Optimization of phase-contrast MRI for the estimation of global cerebral blood flow of mice at 11.7T

被引:14
|
作者
Wei, Zhiliang [1 ,2 ]
Chen, Lin [1 ,2 ]
Lin, Zixuan [3 ]
Jiang, Dengrong [3 ]
Xu, Jiadi [1 ,2 ]
Liu, Peiying [1 ]
van Zijl, Peter C. M. [1 ,2 ]
Lu, Hanzhang [1 ,2 ,3 ]
机构
[1] Johns Hopkins Univ, Sch Med, Russell H Morgan Dept Radiol & Radiol Sci, 600 North Wolfe St,Pk 322, Baltimore, MD 21287 USA
[2] Kennedy Krieger Res Inst, FM Kirby Res Ctr Funct Brain Imaging, Baltimore, MD USA
[3] Johns Hopkins Univ, Sch Med, Dept Biomed Engn, Baltimore, MD 21205 USA
关键词
cerebral blood flow; internal carotid artery; maximum blood velocity; partial volume effect; spatial folding; spatial resolution; vertebral artery; QUANTITATIVE-EVALUATION; OXYGEN-METABOLISM; BRAIN; MOUSE; PET; QUANTIFICATION; RATS;
D O I
10.1002/mrm.27592
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To optimize phase-contrast (PC) MRI for the measurement of global cerebral blood flow (CBF) in the mouse at 11.7T. Methods: We determined proper velocity encoding (VENC) for internal carotid arteries (ICAs) and vertebral arteries (VAs). Next, we optimized spatial resolution of the sequence. To shorten scan time without compromising data quality, we further optimized repetition time and developed a reduced field-of-view (FOV) scheme for ICA and VA PC MRI. Whole-brain volume was determined with T-2-weighted image to obtain unit-volume CBF. Results: Peak flow velocities were 13.8 +/- 1.7, 14.4 +/- 0.6, 6.5 +/- 1.7, and 6.7 +/- 1.3 cm/s for left ICA, right ICA, left VA, and right VA, respectively. Thus, VENC values of 20 and 10 cm/s were chosen for ICA and VA PC MRI, respectively. An in-plane spatial resolution of 50 x 50 mu m(2) was found to provide a reasonable trade-off between reducing partial-volume effects and maintaining signal-to-noise ratio. Because of the fact that saturated spins in the imaging slice are rapidly replaced by fresh spins, TR of the sequence can be decreased to as short as 15 ms without reducing signal intensity, thereby substantially lowering scan time. Moreover, reduced FOV along the phase-encoding direction was able to shorten scan time by 33.3% while maintaining measurement accuracy. With these optimizations, it took 96 seconds to evaluate CBF with a test-retest variability of approximately 5% and an interrater correlation of > 0.95. Global unit-volume CBF was found to be 279.5 +/- 11.1 mL of blood/100 ml of tissue/min. Conclusion: We have optimized PC MRI for noninvasive quantification of blood flow in mice at 11.7T.
引用
收藏
页码:2566 / 2575
页数:10
相关论文
共 50 条
  • [21] Comparison of Global Cerebral Blood Flow Measured by Phase-Contrast Mapping MRI with 15O-H2O Positron Emission Tomography
    Vestergaard, Mark Bitsch
    Lindberg, Ulrich
    Aachmann-Andersen, Niels Jacob
    Lisbjerg, Kristian
    Christensen, Soren Just
    Rasmussen, Peter
    Olsen, Niels Vidiendal
    Law, Ian
    Larsson, Henrik Bo Wiberg
    Henriksen, Otto Molby
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2017, 45 (03) : 692 - 699
  • [22] Assessment of the accuracy and reproducibility of left ventricular mass and volumes measurements in mice using an 11.7T MRI machine
    Vanhoutte, L.
    Esfahani, H.
    Feron, O.
    Balligand, J. L.
    Gallez, B.
    Moniotte, S.
    CARDIOVASCULAR RESEARCH, 2014, 103
  • [23] Longitudinal neuroplasticity after ankle sprain in mice: A voxel-based morphometry study on 11.7T MRI
    Xue, Xiao'ao
    Zhang, Yuwen
    Tao, Weichu
    Wei, Lei
    Li, Qianru
    Ma, Tengjia
    Xu, Xiaoyun
    Wang, Yiran
    Gu, Xicheng
    Xu, Zhangran
    Wang, He
    Hua, Yinghui
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2023, 41 (06) : 1291 - 1298
  • [24] Doppler-Assessed Cerebral Blood Flow Velocity in the Neonate as Estimator of Global Cerebral Blood Volume Flow Measured Using Phase-Contrast Magnetic Resonance Angiography
    Benders, Manon J. N. L.
    Hendrikse, Jeroen
    de Vries, Linda
    Groenendaal, Floris
    van Bel, Frank
    NEONATOLOGY, 2013, 103 (01) : 21 - 26
  • [25] Rapid measurement of renal artery blood flow with ungated spiral phase-contrast MRI
    Park, JB
    Santos, JM
    Hargreaves, BA
    Nayak, KS
    Sommer, G
    Hu, BS
    Nishimura, DG
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 21 (05) : 590 - 595
  • [26] Blood flow quantification in adults by phase-contrast MRI combined with SENSE -: a validation study
    Beerbaum, P
    Körperich, H
    Gieseke, J
    Barth, P
    Peuster, M
    Meyer, H
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2005, 7 (02) : 361 - 369
  • [27] Maximum likelihood estimation of cerebral blood flow in dynamic susceptibility contrast MRI
    Vonken, EPA
    Beekman, FJ
    Bakker, CJG
    Viergever, MA
    MAGNETIC RESONANCE IN MEDICINE, 1999, 41 (02) : 343 - 350
  • [28] Simultaneous Phase-Contrast MRI and PET for Noninvasive Quantification of Cerebral Blood Flow and Reactivity in Healthy Subjects and Patients With Cerebrovascular Disease
    Ishii, Yosuke
    Thamm, Thoralf
    Guo, Jia
    Khalighi, Mohammad Mehdi
    Wardak, Mirwais
    Holley, Dawn
    Gandhi, Harsh
    Park, Jun Hyung
    Shen, Bin
    Steinberg, Gary K.
    Chin, Frederick T.
    Zaharchuk, Greg
    Fan, Audrey Peiwen
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2020, 51 (01) : 183 - 194
  • [29] Inter- and Intra-Rater Reliability of Individual Cerebral Blood Flow Measured by Quantitative Vessel-Flow Phase-Contrast MRI
    Chang, Kwang-Hwa
    Lee, Yuan-Hao
    Chen, Chia-Yuen
    Lin, Ming-Fang
    Lin, Ying Chin
    Chen, Jyh-Horng
    Chan, Wing P.
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (10) : 1 - 11
  • [30] Applications of phase-contrast flow and velocity imaging in cardiovascular MRI
    Peter D. Gatehouse
    Jennifer Keegan
    Lindsey A. Crowe
    Sharmeen Masood
    Raad H. Mohiaddin
    Karl-Friedrich Kreitner
    David N. Firmin
    European Radiology, 2005, 15 : 2172 - 2184