Droplet detector for the continuous flow luminol-hydrogen peroxide chemiluminescence system

被引:6
|
作者
Wen, Yaqiong [1 ]
Yuan, Hongyan [2 ]
Mao, Jianfei [1 ]
Xiao, Dan [1 ,2 ]
Choi, Martin M. F. [3 ]
机构
[1] Sichuan Univ, Coll Chem, Chengdu 610065, Peoples R China
[2] Sichuan Univ, Coll Chem Engn, Chengdu 610065, Peoples R China
[3] Hong Kong Baptist Univ, Dept Chem, Kowloon Tong, Hong Kong, Peoples R China
基金
中国国家自然科学基金;
关键词
LIQUID-CHROMATOGRAPHY; CAPILLARY-ELECTROPHORESIS; SENSITIVE DETERMINATION; TOTAL IRON; INJECTION; IMMUNOASSAY; SEAWATER; MECHANISM; OXIDATION; FE(II);
D O I
10.1039/b812750f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A droplet detector with high sensitivity and low consumption of reagents and sample for continuous flow chemiluminescence (CL) detection has been developed. The proposed system directly mixes sample with luminescence reagents to form a small droplet, thanks to the gravity force and surface tension of water, in front of a photomultiplier tube (PMT) where the CL reaction takes place. The PMT can then instantaneously record the maximum CL light signal emitted from the whole reaction process. This small droplet can replace the conventional spiral-shaped flow cell for flow injection CL analysis and the droplet detector can reduce interference from the reaction products and residual reagents. To evaluate the potential application of the proposed detector, the CL detection of Fe(II) was assessed under the alkaline luminol-hydrogen peroxide conditions. The CL intensity increased linearly with the concentration of Fe(II) ranging from 10.0 to 1000 nmol dm(-3) with a detection limit of 7.16 nmol dm(-3) (S/N = 3). The droplet detector allows highly sensitive, stable and fast CL detection of analytes, and thus is promising for application to other analytes having fast CL reaction rates.
引用
收藏
页码:354 / 360
页数:7
相关论文
共 50 条
  • [21] Luminol-hydrogen peroxide-horseradish peroxidase chemiluminescence intensification by kosmotrope ammonium sulfate
    Karatani, Hajime
    ANALYTICAL SCIENCES, 2022, 38 (03) : 613 - 621
  • [22] Chemiluminescence detection for Co (II) based on luminol-hydrogen peroxide reaction on a microfluidic chip
    Xueye Chen
    Fuqiang Chang
    Xing Wei
    Geng Li
    Journal of Food Science and Technology, 2015, 52 : 601 - 604
  • [23] Chemiluminescence determination of surfactant Triton X-100 in environmental water with luminol-hydrogen peroxide system
    Liu, Xiaoyu
    Li, Aifang
    Zhou, Baohui
    Qiu, Chaokun
    Ren, Hongmin
    CHEMISTRY CENTRAL JOURNAL, 2009, 3
  • [24] Calibration transfer in chemiluminescence analysis -: Application to chromium determination by luminol-hydrogen peroxide reaction
    Tortajada-Genaro, LA
    Campíns-Falcó, P
    Bosch-Reig, F
    ANALYTICA CHIMICA ACTA, 2001, 446 (1-2) : 385 - 392
  • [25] Lophine derivatives and analogues as new phenolic enhancers for the luminol-hydrogen peroxide-horseradish peroxidase chemiluminescence system
    Kuroda, N
    Shimoda, R
    Wada, M
    Nakashima, K
    ANALYTICA CHIMICA ACTA, 2000, 403 (1-2) : 131 - 136
  • [26] Determination of phenol by flow-injection with chemiluminescence detection based on the hemin-catalysed luminol-hydrogen peroxide reaction
    Liu, Wenwen
    Cao, Wei
    Liu, Weihua
    Du, Kang
    Gong, Pixue
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2012, 85 (01) : 283 - 287
  • [27] Chemiluminescence Property of the Luminol-Hydrogen Peroxide-Copper(II) System in the Presence of Surface-Carboxylated Microspheres
    Kazuhiko Tsukagoshi
    Masayuki Sumiyama
    Riichiro Nakajima
    Masamichi Nakayama
    Mizuo Maeda
    Analytical Sciences, 1998, 14 : 409 - 412
  • [28] Enhanced chemiluminescence of the luminol-hydrogen peroxide system by BSA-stabilized Au nanoclusters as a peroxidase mimic and its application
    Deng, Mao
    Xu, Shuangjiao
    Chen, Funan
    ANALYTICAL METHODS, 2014, 6 (09) : 3117 - 3123
  • [29] Flow-injection chemiluminescence determination of cloxacillin in water samples and pharmaceutical preparation by using CuO nanosheets-enhanced luminol-hydrogen peroxide system
    Khataee, Alireza
    Iranifam, Mortaza
    Fathinia, Mehrangiz
    Nikravesh, Mina
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2015, 134 : 210 - 217
  • [30] A flow-injection chemiluminescence method for the determination of some estrogens by enhancement of luminol-hydrogen peroxide-tetrasulfonated manganese phthalocyanine reaction
    Wang, Lun
    Yang, Ping
    Li, Yongxin
    Zhu, Changqin
    TALANTA, 2006, 70 (01) : 219 - 224