Dose reduction in CT urography and vasculature phantom studies using model-based iterative reconstruction

被引:2
|
作者
Page, Leland [1 ]
Wei, Wei [2 ]
Kundra, Vikas [3 ,4 ]
Rong, X. John [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, 1400 Pressler St,Unit 1472, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Biostat, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Diagnost Radiol, Houston, TX 77030 USA
[4] Univ Texas MD Anderson Canc Ctr, Dept Canc Syst Imaging, Houston, TX 77030 USA
来源
关键词
CT; urography; vasculature; iterative reconstruction; dose; FILTERED BACK-PROJECTION; IMAGE QUALITY; OPTIMIZATION;
D O I
10.1120/jacmp.v17i6.6184
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate the feasibility of radiation dose reduction using model-based iterative reconstruction (MBIR) for evaluating the ureters and vasculature in a phantom, a tissue-equivalent CT dose phantom was scanned using a 64-channel CT scanner. Tubes of varying diameters filled with different dilutions of a contrast agent, simulating ureters or vessels, were inserted into the center of the phantom. Each combination was scanned using an existing renal protocol at 140 kVp or 120 kVp, yielding a display volumetric CT dose index (CTDIvol) of 24 mGy. The scans were repeated using reduced scan techniques to achieve lower radiation doses down to 0.8 mGy. The images were reconstructed using filtered back-projection (FBP) and model-based iterative reconstruction (MBIR). The noise and contrast-to-noise ratio (CNR) was measured for each contrast object. Comparisons between the two reconstruction methods at different dose levels were evaluated using a factorial design. At each CTDIvol the measured image noise was lower using MBIR compared to FBP (p < 0.0001). At low doses, the percent change in measured image noise between FBP and MBIR was larger. For the 12 mm object simulating a ureter or large vessel with an HU of 600, the measured CNR using MBIR at a CTDIvol of 1.7 mGy was greater than the CNR of FBP at a CTIDvol of 24 mGy (p < 0.0001). For the 5 mm object simulating a medium-sized vessel with a HU of 250, the measured CNR using MBIR at a CTDIvol of 1.7 mGy was equivalent to that of FBP at a CTDIvol of 24 mGy. For the 2 mm, 100 HU object simulating a small vessel, the measured CNR using MBIR at a CTDIvol of 1.7 mGy was equivalent to that of FBP at a CTDIvol of 24 mGy. Low-dose (3.6 mGy) CT imaging of vasculature and ureter phantoms using MBIR results in similar noise and CNR compared to FBP at approximately one-sixth the dose. This suggests that, using MBIR, a one milliSievert exam of the ureters and vasculature may be clinically possible whilst still maintaining adequate image quality
引用
收藏
页码:334 / 342
页数:9
相关论文
共 50 条
  • [1] Radiation Dose Reduction in Paranasal Sinus CT Using Model-Based Iterative Reconstruction
    Hoxworth, J. M.
    Lal, D.
    Fletcher, G. P.
    Patel, A. C.
    He, M.
    Paden, R. G.
    Hara, A. K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (04) : 644 - 649
  • [2] Limiting CT radiation dose in children with craniosynostosis: phantom study using model-based iterative reconstruction
    Kaasalainen, Touko
    Palmu, Kirsi
    Lampinen, Anniina
    Reijonen, Vappu
    Leikola, Junnu
    Kivisaari, Riku
    Kortesniemi, Mika
    PEDIATRIC RADIOLOGY, 2015, 45 (10) : 1544 - 1553
  • [3] Limiting CT radiation dose in children with craniosynostosis: phantom study using model-based iterative reconstruction
    Touko Kaasalainen
    Kirsi Palmu
    Anniina Lampinen
    Vappu Reijonen
    Junnu Leikola
    Riku Kivisaari
    Mika Kortesniemi
    Pediatric Radiology, 2015, 45 : 1544 - 1553
  • [4] Model-based iterative reconstruction for reduction of radiation dose in abdominopelvic CT: comparison to adaptive statistical iterative reconstruction
    Yasaka, Koichiro
    Katsura, Masaki
    Akahane, Masaaki
    Sato, Jiro
    Matsuda, Izuru
    Ohtomo, Kuni
    SPRINGERPLUS, 2013, 2 : 1 - 9
  • [5] Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction
    Wellenberg, R. H. H.
    Boomsma, M. F.
    van Osch, J. A. C.
    Vlassenbroek, A.
    Milles, J.
    Edens, M. A.
    Streekstra, G. J.
    Slump, C. H.
    Maas, M.
    SKELETAL RADIOLOGY, 2017, 46 (05) : 623 - 632
  • [6] Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction
    R. H. H. Wellenberg
    M. F. Boomsma
    J. A. C. van Osch
    A. Vlassenbroek
    J. Milles
    M. A. Edens
    G. J. Streekstra
    C. H. Slump
    M. Maas
    Skeletal Radiology, 2017, 46 : 623 - 632
  • [7] Investigation of Renal CT Dose Reduction by Using Model Based Iterative Reconstruction
    Page, L.
    Kundra, V.
    Rong, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction
    Kataria, Bharti
    Althen, Jonas Nilsson
    Smedby, Orjan
    Persson, Anders
    Sokjer, Hannibal
    Sandborg, Michael
    EUROPEAN RADIOLOGY, 2018, 28 (06) : 2464 - 2473
  • [9] Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction
    Bharti Kataria
    Jonas Nilsson Althén
    Örjan Smedby
    Anders Persson
    Hannibal Sökjer
    Michael Sandborg
    European Radiology, 2018, 28 : 2464 - 2473
  • [10] Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique
    Masaki Katsura
    Izuru Matsuda
    Masaaki Akahane
    Jiro Sato
    Hiroyuki Akai
    Koichiro Yasaka
    Akira Kunimatsu
    Kuni Ohtomo
    European Radiology, 2012, 22 : 1613 - 1623