High intensity focused ultrasound (HIFU) focal spot localization using harmonic motion imaging (HMI)

被引:25
|
作者
Han, Yang [1 ]
Hou, Gary Yi [1 ]
Wang, Shutao [1 ]
Konofagou, Elisa [1 ,2 ]
机构
[1] Columbia Univ, Dept Biomed Engn, New York, NY 10027 USA
[2] Columbia Univ, Dept Radiol, New York, NY 10027 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 15期
基金
美国国家卫生研究院;
关键词
HMIFU; HIFU; focal spot localization; ACOUSTIC RADIATION FORCE; DIAGNOSTIC ULTRASOUND; IN-VITRO; LIVER; ELASTOGRAPHY; FEASIBILITY; VALIDATION; DEPENDENCE; ABLATION; TISSUES;
D O I
10.1088/0031-9155/60/15/5911
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Several ultrasound-based imaging modalities have been proposed for image guidance and monitoring of high-intensity focused ultrasound (HIFU) treatment. However, accurate localization and characterization of the effective region of treatment (focal spot) remain important obstacles in the clinical implementation of HIFU ablation. Harmonic motion imaging for focused ultrasound (HMIFU) is a HIFU monitoring technique that utilizes radiation-force-induced localized oscillatory displacement. HMIFU has been shown to correctly identify the formation and extent of HIFU thermal ablation lesions. However a significant problem remains in identifying the location of the HIFU focus, which is necessary for treatment planning. In this study, the induced displacement was employed to localize the HIFU focal spot inside the tissue prior to treatment. Feasibility was shown with two separate systems. The 1D HMIFU system consisted of a HIFU transducer emitting an amplitude-modulated HIFU beam for mechanical excitation and a confocal single-element, pulse-echo transducer for simultaneous RF acquisition. The 2D HIFU system consists of a HIFU phased array, and a co-axial imaging phased array for simultaneous imaging. Initial feasibility was first performed on tissue-mimicking gelatin phantoms and the focal zone was defined as the region corresponding to the -3dB full width at half maximum of the HMI displacement. Using the same parameters, in vitro experiments were performed in canine liver specimens to compare the defined focal zone with the lesion. In vitro measurements showed good agreement between the HMI predicted focal zone and the induced HIFU lesion location. HMIFU was experimentally shown to be capable of predicting and tracking the focal region in both phantoms and in vitro tissues. The accuracy of focal spot localization was evaluated by comparing with the lesion location in post-ablative tissues, with a R-2 = 0.821 at p < 0.002 in the 2D HMI system. We demonstrated the feasibility of using this HMI-based technique to localize the HIFU focal spot without inducing thermal changes during the planning phase. The focal spot localization method has also been applied on ex vivo human breast tissue ablation and can be fully integrated into any HMI system for planning purposes.
引用
收藏
页码:5911 / 5924
页数:14
相关论文
共 50 条
  • [1] Multi-parametric monitoring of High Intensity Focused Ultrasound (HIFU) Treatment Using Harmonic Motion Imaging For Focused Ultrasound (HMIFU)
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Konofagou, Elisa
    12TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2012, 1503 : 140 - 145
  • [2] Ultrasound Acoustic Radiation Force Impulse imaging for High Intensity Focused Ultrasound focal spot localization
    Cafarelli, A.
    Chanel, L. A.
    Di Bartolo, F.
    Locteau, H.
    Tognarelli, S.
    Dumont, E.
    Menciassi, A.
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 5713 - 5716
  • [3] Optimization of Real-Time Acoustical and Mechanical Monitoring of High Intensity Focused Ultrasound (HIFU) treatment using Harmonic Motion Imaging for High Focused Ultrasound (HMIFU)
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Konofagou, Elisa E.
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 6281 - 6284
  • [4] Real-time Monitoring of High Intensity Focused Ultrasound (HIFU) Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU)
    Grondin, Julien
    Payen, Thomas
    Wang, Shutao
    Konofagou, Elisa E.
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2015, (105):
  • [5] Coagulation Using High Intensity Focused Ultrasound (HIFU)
    Nguyen, Phuc V.
    Oh, Junghwan
    Kang, Hyun Wook
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS X, 2014, 8926
  • [6] High intensity focused ultrasound (HIFU)
    Barkin, Jack
    CANADIAN JOURNAL OF UROLOGY, 2011, 18 (02) : 5634 - 5643
  • [7] High intensity focused ultrasound (HIFU)
    Ganzer, R.
    RADIOLOGE, 2017, 57 (08): : 659 - 664
  • [8] Real-time Focused Ultrasound Surgery (FUS) Monitoring Using Harmonic Motion Imaging (HMI)
    Maleke, Caroline
    Konofagou, Elisa E.
    8TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2009, 1113 : 236 - 240
  • [9] Robotized High Intensity Focused Ultrasound (HIFU) system for treatment of mobile organs using motion tracking by ultrasound imaging: An in vitro study
    Chanel, Laure-Anais
    Nageotte, Florent
    Vappou, Jonathan
    Luo, Jianwen
    Cuvillon, Loic
    de Mathelin, Michel
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 2571 - 2575
  • [10] Multi-parametric monitoring and assessment of high-intensity focused ultrasound (HIFU) boiling by harmonic motion imaging for focused ultrasound (HMIFU): an ex vivo feasibility study
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Konofagou, Elisa E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (05): : 1121 - 1145