Magnetic Resonance Imaging in Patients with Recently Implanted Pacemakers

被引:36
|
作者
Friedman, Hannah L. [1 ]
Acker, Nancy [1 ]
Dalzell, Connie [1 ]
Shen, Win K. [2 ]
Asirvatham, Samuel J. [1 ]
Cha, Yong-Mei [1 ]
Hodge, David [3 ]
Felmlee, Joel [4 ]
Watson, Robert [4 ]
Friedman, Paul A. [1 ]
机构
[1] Mayo Clin, Heart Rhythm Serv, Rochester, MN 55905 USA
[2] Mayo Clin, Serv Cardiol, Scottsdale, AZ USA
[3] Mayo Clin, Serv Biostat, Rochester, MN 55905 USA
[4] Mayo Clin, Serv Radiol, Rochester, MN 55905 USA
来源
关键词
cardiac pacemaker; magnetic resonance imaging; INDUSTRY VIEWPOINT; CARDIOVERTER-DEFIBRILLATORS; MRI; ICDS; SAFETY;
D O I
10.1111/pace.12213
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Recent studies have shown that magnetic resonance imaging (MRI) of patients with pacemakers can be safely performed under careful monitoring, but they excluded patients with recently implanted devices. Patients with recent implants may be at a greater risk for complications during MRI imaging due to lack of lead and wound maturity. Methods We implemented a clinical protocol for MRI imaging of patients with implanted cardiac devices, and prospectively collected data. For this study, we retrospectively analyzed two groups of patients: those with recently implanted (42 days) and nonrecently implanted (>42 days) leads at the time of MRI scanning. All devices were interrogated before and after scanning, and were reprogrammed during the scan as per protocol. Results Of the 219 scans (in 171 patients), eight included patients with recently implanted (range: 7-36 days) and 211 with only nonrecently implanted pacemaker leads. During the scan, there were no complications in the early or late group. In one patient imaged 79 days postimplant, frequent premature ventricular complexes were noted during the scan, requiring no action. No patient reported pain during or immediately after the procedure. No clinically significant changes in function were seen at subsequent follow up (average 104 days post-MRI). Compared to patients with nonrecently implanted leads, there was no difference in any parameter between the two groups. Conclusions With a strong clinical indication and with careful monitoring, MRI imaging is feasible in patients with recently implanted pacemakers, although experience is limited.
引用
收藏
页码:1090 / 1095
页数:6
相关论文
共 50 条
  • [41] Quantitative Assessment of Artifacts on Cardiac Magnetic Resonance Imaging of Patients with Pacemakers and Implantable Defibrillators
    Sasaki, Takeshi
    Hansford, Rozann
    Zviman, Menekhem M.
    Berger, Ronald D.
    Calkins, Hugh
    Bluemke, David A.
    Halperin, Henry R.
    Nazarian, Saman
    CIRCULATION, 2010, 122 (21)
  • [42] 3.0 T magnetic resonance imaging scanning on different body regions in patients with pacemakers
    Xiaohui Ning
    Xiaofei Li
    Xiaohan Fan
    Keping Chen
    Wei Hua
    Zhimin Liu
    Yan Dai
    Xiuyu Chen
    Minjie Lu
    Shihua Zhao
    Shu Zhang
    Journal of Interventional Cardiac Electrophysiology, 2021, 61 : 545 - 550
  • [43] Magnetic Resonance Imaging in Nondependent Pacemaker Patients with Pacemakers and Defibrillators with a Nearly Depleted Battery
    Okamura, Hideo
    Padmanabhan, Deepak
    Watson, Robert E., Jr.
    Dalzell, Connie
    Acker, Nancy
    Jondal, Mary
    Romme, Abby L.
    Cha, Yong-mei
    Asirvatham, Samuel J.
    Felmlee, Joel P.
    Friedman, Paul A.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2017, 40 (05): : 476 - 481
  • [44] Magnetic resonance imaging in patients with pacemakers and implantable cardioverter-defibrillators: A systematic review
    Martins, R. -P.
    Baruteau, A. -E.
    Treguer, F.
    Cesari, O.
    Carsin-Nicol, B.
    Langella, B.
    Leclercq, C.
    Daubert, J. -C.
    Mabo, P.
    ANNALES DE CARDIOLOGIE ET D ANGEIOLOGIE, 2010, 59 (04): : 221 - 228
  • [45] 3.0 T magnetic resonance imaging scanning on different body regions in patients with pacemakers
    Ning, Xiaohui
    Li, Xiaofei
    Fan, Xiaohan
    Chen, Keping
    Hua, Wei
    Liu, Zhimin
    Dai, Yan
    Chen, Xiuyu
    Lu, Minjie
    Zhao, Shihua
    Zhang, Shu
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 61 (03) : 545 - 550
  • [46] Initial experience with magnetic resonance imaging-safe pacemakers
    Jung, Werner
    Zvereva, Vlada
    Hajredini, Bajram
    Jaeckle, Sebastian
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2011, 32 (03) : 213 - 219
  • [47] THE BEHAVIOR OF DDD PACEMAKERS DURING MAGNETIC-RESONANCE IMAGING
    ERLEBACHER, JA
    CAHILL, P
    PANNIZZO, F
    KNOWLES, J
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (02): : 291 - 291
  • [48] Do we need pacemakers resistant to magnetic resonance imaging?
    Naehle, CP
    Litt, H
    Lewalter, T
    Sommer, T
    EUROPACE, 2006, 8 (05): : 388 - 388
  • [49] Pacemakers and magnetic resonance imaging Current status and survey in Switzerland
    Weishaupt, Dominik
    Bremerich, Jens
    Duru, Firat
    Hoppe, Hanno
    Rizzo, Elena
    Votik, Peter
    Luechinger, Roger
    SWISS MEDICAL WEEKLY, 2011, 141
  • [50] Effect of magnetic resonance imaging in non-compatible pacemakers
    Fernandez, M. J.
    Soto, M.
    Torres, M.
    Monsalve, R.
    Ardiles, J.
    Moreno, V.
    Hirsch, M.
    Quininir, L.
    Diaz, A.
    Neira, V.
    Miranda, R.
    EUROPEAN HEART JOURNAL, 2017, 38 : 364 - 365