Immediate and day-to-day reproducibility of the signal-averaged electrocardiogram in patients with coronary artery disease

被引:1
|
作者
Paquay, JL
Zimmermann, M
Mermillod, B
Adamec, R
机构
[1] UNIV HOSP GENEVA,CTR CARDIOL,CH-1211 GENEVA 14,SWITZERLAND
[2] UNIV HOSP GENEVA,CTR INFORMAT HOSP,CH-1211 GENEVA 14,SWITZERLAND
来源
关键词
signal-averaging; late potentials;
D O I
10.1111/j.1540-8159.1996.tb06515.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to prospectively evaluate the immediate (5 minutes) and short-term (1 day) reproducibility of the signal-averaged electrocardiogram (SAECG) in patients with coronary artery disease. A total of 59 consecutive patients with coronary artery disease (50 male, 9 female, mean age 59 +/- 13 years) were included, Analysis was performed using a commercially available system with high-pass filters of 25, 40, and 80 Hz without any change in electrode position. The following time-domain parameters were measured: (1) total filtered QRS duration; (2) duration of high-frequency low-amplitude signals (HFLA); and (3) root-mean-square voltage of She terminal 40 ms of the QRS complex (RMS40). Correlation between pairs of measurements was excellent, ranging from 0.91-0.99 for QRS and HFLA duration, and from 0.83-0.98 for RMS40 value. However, the range of differences was wide for all parameters, especially at 25 and 80 Hz, revealing marked individual variations in selected cases. The immediate reproducibility of an initially normal SAECG was 95% at 25 Hz, 100% at 40 Hz, and 81% at 80 Iii; figures for an initially abnormal SAECG were 89%, 91%, and 93%, respectively, and the coefficient of agreement kappa was highest at 40 Hz than at 25 or 80 Hz. Discordant results were most frequently observed in patients with borderline values, or in cases with nonsimilar residual noise levels. In conclusion the immediate and day-to-day reproducibility of the SAECG is excellent in patients with chronic stable coronary artery disease. However, individual variations are observed in selected cases, leading to modification in interpretation of the result. For this reason, individual reproducibility should be considered before evaluating the effect of any therapeutic intervention on the SAECG, especially in patients with borderline values of the quantitative parameters of the SAECG.
引用
收藏
页码:443 / 454
页数:12
相关论文
共 50 条
  • [1] REPRODUCIBILITY OF THE SIGNAL-AVERAGED ELECTROCARDIOGRAM
    ENGEL, TR
    PIERCE, DL
    PATIL, KD
    AMERICAN HEART JOURNAL, 1991, 122 (06) : 1652 - 1660
  • [2] THE SIGNAL-AVERAGED ELECTROCARDIOGRAM IN PREDICTING CORONARY-ARTERY DISEASE
    SOLOMON, AJ
    TRACY, CM
    AMERICAN HEART JOURNAL, 1991, 122 (05) : 1334 - 1339
  • [3] Effect of dobutamine on signal-averaged electrocardiogram of patients with suspected coronary artery disease
    Khan, AH
    Herman, SD
    Ahlberg, A
    Heller, GV
    AMERICAN JOURNAL OF NONINVASIVE CARDIOLOGY, 1994, 8 (06): : 309 - 312
  • [4] Reproducibility of the signal-averaged electrocardiogram.
    Copie, X
    Blankoff, I
    Fei, L
    Murgatroyd, FD
    Hnatkova, K
    Malik, M
    Camm, AJ
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1996, 89 (03): : 325 - 330
  • [5] REPRODUCIBILITY OF DATA OF THE SIGNAL-AVERAGED ELECTROCARDIOGRAM
    GOEDELMEINEN, L
    HOFMANN, M
    SCHMIDT, G
    HOGLSPERGER, H
    JAHNS, G
    BAEDEKER, W
    BLOMER, H
    CIRCULATION, 1987, 76 (04) : 31 - 31
  • [6] EFFECTS OF EXERCISE ON THE SIGNAL-AVERAGED ELECTROCARDIOGRAM IN CORONARY-ARTERY DISEASE
    CAREF, EB
    GOLDBERG, N
    MENDELSON, L
    HANLEY, G
    OKEREKE, R
    STEIN, RA
    ELSHERIF, N
    AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (01): : 54 - 58
  • [7] SHORT-TERM AND LONG-TERM REPRODUCIBILITY OF THE SIGNAL-AVERAGED ELECTROCARDIOGRAM IN CORONARY-ARTERY DISEASE
    BORBOLA, J
    DENES, P
    AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (13): : 1123 - 1124
  • [8] Immediate and short-term reproducibility of the P wave signal-averaged electrocardiogram
    Ehlert, FA
    Zaman, N
    Steinberg, JS
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (06): : 1636 - 1645
  • [9] THE EFFECTS OF ATRIAL-PACING ON THE SIGNAL-AVERAGED ELECTROCARDIOGRAM IN PATIENTS WITH CORONARY-ARTERY DISEASE
    GREENSPON, AJ
    VOLOSIN, K
    AMERICAN HEART JOURNAL, 1990, 119 (01) : 29 - 34
  • [10] Effects of eplerenone on P-wave signal-averaged electrocardiogram in hypertensive patients with coronary artery disease
    Kurisu, Satoshi
    Watanabe, Noriaki
    Ikenaga, Hiroki
    Shimonaga, Takashi
    Higaki, Tadanao
    Iwasaki, Toshitaka
    Mitsuba, Naoya
    Ishibashi, Ken
    Dohi, Yoshihiro
    Kihara, Yasuki
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2014, 172 (01) : E180 - E181