The Cross-Talk Between Spinal Cord Stimulators and the Confirm™ Cardiac Monitor

被引:1
|
作者
Chaiban, Gassan Mitri [1 ]
Davidov, Mark [2 ]
Cummings, Jennifer Eiko [3 ]
Lehrian, Laura Elizabeth [2 ]
Yee, Eleanore Mun Yun [2 ]
Dugan, Amy [2 ]
Almualim, Mohammed Salman [1 ]
Atallah, Joseph N. [2 ]
机构
[1] Univ Toledo, Med Ctr, Toledo, OH 43606 USA
[2] Univ Toledo, Toledo, OH 43606 USA
[3] Univ Toledo, Dept Internal Med, Div Cardiol, Med Ctr, Toledo, OH 43606 USA
来源
NEUROMODULATION | 2014年 / 17卷 / 01期
关键词
ECG recording; electromagnetic interference; interaction; loop recorder; spinal cord stimulators; IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; REFRACTORY ANGINA; PATIENT; NEUROSTIMULATORS; INTERFERENCE; DEVICES;
D O I
10.1111/ner.12060
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
IntroductionSpinal cord stimulators (SCS) function by transmission of electrical impulses to electrode contacts placed within the epidural space depending on the painful area to be treated. Because of the electrical nature of the SCS, there has been concern about the interaction between these devices and devices that monitor or augment the cardiac system. Implantable loop recorders help to identify the causes of syncope or palpitation by continuously evaluating and recording portions of an electrocardiograph in patients being evaluated for cardiac conduction arrhythmias. The purpose of the study is to simulate the possible effects of spinal cord stimulation on a Confirm cardiac monitor (St. Jude Medical, St. Paul, MN, USA). MethodsTwenty patients without preexisting cardiac disease, who were successfully being treated with SCS, were enrolled. Confirm loop recorders (St. Jude Medical) were placed on their chest wall in a noninvasive manner, with all programmed at identical settings. Multiple stimulation settings were adjusted on the stimulators and the recordings from the Confirm loop recorder were analyzed for evidence of interference. ResultsFifteen of the patients had no electrical noise detected at any of the tested combinations of stimulation. Five patients had some electrical noise detected by the loop recorder, but it did not inhibit the cardiologist evaluating the recording from analyzing the electrocardiograph for diagnostic purposes. At no point with any of the patients at any tested setting was there an appearance of a life-threatening arrhythmia. ConclusionOur study demonstrates that spinal cord stimulation is unlikely to interfere with the data collected by the Confirm loop recorder, and the presence of an SCS should not interfere with the ability to use a loop recorder for diagnostic purposes.
引用
收藏
页码:42 / 47
页数:6
相关论文
共 50 条
  • [41] Cross-talk between apparently independent receptors
    Searl, TJ
    Silinsky, EM
    JOURNAL OF PHYSIOLOGY-LONDON, 1998, 513 (03): : 629 - 629
  • [42] Cross-talk between the kidney and the cardiovascular system
    Amann, Kerstin
    Wanner, Christoph
    Ritz, Eberhard
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2006, 17 (08): : 2112 - 2119
  • [43] Cross-talk between arterioles and tubules in the kidney
    Ren, Yilin
    Garvin, Jeffrey L.
    Liu, Ruisheng
    Carretero, Oscar A.
    PEDIATRIC NEPHROLOGY, 2009, 24 (01) : 31 - 35
  • [44] Cross-talk between mast cells and eosinophils
    Levi-Schaffer, F
    ALLERGY, 1999, 54 : 36 - 38
  • [45] Cross-talk between adenosine and opioid receptors
    Peart, JN
    Gross, GJ
    DRUG NEWS & PERSPECTIVES, 2005, 18 (04) : 237 - 242
  • [46] THEORY OF CROSS-TALK BETWEEN TWISTED PAIRS
    BELEVITCH, V
    PHILIPS RESEARCH REPORTS, 1977, 32 (5-6): : 365 - 372
  • [47] Cross-talk between arterioles and tubules in the kidney
    YiLin Ren
    Jeffrey L. Garvin
    Ruisheng Liu
    Oscar A. Carretero
    Pediatric Nephrology, 2009, 24 : 31 - 35
  • [48] Cross-Talk between Apolipoprotein E and Cytokines
    Zhang, Hongliang
    Wu, Li-Min
    Wu, Jiang
    MEDIATORS OF INFLAMMATION, 2011, 2011
  • [49] Optical cross-talk between quantum wells
    Weber, D
    Hellmann, R
    Feldmann, J
    Gobel, EO
    Citrin, DS
    Kohler, K
    ULTRAFAST PROCESSES IN SPECTROSCOPY, 1996, : 189 - 193
  • [50] The cross-talk between the skeleton and energy metabolism
    Liu, Jian-min
    JOURNAL OF DIABETES, 2013, 5 (01) : 10 - 12