A rapid analytical method for the quantitative determination of the sugar in acarbose fermentation by infrared spectroscopy and chemometrics

被引:10
|
作者
Zhou, Wanzhu [1 ]
Guo, Pengran [1 ]
Chen, Jianghan [1 ]
Lei, Yongqian [1 ]
机构
[1] Guangdong Acad Sci, Guangdong Prov Engn Res Ctr Online Monitoring Wat, Guangdong Prov Key Lab Emergency Test Dangerous C, Guangdong Inst Anal,China Natl Analyt Ctr Guangzh, Guangzhou 510070, Peoples R China
基金
中国国家自然科学基金;
关键词
Infrared spectroscopy; Acarbose; Glucose; Maltose; Chemometrics; REFLECTANCE SPECTROSCOPY; PHARMACEUTICAL PRODUCTS; SUSTAINED-RELEASE; IR SPECTROSCOPY; REGRESSION; CLASSIFICATION; FEASIBILITY; POLYMORPHS; MIXTURES; GENUINE;
D O I
10.1016/j.saa.2020.118571
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this article, a rapid analytical method for the quantitative determination of the glucose and maltose in the industrial acarbose fermentation was established by the combination application of infrared spectroscopy and chemometrics. The spectra of the 398 acarbose samples were collected by a portable infrared fast analyzer and the concentration of the glucose and maltose in the acarbose fermentation solution were determinate by high performance liquid chromatography (HPLC) as the referent database. Four spectral pretreatment methods, first derivative (FD), second derivative (SD), Savitzky-Golay (SG) convolution smoothing and mean center (MC) were employed to eliminate the optical interference from background and other noise information. The best re-sult was obtained with FD+SG(21, 3)+MC method. The effects of different principal component numbers (PCs) on the parameters were also optimized. Two models of PLS and MLR, were used to predict the concentration of the glucose and maltose. The FD+SG(21, 3)+MC method was chosen as best method, with 12 PCs for glucose and 11 for maltose as optimized parameters. The PLS model was significantly better than the MLR model. Furthermore, both the predicted values and the reference values of glucose and maltose models showed superior linear relationship within the calibration range. The absolute errors of the predicted values and their corresponding reference values of glucose and maltose in the PLS model were within +/- 0.14 and +/- 0.35 confidence intervals, respectively. The prediction correct rate was 98.3%, which indicated that the prediction results of model were excellent. (c) 2020 Elsevier B.V. All rights reserved.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] Rapid determination of polysaccharides and antioxidant activity of Poria cocos using near-infrared spectroscopy combined with chemometrics
    Yi, Yuan
    Hua, Haimin
    Sun, Xuefen
    Guan, Ying
    Chen, Chao
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2020, 240
  • [42] Rapid determination of ethyl pentanoate in liquor using Fourier transform near-infrared spectroscopy coupled with chemometrics
    Han, Sihai
    Zhang, Weiwei
    Li, Xuan
    Li, Peiyan
    Liu, Jianxue
    Luo, Denglin
    Xu, Baocheng
    SPECTROSCOPY LETTERS, 2016, 49 (07) : 464 - 468
  • [43] Rapid determination of total phenolic content of whole wheat flour using near-infrared spectroscopy and chemometrics
    Tian, Wenfei
    Chen, Gengjun
    Zhang, Guorong
    Wang, Donghai
    Tilley, Michael
    Li, Yonghui
    FOOD CHEMISTRY, 2021, 344
  • [44] Online quantitative substrate, product, and cell concentration in citric acid fermentation using near-infrared spectroscopy combined with chemometrics
    Jin, Sai
    Sun, Fuxin
    Hu, Zhijie
    Li, Youran
    Zhao, Zhonggai
    Du, Guocheng
    Shi, Guiyang
    Chen, Jian
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2023, 285
  • [45] Analytical Method Development for Analysis of Soybean Oil in the Mixture with Virgin Coconut Oil Using Infrared Spectroscopy and Chemometrics
    Rohman, Abdul
    Man, Yaakob B. Che
    AGRITECH, 2012, 32 (02): : 111 - 116
  • [46] Multivariate calibration and chemometrics for near infrared spectroscopy: which method?
    Dardenne, P
    Sinnaeve, G
    Baeten, V
    JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2000, 8 (04) : 229 - 237
  • [47] RAPID ANALYTICAL METHOD FOR THE QUANTITATIVE-DETERMINATION OF TRICHOTHECENE TOXINS IN FOOD AND FEEDS
    BATA, A
    VANYI, A
    LASZTITY, R
    ACTA VETERINARIA HUNGARICA, 1984, 32 (1-2) : 51 - 56
  • [48] ENZYME CAPTORS IN FERMENTATION - APPLICATION TO QUANTITATIVE-DETERMINATION OF SUGAR
    ROMETTE, JL
    THOMAS, D
    ANNALES DE MICROBIOLOGIE, 1980, A131 (02): : 227 - 228