Dose reduction in CT urography and vasculature phantom studies using model-based iterative reconstruction

被引:2
|
作者
Page, Leland [1 ]
Wei, Wei [2 ]
Kundra, Vikas [3 ,4 ]
Rong, X. John [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, 1400 Pressler St,Unit 1472, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Biostat, Houston, TX 77030 USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Diagnost Radiol, Houston, TX 77030 USA
[4] Univ Texas MD Anderson Canc Ctr, Dept Canc Syst Imaging, Houston, TX 77030 USA
来源
关键词
CT; urography; vasculature; iterative reconstruction; dose; FILTERED BACK-PROJECTION; IMAGE QUALITY; OPTIMIZATION;
D O I
10.1120/jacmp.v17i6.6184
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate the feasibility of radiation dose reduction using model-based iterative reconstruction (MBIR) for evaluating the ureters and vasculature in a phantom, a tissue-equivalent CT dose phantom was scanned using a 64-channel CT scanner. Tubes of varying diameters filled with different dilutions of a contrast agent, simulating ureters or vessels, were inserted into the center of the phantom. Each combination was scanned using an existing renal protocol at 140 kVp or 120 kVp, yielding a display volumetric CT dose index (CTDIvol) of 24 mGy. The scans were repeated using reduced scan techniques to achieve lower radiation doses down to 0.8 mGy. The images were reconstructed using filtered back-projection (FBP) and model-based iterative reconstruction (MBIR). The noise and contrast-to-noise ratio (CNR) was measured for each contrast object. Comparisons between the two reconstruction methods at different dose levels were evaluated using a factorial design. At each CTDIvol the measured image noise was lower using MBIR compared to FBP (p < 0.0001). At low doses, the percent change in measured image noise between FBP and MBIR was larger. For the 12 mm object simulating a ureter or large vessel with an HU of 600, the measured CNR using MBIR at a CTDIvol of 1.7 mGy was greater than the CNR of FBP at a CTIDvol of 24 mGy (p < 0.0001). For the 5 mm object simulating a medium-sized vessel with a HU of 250, the measured CNR using MBIR at a CTDIvol of 1.7 mGy was equivalent to that of FBP at a CTDIvol of 24 mGy. For the 2 mm, 100 HU object simulating a small vessel, the measured CNR using MBIR at a CTDIvol of 1.7 mGy was equivalent to that of FBP at a CTDIvol of 24 mGy. Low-dose (3.6 mGy) CT imaging of vasculature and ureter phantoms using MBIR results in similar noise and CNR compared to FBP at approximately one-sixth the dose. This suggests that, using MBIR, a one milliSievert exam of the ureters and vasculature may be clinically possible whilst still maintaining adequate image quality
引用
收藏
页码:334 / 342
页数:9
相关论文
共 50 条
  • [41] Hybrid and Model-Based Iterative Reconstruction Techniques for Pediatric CT
    den Harder, Annemarie M.
    Willemink, Martin J.
    Budde, Ricardo P. J.
    Schilham, Arnold M. R.
    Leiner, Tim
    de Jong, Pim A.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (03) : 645 - 653
  • [42] Renovascular CT: comparison between adaptive statistical iterative reconstruction and model-based iterative reconstruction
    Noda, Y.
    Goshima, S.
    Koyasu, H.
    Shigeyama, S.
    Miyoshi, T.
    Kawada, H.
    Kawai, N.
    Matsuo, M.
    CLINICAL RADIOLOGY, 2017, 72 (10) : 901.e13 - 901.e19
  • [43] CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction
    Yasutaka Ichikawa
    Kakuya Kitagawa
    Naoki Nagasawa
    Shuichi Murashima
    Hajime Sakuma
    BMC Medical Imaging, 13
  • [44] CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction
    Ichikawa, Yasutaka
    Kitagawa, Kakuya
    Nagasawa, Naoki
    Murashima, Shuichi
    Sakuma, Hajime
    BMC MEDICAL IMAGING, 2013, 13
  • [45] Image quality improvement using model-based iterative reconstruction in low dose chest CT for children with necrotizing pneumonia
    Sun, Jihang
    Yu, Tong
    Liu, Jinrong
    Duan, Xiaomin
    Hu, Di
    Liu, Yong
    Peng, Yun
    BMC MEDICAL IMAGING, 2017, 17
  • [46] Ultra low dose CT for left atrium and pulmonary veins imaging using VEO (model-based iterative reconstruction)
    Annoni, A.
    Formenti, A.
    Andreini, D.
    Pontone, G.
    Petulla', M.
    Ballerini, G.
    Nobili, E.
    Baggiano, A.
    Mushtaq, S.
    Pepi, M.
    EUROPEAN HEART JOURNAL, 2013, 34 : 845 - 845
  • [47] Image quality improvement using model-based iterative reconstruction in low dose chest CT for children with necrotizing pneumonia
    Jihang Sun
    Tong Yu
    Jinrong Liu
    Xiaomin Duan
    Di Hu
    Yong liu
    Yun Peng
    BMC Medical Imaging, 17
  • [48] Reducing the radiation dose for computed tomography colonography using model-based iterative reconstruction
    Millerd, Patrick J.
    Paden, Robert G.
    Lund, Jeffrey T.
    Hara, Amy K.
    Stiles, Wendy L.
    He, Miao
    Wu, Qing
    Johnson, C. Daniel
    ABDOMINAL IMAGING, 2015, 40 (05): : 1183 - 1189
  • [49] Reducing the radiation dose for computed tomography colonography using model-based iterative reconstruction
    Patrick J. Millerd
    Robert G. Paden
    Jeffrey T. Lund
    Amy K. Hara
    Wendy L. Stiles
    Miao He
    Qing Wu
    C. Daniel Johnson
    Abdominal Imaging, 2015, 40 : 1183 - 1189
  • [50] Improved visualization of the coronary arteries using model-based iterative reconstruction for cardiac CT
    Karolyi, M.
    Szilveszter, B.
    Jermendy, A. L.
    Panajotu, A.
    Bartykowszki, A.
    Celeng, C.
    Raaijmakers, R.
    Giepmans, W.
    Merkely, B.
    Maurovich-Horvat, P.
    EUROPEAN HEART JOURNAL, 2015, 36 : 770 - 770