Evaluation of inhomogeneity correction factors for 6 MV flattening filter-free beams with brass compensators

被引:2
|
作者
Robinson, Joshua [1 ,2 ]
Opp, Daniel [1 ]
Zhang, Geoffrey [1 ]
Feygelman, Vladimir [1 ]
机构
[1] H Lee Moffitt Canc Ctr & Res Inst, Dept Radiat Oncol, Tampa, FL 33612 USA
[2] Univ S Florida, Dept Phys, Tampa, FL 33620 USA
来源
关键词
flattening filter-free; compensator; inhomogeneity correction factor; dose calculation accuracy; lung; MANAGEMENT; ACCURACY; DELIVERY; MOTION;
D O I
10.1120/jacmp.v14i3.3990
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The 6 MV flattening filter-free (FFF) beam has been commissioned for use with compensators at our institution. This novel combination promises advantages in mitigating tumor motion due to the reduced treatment time made possible by the greatly increased dose rate of the FFF beam. Given the different energy spectrum of the FFF beam and the beam hardening effect of the compensator, the accuracy of the treatment planning system (TPS) model in the presence of low-density heterogeneities cannot be assumed. Therefore, inhomogeneity correction factors (ICF) for an FFF beam attenuated by brass slabs were measured and compared to the TPS calculations in this work. The ICF is the ratio of the point dose in the presence of inhomogeneity to the dose in the same point in a homogeneous medium. The ICFs were measured with an ion chamber at a number of points in a flat water-equivalent slab phantom containing a 7.5 cm deep heterogeneity (air or 0.27 g/cm(3) wood). Comparisons for the FFF beam were carried out for the field sizes from 5 x 5 to 20 x 20 cm(2) with the brass slabs ranging from 0 to 5 cm in thickness. For a low-density wood heterogeneity in a slab phantom, with the exception of the point 1 cm beyond the proximal buildup interface, the TPS handles the inhomogeneity correction with the brass-filtered 6 MV FFF beam at the requisite 2% error level. The combinations of field sizes and compensator thicknesses when the error exceeds 2% (2.6% maximum) are not likely to be experienced in clinical practice. In terms of heterogeneity corrections, the beam model is adequate for clinical use.
引用
收藏
页码:226 / 231
页数:6
相关论文
共 50 条
  • [41] Dose Rate Response of Digital Megavolt Imager Detector for Flattening Filter-Free Beams
    Xu, Z.
    Kim, J.
    Han, J.
    Hsia, A.
    Ryu, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2889 - 2889
  • [42] Applicability of an Aperture-Morphing Based Replanning Algorithm for Flattening Filter-Free Beams
    Ahunbay, E. E.
    Li, A. X.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S930 - S931
  • [43] Impact of 6 MV & 10 MV Flattened and Flattening Filter Free Beams in Whole Brain Radiotherapy: A Treatment Planning Study
    Suresh T.
    Madeswaran S.
    Iranian Journal of Medical Physics, 2021, 18 (03) : 218 - 225
  • [44] Improving the Therapeutic Ratio for Lung SBRT Through the Use of Flattening Filter-Free Beams
    Vassiliev, O. N.
    Wang, H.
    Peterson, C. B.
    Chang, J. Y.
    Mohan, R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E251 - E251
  • [45] Stereotactic Radiosurgery for Multiple Brain Metastases Using Flattening Filter-Free Beam and Comparison Study with the Regular 6MV
    Wang, J.
    Murphy, K.
    Rice, R.
    Pawlicki, T.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [46] Evaluation of dosimetric properties of 6 & 10 MV photon beams from a liner accelerator with no flattening filter
    Pearson, D.
    Parsai, E.
    Feldmeier, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2099 - 2099
  • [47] Evaluation of beam modeling for small fields using a flattening filter-free beam
    Kawahara D.
    Ozawa S.
    Nakashima T.
    Aita M.
    Tsuda S.
    Ochi Y.
    Okumura T.
    Masuda H.
    Ohno Y.
    Murakami Y.
    Nagata Y.
    Radiological Physics and Technology, 2017, 10 (1) : 33 - 40
  • [48] Stereotactic body radiation therapy for prostate cancer: a dosimetric comparison of IMRT and VMAT using flattening filter and flattening filter-free beams
    El-Sayed, Sherif M.
    El-Gebaly, Reem H.
    Fathy, Mohamed M.
    Abdelaziz, Dina M.
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2024, 63 (03) : 423 - 431
  • [49] Influence of the Applied Voltage On the Ion Recombination Correction Factor Using the Two-Voltage Technique in Flattening Filter-Free Beams
    Yada, R.
    Hayashi, N.
    Akasaka, H.
    Mukumoto, N.
    Sasaki, R.
    MEDICAL PHYSICS, 2017, 44 (06) : 3220 - 3220
  • [50] A Simple Source Model for 6 MV Flattening Filter Free Photon Beams Monte Carlo Dose Calculations
    Ezzati, Ahad Ollah
    Studenski, Matthew T.
    Jamshidi, Negin
    PHYSICS OF PARTICLES AND NUCLEI LETTERS, 2021, 18 (07) : 791 - 798