SPR detection of cardiac troponin T for acute myocardial infarction

被引:69
|
作者
Pawula, Maria [1 ]
Altintas, Zeynep [1 ]
Tothill, Ibtisam E. [1 ]
机构
[1] Cranfield Univ, Cranfield MK43 0AL, Beds, England
关键词
Cardiac troponin T; Acute myocardial infarction; SPR; Immunosensor; Early diagnosis; SURFACE-PLASMON RESONANCE; QUARTZ-CRYSTAL MICROBALANCE; SENSITIVE DETECTION; GOLD NANOPARTICLES; SERUM SAMPLES; REAL-TIME; IMMUNOSENSOR; BIOMARKER; CHEMISTRY; BIOSENSOR;
D O I
10.1016/j.talanta.2015.06.006
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A surface plasmon resonance (SPR) sensor developed for the rapid, sensitive and specific detection of cardiac troponin T (cTnT) in serum samples is reported in this work. An extensive optimisation of assay parameters was conducted to achieve optimal detection strategy. Both direct and sandwich immunoassay formats were investigated and optimised. The response obtained was enhanced further by the use of gold nanoparticles (AuNPs) conjugated to the anti-cTnT detection antibody. A regeneration method was developed to enable the reuse of the SPR sensor for multiple sample application. The SPR immunosensor showed good reproducibility for cTnT detection in the concentration range of 25-1000 ng mL(-1) and 5-400 ng mL(-1) for the direct and sandwich assays in buffer, respectively. The linear regression analysis was performed and R-2 value was found as 0.99 for both assays. In order to optimise the sensor for serum analysis, nonspecific binding of serum proteins was reduced through the use of additives in the dilution buffer. To achieve greater sensitivity, the performance of the cTnT immunosensor sandwich assay in human serum was evaluated using non-modified and AuNP modified detector antibodies. A detection limit (LOD) for the immunosensor in 50% serum was assessed as 5 ng mL(-1) cTnT for the standard sandwich assay and 0.5 ng mL(-1) cTnT when using AuNP conjugated detector antibodies with a linear dynamic range of 0.5-40 ng mL(-1). The dissociation constant was found as 3.28 x 10(-9) M using Langmuir binding model which indicates high affinity between cTnT and its antibody. The proposed SPR immunosensor has a promising potential to be developed for point-of-care testing for the early diagnosis of acute myocardial infarction (AMI). This method can also be used for the rapid detection of biomarkers in central nervous system diseases. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:823 / 830
页数:8
相关论文
共 50 条
  • [1] Cardiac troponin T as a postmortem biomarker for acute myocardial infarction
    Moridi, Michael
    Magnusson, Cecilia
    Zilg, Brita
    FORENSIC SCIENCE INTERNATIONAL, 2022, 341
  • [2] Early diagnostic efficiency of cardiac troponin I and troponin T for acute myocardial infarction
    Tucker, JF
    Collins, RA
    Anderson, AJ
    Hauser, J
    Kalas, J
    Apple, FS
    ACADEMIC EMERGENCY MEDICINE, 1997, 4 (01) : 13 - 21
  • [3] A model of cardiac troponin T release in patient with acute myocardial infarction
    Procopio, Anna
    De Rosa, Salvatore
    Covello, Caterina
    Merola, Alessio
    Sabatino, Jolanda
    De Luca, Alessia
    Indolfi, Ciro
    Amato, Francesco
    Cosentino, Carlo
    2017 IEEE 56TH ANNUAL CONFERENCE ON DECISION AND CONTROL (CDC), 2017,
  • [4] CARDIAC TROPONIN-T IN DIAGNOSIS OF ACUTE MYOCARDIAL-INFARCTION
    MAIR, J
    ARTNERDWORZAK, E
    LECHLEITNER, P
    SMIDT, J
    WAGNER, I
    DIENSTL, F
    PUSCHENDORF, B
    CLINICAL CHEMISTRY, 1991, 37 (06) : 845 - 852
  • [5] Fragmentation of human cardiac troponin T after acute myocardial infarction
    Katrukha, Ivan A.
    Riabkova, Natalia S.
    Kogan, Alexander E.
    Vylegzhanina, Alexandra V.
    Mukharyamova, Kadriya Sh.
    Bogomolova, Agnessa P.
    Zabolotskii, Artur I.
    Koshkina, Ekaterina V.
    Bereznikova, Anastasia V.
    Katrukha, Alexey G.
    CLINICA CHIMICA ACTA, 2023, 542
  • [6] Diagnosis of myocardial infarction: Cardiac troponin I or troponin T?
    Lee, Kuan Ken
    Shah, Anoop S. V.
    Mills, Nicholas L.
    CLINICAL BIOCHEMISTRY, 2014, 47 (06) : 319 - 320
  • [7] Investigation of release and degradation of cardiac troponin T in patients with acute myocardial infarction
    Michielsen, Etienne C. H. J.
    Diris, Jart H. C.
    Kleijnen, Vincent W. V. C.
    Wodzig, Will K. W. H.
    Van Dieijen-Visser, Marja P.
    CLINICAL BIOCHEMISTRY, 2007, 40 (12) : 851 - 855
  • [8] CARDIAC TROPONIN-T IMMUNOASSAY FOR DIAGNOSIS OF ACUTE MYOCARDIAL-INFARCTION
    WU, AHB
    VALDES, R
    APPLE, FS
    GORNET, T
    STONE, MA
    MAYFIELDSTOKES, S
    INGERSOLLSTROUBOS, AM
    WILER, B
    CLINICAL CHEMISTRY, 1994, 40 (06) : 900 - 907
  • [9] Voltammetric biosensors for analytical detection of cardiac troponin biomarkers in acute myocardial infarction
    Pourali, Ali
    Rashidi, Mohammad Reza
    Barar, Jaleh
    Pavon-Djavid, Graciela
    Omidi, Yadollah
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2021, 134
  • [10] Implementation of serum cardiac troponin I as marker for detection of acute myocardial infarction
    Falahati, A
    Sharkey, SW
    Christensen, D
    McCoy, M
    Miller, EA
    Murakami, MA
    Apple, FS
    AMERICAN HEART JOURNAL, 1999, 137 (02) : 332 - 337