Confocal dual-frequency enhances the damaging effect of high-intensity focused ultrasound in tissue-mimicking phantom

被引:5
|
作者
Kuang, Sheng-Li [1 ]
Hu, Bing [1 ]
Jiang, Li-Xin [1 ]
He, Pei-Zhong [2 ]
Xia, Rong-Min [3 ]
Shou, Wen-De [3 ]
机构
[1] Shanghai Jiao Tong Univ, Dept Ultrasound Med, Peoples Hosp 6, Shanghai 200233, Peoples R China
[2] Xuzhou Normal Univ, Dept Phys, Xuzhou, Peoples R China
[3] Shanghai Jiao Tong Univ, Dept Biomed Engn, Shanghai 200233, Peoples R China
基金
中国国家自然科学基金;
关键词
HIFU; ultrasound; tissue-mimicking phantom; dual-frequency; lesion;
D O I
10.1080/13645700802274281
中图分类号
R61 [外科手术学];
学科分类号
摘要
The volume of the lesions created by conventional single-frequency high-intensity focused ultrasound (HIFU) is small, which leads to long treatment duration in patients who are undergoing tumor ablation. In this study, the lesions induced by confocal dual-frequency HIFU in an optically transparent tissue-mimic king phantom were investigated and compared with the lesions created by conventional single-frequency HIFU. The results show that using different exposure times resulted in lesions of different sizes in both dual-frequency and single-frequency HIFU modes at the same spatially averaged intensity level (ISAL=4900 W cm(2)), but the lesion dimensions made in dual-frequency mode were significantly larger than those made in single-frequency mode. Difference frequency acoustic fields that exist in the confocal region of dual-frequency HIFU may be the reason for the enlargement of the lesions' dimensions. The dual-frequency HIFU mode may represent a new technique to improve the ablation efficiency of HIFU. The total time for the ablation of a tumor can be reduced, thus requiring less therapy time and reducing possible patient complications.
引用
收藏
页码:285 / 291
页数:7
相关论文
共 50 条
  • [1] Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin
    Hee-Jin Kim
    Han Gu Kim
    Zhenlong Zheng
    Hyoun Jun Park
    Jeung Hyun Yoon
    Wook Oh
    Cheol Woo Lee
    Sung Bin Cho
    Lasers in Medical Science, 2015, 30 : 2251 - 2258
  • [2] Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin
    Kim, Hee-Jin
    Kim, Han Gu
    Zheng, Zhenlong
    Park, Hyoun Jun
    Yoon, Jeung Hyun
    Oh, Wook
    Lee, Cheol Woo
    Cho, Sung Bin
    LASERS IN MEDICAL SCIENCE, 2015, 30 (09) : 2251 - 2258
  • [3] Development and Characterization of a Tissue-Mimicking Material for High-Intensity Focused Ultrasound
    King, Randy L.
    Liu, Yunbo
    Maruvada, Subha
    Herman, Bruce A.
    Wear, Keith A.
    Harris, Gerald R.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (07) : 1397 - 1405
  • [4] Boiling activity induced by high intensity focused ultrasound in a tissue-mimicking phantom
    Kang, Hwi Suk
    Lee, Kang Il
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2020, 59 (02)
  • [5] AN IMPROVED TISSUE-MIMICKING POLYACRYLAMIDE HYDROGEL PHANTOM FOR VISUALIZING THERMAL LESIONS WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    Guntur, Sitaramanjaneya Reddy
    Choi, Min Joo
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (11): : 2680 - 2691
  • [6] High-Intensity Focused Ultrasound Ablation by the Dual-Frequency Excitation
    Law, Siu Kwun Berlinda
    Zhou, Yufeng
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2019, 66 (01) : 18 - 25
  • [7] Thrombolysis enhanced by dual-frequency high-intensity focused ultrasound
    Guo, Sijia
    Suo, Dingjie
    Jing, Yun
    Jiang, Xiaoning
    Frank, Jonathan
    Lin, Weili
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 2173 - 2176
  • [8] SIMULTANEOUS EVALUATION OF THERMAL AND NON-THERMAL EFFECTS OF HIGH-INTENSITY FOCUSED ULTRASOUND ON A TISSUE-MIMICKING PHANTOM
    Kim, Yong Tae
    Ma, Donghee
    Sim, Jai Kyoung
    Kim, Se-Hwa
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2018, 44 (08): : 1799 - 1809
  • [9] The affection on the tissue lesions of difference frequency in dual-frequency high-intensity focused ultrasound (HIFU)
    He, PZ
    Xia, RM
    Duan, SM
    Shou, WD
    Qian, DC
    ULTRASONICS SONOCHEMISTRY, 2006, 13 (04) : 339 - 344
  • [10] Development of a tissue-mimicking phantom for evaluating the focusing performance of high intensity focused ultrasound
    Jing Zongyu
    Li Faqi
    Zou Hangzhong
    Wang Zhibiao
    THERAPEUTIC ULTRASOUND, 2006, 829 : 648 - +