx/Nb technology using an i-line stepper combined with a mask aligner lithography. The stepper and the aligner are used to fabricate fine-scale patterns such as the Josephson junctions and large-scale patterns such as pickup coil, respectively, on the same wafer with a 4-inch diameter. The SQUID is designed as a first-order gradiometer with a dual-washer configuration. Two sets of multiturn spiral input coils with opposite directions are integrated on the washer and connected to the pickup coil to form a flux transformer. In order to improve the balance of the gradiometer, a fully symmetric configuration is introduced with a mirrored SQUID. The pickup coil is a first-order gradiometric type with a base line of 3.5 cm. The size of each loop is 12.65 mm X 11.65 mm, which results in a gradient magnetic field sensitivity of 0.25 nT/(cm.Phi(0)). The measured gradient magnetic field noise is 75 fT/(m.root Hz) in the white noise region. Moreover, a sapphire holder is used to fix and package the chip in order to maintain the high intrinsic balance (typical 10(-4)). It makes the SQUID gradiometer more attractive for an unshielded operation in noisy environments.
引用
收藏
页数:3
相关论文
共 50 条
  • [21] A PLANAR 2ND-ORDER DC-SQUID GRADIOMETER
    CARELLI, P
    CHIAVENTI, L
    LEONI, R
    PULLANO, M
    SPAGNOLO, GS
    CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1991, 12 : 13 - 19
  • [22] Noise Compensation of a Mobile LTS SQUID Planar Gradiometer for Aeromagnetic Detection
    Song, Zhengwei
    Dai, Haibin
    Rong, Liangliang
    Dong, Hui
    Wu, Jun
    Qiu, Longqing
    Zhang, Guofeng
    Wang, Yongliang
    Tao, Quan
    Pei, Yifeng
    Zhang, Shulin
    Xie, Xiaoming
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2019, 29 (08)
  • [23] 3 SQUID GRADIOMETER
    KOCH, RH
    ROZEN, JR
    SUN, JZ
    GALLAGHER, WJ
    APPLIED PHYSICS LETTERS, 1993, 63 (03) : 403 - 405
  • [24] High-Tc SQUID gradiometer system for immunoassays
    Oisjoen, F.
    Magnelind, P.
    Kalabukhov, A.
    Winkler, D.
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2008, 21 (03):
  • [25] 2ND ORDER GRADIOMETER AND DC SQUID INTEGRATED ON A PLANAR SUBSTRATE
    VANNIEUWENHUYZEN, GJ
    DEWAAL, VJ
    APPLIED PHYSICS LETTERS, 1985, 46 (04) : 439 - 441
  • [26] YBCO step-edge junction DC-squid planar gradiometer
    Han, B
    Chen, GH
    Xu, FZ
    Zhao, SP
    Yang, QS
    ACTA PHYSICA SINICA, 2000, 49 (10) : 2051 - 2054
  • [27] YBCO step-edge junction dc-SQUID planar gradiometer
    2000, Sci Publ House (49):
  • [28] OPTIMIZATION OF A SQUID VECTOR GRADIOMETER
    LEKKALA, JO
    MALMIVUO, JAV
    CRYOGENICS, 1985, 25 (06) : 291 - 303
  • [29] First-order gradiometer of high Tc rf SQUID
    He, DF
    Zhang, Y
    PHYSICA C, 1997, 282 : 2481 - 2482
  • [30] First-order high-Tc SQUID gradiometer
    Kong, X. Y.
    Nakatani, Y.
    Yutani, A.
    Maki, T.
    Itozaki, H.
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2008, 468 (15-20): : 1946 - 1949