Real-time Monitoring of High Intensity Focused Ultrasound (HIFU) Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU)

被引:9
|
作者
Grondin, Julien [1 ]
Payen, Thomas [1 ]
Wang, Shutao [1 ]
Konofagou, Elisa E. [1 ,2 ]
机构
[1] Columbia Univ, Dept Biomed Engn, New York, NY 10027 USA
[2] Columbia Univ, Dept Radiol, New York, NY 10027 USA
来源
基金
美国国家卫生研究院;
关键词
Bioengineering; Issue; 105; HIFU ablation; Harmonic motion imaging; Real-time monitoring; High frame rate imaging; Elastography; Lesion monitoring; Liver ablation; Canine liver; THERMAL ABLATION; ELASTOGRAPHY;
D O I
10.3791/53050
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Harmonic Motion Imaging for Focused Ultrasound (HMIFU) is a technique that can perform and monitor high-intensity focused ultrasound (HIFU) ablation. An oscillatory motion is generated at the focus of a 93-element and 4.5 MHz center frequency HIFU transducer by applying a 25 Hz amplitude-modulated signal using a function generator. A 64-element and 2.5 MHz imaging transducer with 68kPa peak pressure is confocally placed at the center of the HIFU transducer to acquire the radio-frequency (RF) channel data. In this protocol, real-time monitoring of thermal ablation using HIFU with an acoustic power of 7 W on canine livers in vitro is described. HIFU treatment is applied on the tissue during 2 min and the ablated region is imaged in real-time using diverging or plane wave imaging up to 1,000 frames/second. The matrix of RF channel data is multiplied by a sparse matrix for image reconstruction. The reconstructed field of view is of 90 degrees for diverging wave and 20 mm for plane wave imaging and the data are sampled at 80 MHz. The reconstruction is performed on a Graphical Processing Unit (GPU) in order to image in real-time at a 4.5 display frame rate. 1-D normalized cross-correlation of the reconstructed RF data is used to estimate axial displacements in the focal region. The magnitude of the peak-to-peak displacement at the focal depth decreases during the thermal ablation which denotes stiffening of the tissue due to the formation of a lesion. The displacement signal-to-noise ratio (SNRd) at the focal area for plane wave was 1.4 times higher than for diverging wave showing that plane wave imaging appears to produce better displacement maps quality for HMIFU than diverging wave imaging.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Optimization of Real-Time Acoustical and Mechanical Monitoring of High Intensity Focused Ultrasound (HIFU) treatment using Harmonic Motion Imaging for High Focused Ultrasound (HMIFU)
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Konofagou, Elisa E.
    [J]. 2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 6281 - 6284
  • [2] Multi-parametric monitoring of High Intensity Focused Ultrasound (HIFU) Treatment Using Harmonic Motion Imaging For Focused Ultrasound (HMIFU)
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Konofagou, Elisa
    [J]. 12TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2012, 1503 : 140 - 145
  • [3] High-Intensity Focused Ultrasound Monitoring Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU) Under Boiling or Slow Denaturation Conditions
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Apostolakis, Iason-Zacharias
    Konofagou, Elisa E.
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (07) : 1308 - 1319
  • [4] Multi-parametric monitoring and assessment of high-intensity focused ultrasound (HIFU) boiling by harmonic motion imaging for focused ultrasound (HMIFU): an ex vivo feasibility study
    Hou, Gary Y.
    Marquet, Fabrice
    Wang, Shutao
    Konofagou, Elisa E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (05): : 1121 - 1145
  • [5] Techniques for real-time monitoring and control for HIFU (High Intensity Focused Ultrasound) ablation in porcine brains in vitro studies
    Long, Tao
    Amin, Viren
    McClure, Scott
    Robert, Ronald
    Wu, Liangshou
    Thompson, R. B.
    Ryken, Timothy
    [J]. 2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 1788 - +
  • [6] Simulation of HMIFU (Harmonic Motion Imaging for Focused Ultrasound) with in-vitro validation
    Hou, Gary Y.
    Luo, Jianwen
    Maleke, Caroline
    Konofagou, Elisa E.
    [J]. 2010 IEEE 36TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2010,
  • [7] Fluorescence imaging for real-time monitoring of high-intensity focused ultrasound cardiac ablation
    Deng, CRX
    Qu, FJ
    Nikolski, VP
    Zhou, Y
    Efimov, IR
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2005, 33 (10) : 1352 - 1359
  • [8] Fluorescence Imaging for Real-Time Monitoring of High-Intensity Focused Ultrasound Cardiac Ablation
    Cheri X. Deng
    Fujian Qu
    Vladimir P. Nikolski
    Yun Zhou
    Igor R. Efimov
    [J]. Annals of Biomedical Engineering, 2005, 33 : 1352 - 1359
  • [9] Real-time Focused Ultrasound Surgery (FUS) Monitoring Using Harmonic Motion Imaging (HMI)
    Maleke, Caroline
    Konofagou, Elisa E.
    [J]. 8TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2009, 1113 : 236 - 240
  • [10] Harmonic motion imaging for focused ultrasound (HMIFU): a fully integrated technique for sonication and monitoring of thermal ablation in tissues
    Maleke, C.
    Konofagou, E. E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (06): : 1773 - 1793