Design and Fabrication of a CdMnTe Nuclear Radiation Detection System

被引:1
|
作者
Egarievwe, Stephen U. [1 ]
Israel, Mordecai B. [1 ]
Banks, Alex D. [1 ]
Drabo, Mebougna L. [1 ]
Dunning, Keiandrea L. [1 ]
Cook, Vintrell J. [1 ]
Johnson, Frederick D. [1 ]
Palmer, Shaunay M. [1 ]
Roy, Utpal N. [2 ]
James, Ralph B. [3 ]
机构
[1] Alabama A&M Univ, Dept Elect Engn, Normal, AL 35762 USA
[2] Brookhaven Natl Lab, Dept Noneproliferat & Natl Secur, Upton, NY 11973 USA
[3] Savannah River Natl Lab, Sci & Technol, Aiken, SC USA
来源
2019 IEEE SOUTHEASTCON | 2019年
关键词
CdMnTe nuclear detectors; gamma-rays; X-rays; CDZNTE X-RAY; CONTACTS; PERFORMANCE;
D O I
10.1109/southeastcon42311.2019.9020612
中图分类号
TP301 [理论、方法];
学科分类号
081202 ;
摘要
Cadmium manganese telluride (CdMnTe) is one of the ternary compounds of cadmium telluride (CdTe) that has shown great potential applications in the development of X-ray and gamma-ray detectors used in medical imaging and in the detection of radiological and nuclear threats. CdMnTe nuclear detectors are capable of operating at room temperature without cryogenic cooling. The goal of this project is to design and fabricate CdMnTe detection system that can be used in laboratory experiments for characterizing detector materials and for identification of radionuclides. The engineering design objectives include a spring-based sample holder, ability to measure electrical properties of CdTe-based semiconductor nuclear detectors, and the detection of nuclear radiations. This paper presents the design of the sample holder and the fabrication of a CdMnTe detector. The current-voltage characterization experimental results showed that the CdMnTe has the high resistivity needed for CdTe-based nuclear detectors. The results also include the response of the CdiVInTe detector to the 59.5-keV gamma peak of Am-241 nuclear radiation source.
引用
收藏
页数:4
相关论文
共 50 条
  • [41] Radiation Detection System
    Tatarnikov, Denis A.
    Godovykh, Aleksey V.
    HIGH TECHNOLOGY: RESEARCH AND APPLICATIONS, 2014, 1040 : 980 - 984
  • [42] Investigation of the Charge Collection Efficiency of CdMnTe Radiation Detectors
    Rafiei, R.
    Boardman, D.
    Sarbutt, A.
    Prokopovich, D. A.
    Kim, K.
    Reinhard, M. I.
    Bolotnikov, A. E.
    James, R. B.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (03) : 634 - 641
  • [43] Plastic scintillator-based radiation detector for mobile radiation detection system against nuclear/radiological terrorism
    Kwak, Sung-Woo
    Yoo, Ho-Sik
    Jang, Sung Soon
    Kim, Jung Soo
    Yoon, Wan-Ki
    Jun, In Sub
    Kim, Kwang Hyun
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 604 (1-2): : 161 - 163
  • [44] Design features and engineering practice of radiation monitoring system in CEPR nuclear power plant
    Peng, Xuewei
    Wang, Ensheng
    Ye, Jianqiang
    Ma, Boyang
    Hedongli Gongcheng/Nuclear Power Engineering, 2015, 36 : 63 - 66
  • [45] Web Design and Development for Centralize Area Radiation Monitoring System in Malaysian Nuclear Agency
    Ibrahim, Maslina Mohd
    Yussup, Nolida
    Haris, Mohd Fauzi
    Shaari, Syirrazie Che Soh
    Azman, Azraf
    Razalim, Faizal Azrin B. Abdul
    Yapp, Raymond
    Hasim, Harzawardi
    Aslan, Mohd Dzul Aiman
    ADVANCING NUCLEAR SCIENCE AND ENGINEERING FOR SUSTAINABLE NUCLEAR ENERGY KNOWLEDGE, 2017, 1799
  • [46] Design, fabrication, and evaluation a laboratory dry-peeling system for hazelnut using infrared radiation
    Eskandari, Jalal
    Kermani, Ali M.
    Kouravand, Shahriar
    Zarafshan, Payam
    LWT-FOOD SCIENCE AND TECHNOLOGY, 2018, 90 : 570 - 576
  • [47] Millimeter wave detection of nuclear radiation: An alternative detection mechanism
    Gopalsami, N.
    Chien, H. T.
    Heifetz, A.
    Koehl, E. R.
    Raptis, A. C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (08):
  • [48] Characterization of CdMnTe radiation detectors using current and charge transients
    R.Rafiei
    M.I.Reinhard
    A.Sarbutt
    S.Uxa
    D.Boardman
    G.C.Watt
    E.Belas
    K.Kim
    A.E.Bolotnikov
    R.B.James
    Journal of Semiconductors, 2013, (07) : 19 - 25
  • [49] NUCLEAR RADIATION DETECTION BY SOLID STATE DEVICES
    DEARNALEY, G
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1966, 43 (12): : 869 - +
  • [50] Covariance spectroscopy applied to nuclear radiation detection
    Trainham, R.
    Tinsley, J.
    Keegan, R.
    Quam, W.
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XIII, 2011, 8142