Design and Ex Vivo Evaluation of a 3D High Intensity Focused Ultrasound System for Tumor Treatment with Tissue Ablation

被引:0
|
作者
Lweesy, K. [1 ]
Fraiwan, L. [1 ]
Al-Shalabi, M. [1 ]
Mohammad, L. [1 ]
Al-Oglah, R. [1 ]
机构
[1] Jordan Univ Sci & Technol, Fac Engn, Dept Biomed Engn, Irbid 22110, Jordan
关键词
Geometrically focused transducer; high intensity focused ultrasound; lesion; sonication; treatment planning; BREAST-CANCER;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the design, construction, and evaluation of a three dimensional (3D) ultrasound system to be used to treat different kinds of tumors using high intensity focused ultrasound (HIFU). The system consists of two major parts: an ultrasonic therapy part and a treatment planning part. The ultrasonic therapy part consists of an ultrasound bowl shaped transducer (made from Lead Zirconate Titanate (PZT) and has a resonance frequency of 0.5 MHz), a lossless electrical impedance matching circuit built to ensure maximum electrical power delivery to the transducer, a function generator, and a high power amplifier. The ultrasonic therapy part is responsible for generating a high-power focus at the location of the geometric focus of the bowl shaped ultrasound transducer. The treatment planning part consists of three stepper motors (responsible for moving the setup in the x- y- and z-directions), three high-voltage high-current darlington arrays (to supply the stepper motors with the required voltages and currents), and a C# software to perform the treatment planning. To assess the movement of the treatment planner, each of the three stepper motors was moved forward and backward from end to end. Then the treatment planner was successfully driven to cover cubes of dimensions of 1 x 1 x 1 cm(3), 2 x 2 x 2 cm(3), 4 x 4 x 4 cm(3), and 8 x 8 x 8 cm(3), with step sizes 0.5, 1, 2, and 4 mm, respectively. Ex vivo experiments using fresh bovine liver were performed and indicated the capability of the system to generate lesions both on- and off-axis. Lesions at different depths were successfully generated at the intended locations. Temperature distributions were recorded both inside and outside the lesion and indicated that the temperature reached about 60 degrees C inside the lesion and remained below 39 degrees C outside it.
引用
收藏
页码:556 / 559
页数:4
相关论文
共 50 条
  • [1] Design and ex vivo kidney evaluation of a high-intensity focused ultrasound transducer and 3D positioner
    Khaldon Lweesy
    Luay Fraiwan
    Alaa Shatat
    Ghassan Abdo
    Abduallah Dawodiah
    Mohammad Sameer
    Medical & Biological Engineering & Computing, 2010, 48 : 269 - 276
  • [2] Design and ex vivo kidney evaluation of a high-intensity focused ultrasound transducer and 3D positioner
    Lweesy, Khaldon
    Fraiwan, Luay
    Shatat, Alaa
    Abdo, Ghassan
    Dawodiah, Abduallah
    Sameer, Mohammad
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2010, 48 (03) : 269 - 276
  • [3] High-intensity focused ultrasound treatment for skin: ex vivo evaluation
    Park, J. -H.
    Lim, S. -D.
    Oh, S. H.
    Lee, J. H.
    Yeo, U. C.
    SKIN RESEARCH AND TECHNOLOGY, 2017, 23 (03) : 384 - 391
  • [4] EX VIVO CHARACTERIZATION OF CANINE LIVER TISSUE VISCOELASTICITY AFTER HIGH-INTENSITY FOCUSED ULTRASOUND ABLATION
    Shahmirzadi, Danial
    Hou, Gary Y.
    Chen, Jiangang
    Konofagou, Elisa E.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (02): : 341 - 350
  • [5] Control parameters for high-intensity focused ultrasound (HIFU) for tissue ablation in the ex-vivo kidney
    Köhrmann, KU
    Michel, MS
    Steidler, A
    Marlinghaus, EH
    Kraut, O
    Alken, P
    AKTUELLE UROLOGIE, 2002, 33 (01) : 58 - 63
  • [6] HIGH-INTENSITY FOCUSED ULTRASOUND ABLATION OF EX VIVO BOVINE ACHILLES TENDON
    Muratore, Robert
    Akabas, Tal
    Muratore, Isabella B.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2008, 34 (12): : 2043 - 2050
  • [7] Characterization and Ex Vivo evaluation of an extracorporeal high-intensity focused ultrasound (HIFU) system
    Zhou, Yufeng
    Cunitz, Bryan W.
    Dunmire, Barbrina
    Wang, Yak-Nam
    Karl, Steven G.
    Warren, Cinderella
    Mitchell, Stuart
    Hwang, Joo Ha
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (09): : 345 - 359
  • [8] Experimental evaluation of high intensity focused ultrasound for fat reduction of ex vivo porcine adipose tissue
    Antria Filippou
    Christakis Damianou
    Journal of Ultrasound, 2022, 25 : 815 - 825
  • [9] Experimental evaluation of high intensity focused ultrasound for fat reduction of ex vivo porcine adipose tissue
    Filippou, Antria
    Damianou, Christakis
    JOURNAL OF ULTRASOUND, 2022, 25 (04) : 815 - 825
  • [10] High-intensity focused ultrasound for ex vivo kidney tissue ablation:: Influence of generator power and pulse duration
    Häcker, A
    Köhrmann, KU
    Knoll, T
    Langbein, S
    Steidler, A
    Kraut, O
    Marlinghaus, E
    Alken, P
    Michel, MS
    JOURNAL OF ENDOUROLOGY, 2004, 18 (09) : 917 - 924