Development of a tissue-mimicking phantom for evaluating the focusing performance of high intensity focused ultrasound

被引:0
|
作者
Jing Zongyu [1 ]
Li Faqi [1 ]
Zou Hangzhong [1 ]
Wang Zhibiao [1 ]
机构
[1] Chongqing Univ Med Sci, Dept Biomed Engn, Inst Ultrasound Engn Med, Chongqing 400016, Peoples R China
来源
THERAPEUTIC ULTRASOUND | 2006年 / 829卷
关键词
ultrasound;
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Obiectives: To develop a tissue mimicking phantom which can be used to evaluate the focusing performance of the HIFU transducer, and the phantom should has the same acoustic characteristic and thermotics characteristic as the biological tissue. Materials and methods: The tissue mimicking phantom was made from water, gelatin, fresh biologic tissue Its ultrasonic parameters (attenuation coefficient) of the phantom was measured by the method of radiation pressure. and thermotics parameters of the phantom, including thermal conductivity, specific heat/fusion point et al were tested under the Measurement meter. The HIFU biological effect of the phantom was evaluated under the Model JC focused ultrasound tumor therapeutic system, developed and produced by Chongqing HIFU Technology Co. Ltd (working frequency: 0.7MHz; acoustic power: 200W; focal distance: 135mm; Acoustic focal region: 3x3x25 cubic mm). Results: The self-made phantom is sable, has smooth and glossy appearance, well-distributed construction, and good elasticity. We measured the followed values for acoustic and thermal properties: density 1049 +/- 2 kg/m(3); attenuation 0.532 +/- 0.017 dB/cm (0.8 MHz), 0.612 +/- 0.021 dB/cm (1.0 MHz); thermal conductivity 0.76 +/- 0.08 W/m/-degrees C; specific heat 3653 143 J/kg-degrees C; fusion point154 +/- 8 degrees C. The BFR induced in the phantom after HIFU exposure was stable in its size and appearance. Conclusion: We produced and improved one tissue mimicking phantom successfully which had semblable ultrasound and thermphysical properties like the soft tissue, and can replace the bovine liver to investigate the HIFU biological effect and to detect the focusing performance of the HIFU energy transducer.
引用
收藏
页码:648 / +
页数:2
相关论文
共 50 条
  • [1] Boiling activity induced by high intensity focused ultrasound in a tissue-mimicking phantom
    Kang, Hwi Suk
    Lee, Kang Il
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2020, 59 (02)
  • [2] Prediction and Measurement of Temperature Rise Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Kang Il Lee
    [J]. Journal of the Korean Physical Society, 2018, 72 : 1313 - 1319
  • [3] Development and Characterization of a Tissue-Mimicking Material for High-Intensity Focused Ultrasound
    King, Randy L.
    Liu, Yunbo
    Maruvada, Subha
    Herman, Bruce A.
    Wear, Keith A.
    Harris, Gerald R.
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (07) : 1397 - 1405
  • [4] Prediction and Measurement of Temperature Rise Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Lee, Kang Il
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (11) : 1313 - 1319
  • [5] Prediction and Measurement of the Size of Thermal Lesion Induced by High Intensity Focused Ultrasound in a Tissue-Mimicking Phantom
    Lee, Kang Il
    Choi, Min Joo
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2009, 48 (02)
  • [6] Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin
    Hee-Jin Kim
    Han Gu Kim
    Zhenlong Zheng
    Hyoun Jun Park
    Jeung Hyun Yoon
    Wook Oh
    Cheol Woo Lee
    Sung Bin Cho
    [J]. Lasers in Medical Science, 2015, 30 : 2251 - 2258
  • [7] Coagulation and ablation patterns of high-intensity focused ultrasound on a tissue-mimicking phantom and cadaveric skin
    Kim, Hee-Jin
    Kim, Han Gu
    Zheng, Zhenlong
    Park, Hyoun Jun
    Yoon, Jeung Hyun
    Oh, Wook
    Lee, Cheol Woo
    Cho, Sung Bin
    [J]. LASERS IN MEDICAL SCIENCE, 2015, 30 (09) : 2251 - 2258
  • [8] AN IMPROVED TISSUE-MIMICKING POLYACRYLAMIDE HYDROGEL PHANTOM FOR VISUALIZING THERMAL LESIONS WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    Guntur, Sitaramanjaneya Reddy
    Choi, Min Joo
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (11): : 2680 - 2691
  • [9] SIMULTANEOUS EVALUATION OF THERMAL AND NON-THERMAL EFFECTS OF HIGH-INTENSITY FOCUSED ULTRASOUND ON A TISSUE-MIMICKING PHANTOM
    Kim, Yong Tae
    Ma, Donghee
    Sim, Jai Kyoung
    Kim, Se-Hwa
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2018, 44 (08): : 1799 - 1809
  • [10] Confocal dual-frequency enhances the damaging effect of high-intensity focused ultrasound in tissue-mimicking phantom
    Kuang, Sheng-Li
    Hu, Bing
    Jiang, Li-Xin
    He, Pei-Zhong
    Xia, Rong-Min
    Shou, Wen-De
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2008, 17 (05) : 285 - 291