Impact of electrode tip shape on catheter performance in cardiac radiofrequency ablation

被引:4
|
作者
Petras, Argyrios [1 ]
Moreno Weidmann, Zoraida [2 ]
Echeverria Ferrero, Marina [3 ]
Leoni, Massimiliano
Guerra, Jose M. [2 ]
Gerardo-Giorda, Luca [1 ,4 ]
机构
[1] Austrian Acad Sci, RICAM, Linz, Austria
[2] Univ Autonoma Barcelona, Hosp Santa Creu & St Pau, Dept Cardiol, IIB ST PAU,CIBERCV, Barcelona, Spain
[3] BCAM Basque Ctr Appl Math, Bilbao, Spain
[4] Johannes Kepler Univ Linz, Inst Math Methods Med & Data Based Modelling, Linz, Austria
来源
HEART RHYTHM O2 | 2022年 / 3卷 / 06期
关键词
Cardiac ablation; Computational modeling and simulation; Electrode tip; Lesion science; Open irrigation; Radiofrequency; Steam pop; CONTACT FORCE; LESION SIZE; TEMPERATURE;
D O I
10.1016/j.hroo.2022.07.014
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The role of catheter tip shape on the safety and efficacy of radiofrequency (RF) ablation has been overlooked, although differences have been observed in clinical and research fields. OBJECTIVE The purpose of this study was to analyze the role of electrode tip shape in RF ablation using a computational model. METHODS We simulated 108 RF ablations through a realistic 3-dimensional computational model considering 2 clinically used, open-irrigated catheters (spherical and cylindrical tip), varying contact force (CF), blood flow, and irrigation. Lesions are defined by the 50 degrees C isotherm contour and evaluated by means of width, depth, depth at maximum width, and volume. Ablations are deemed as safe, critical (tissue temperature.90 degrees C), and pop (tissue temperature.100 degrees C). RESULTS Tissue-electrode contact is less for the spherical tip at low CF but the relationship is inverted at high CF. At low CF, the cylindrical tip generates deeper and wider lesions and a 4-fold larger volume. With increasing CF, the lesions generated by the spherical tip become comparable to those generated by the cylindrical tip. The 2 tips feature different safety profiles: CF and power are the main determinants of pops for the spherical tip; power is the main factor for the cylindrical tip; and CF has a marginal effect. The cylindrical tip is more prone to pop generation at higher powers. Saline irrigation and blood flow effect do not depend on tip shape. CONCLUSION Tip shape determines the performance of ablation catheters and has a major impact on their safety profile. The cylindrical tip shows more predictable behavior in a wide range of CF values.
引用
收藏
页码:699 / 705
页数:7
相关论文
共 50 条
  • [41] A rare complication of cardiac radiofrequency catheter ablation: Esophageal irritation
    Celebi, Ozlem Ozcan
    Canbay, Alper
    Diker, Erdem
    TURK KARDIYOLOJI DERNEGI ARSIVI-ARCHIVES OF THE TURKISH SOCIETY OF CARDIOLOGY, 2015, 43 (04): : 389 - 391
  • [42] Ventricular Tachycardia in Cardiac Sarcoidosis Controlled by Radiofrequency Catheter Ablation
    Bando, Mika
    Soeki, Takeshi
    Niki, Toshiyuki
    Kusunose, Kenya
    Tomita, Noriko
    Yamaguchi, Koji
    Koshiba, Kunihiko
    Taketani, Yoshio
    Iwase, Takashi
    Yamada, Hirotsugu
    Wakatsuki, Tetsuzo
    Akaike, Masashi
    Sata, Masataka
    INTERNAL MEDICINE, 2011, 50 (11) : 1201 - 1206
  • [43] The present role of radiofrequency catheter ablation in the management of cardiac arrhythmias
    Kay, GN
    Plumb, VJ
    AMERICAN JOURNAL OF MEDICINE, 1996, 100 (03): : 344 - 356
  • [44] Effect of Tissue Elasticity in Cardiac Radiofrequency Catheter Ablation Models
    Petras, Argyrios
    Leoni, Massimiliano
    Guerra, Jose M.
    Jansson, Johan
    Gerardo-Giorda, Luca
    2018 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), 2018, 45
  • [45] Role of Catheter Contact Force on Biophysical Properties of the Ablation Lesion Formation in Radiofrequency Catheter Cardiac Ablation
    Masnok, Kriengsak
    Watanabe, Nobuo
    2021 IEEE REGION 10 SYMPOSIUM (TENSYMP), 2021,
  • [46] Effects of radiofrequency catheter ablation on quality of life and cardiac performance in patients with typical atrial flutter
    Lee, SH
    Chen, SA
    Cheng, JJ
    Roc, TT
    Ding, YA
    Chang, MS
    CIRCULATION, 1998, 98 (17) : 181 - 181
  • [47] Impact of paroxysmal atrial fibrillation ablation on cardiac sympathetic nervous system: a prospective randomized comparative study with cryoballoon catheter or radiofrequency ablation catheter
    Kawasaki, M.
    Yamada, T.
    Morita, T.
    Furukawa, Y.
    Tamaki, S.
    Iwasaki, Y.
    Kikuchi, A.
    Kawai, T.
    Seo, M.
    Ikeda, I.
    Fukuhara, E.
    Abe, M.
    Nakamura, J.
    Fukunami, M.
    EUROPEAN HEART JOURNAL, 2018, 39 : 1415 - 1415
  • [48] Radiofrequency ablation lesions produced by a cool tip catheter in human infarction.
    Stevenson, WG
    Mitchell, R
    Friedman, PL
    Stewart, ST
    Seifert, MJ
    CIRCULATION, 1996, 94 (08) : 124 - 124
  • [49] Probabilistic guidance for catheter tip motion in cardiac ablation procedures
    Constantinescu, Mihaela A. M.
    Lee, Su-Lin
    Ernst, Sabine
    Hemakom, Apit
    Mandic, Danilo
    Yang, Guang-Zhong
    MEDICAL IMAGE ANALYSIS, 2018, 47 : 1 - 14
  • [50] RADIOFREQUENCY CATHETER ABLATION - THE EFFECT OF ELECTRODE SIZE ON LESION VOLUME INVIVO
    LANGBERG, JJ
    LEE, MA
    CHIN, MC
    ROSENQVIST, M
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (10): : 1242 - 1248