Potential thermal artifacts in hip joint wear simulators

被引:1
|
作者
Lu, Z [1 ]
McKellop, H
Liao, P
Benya, P
机构
[1] Los Angeles Orthopaed Hosp, J Vernon Luck Orthopaed Res Ctr, Los Angeles, CA USA
[2] Univ So Calif, Sch Med, Dept Orthopaed, Los Angeles, CA 90033 USA
[3] Univ So Calif, Dept Biomed Engn, Los Angeles, CA 90089 USA
来源
关键词
wear; hip prosthesis; polyethylene; frictional heat; hip simulator;
D O I
10.1002/(SICI)1097-4636(1999)48:4<458::AID-JBM9>3.3.CO;2-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Frictional heat was monitored during wear tests of ultrahigh molecular weight polyethylene acetabular cups bearing against femoral balls of metal or ceramic in a hip simulator, using bovine serum as a lubricant. About 1 to 2 h of continuous cycling were required for the temperature in the zone of contact between the cup and ball to rise to its maximum steady value, and this equilibrium temperature was markedly higher with increased load and/or cycling rate, Frictional heating caused substantial precipitation of the proteins from the serum and, in some of the tests running at 1.5 or 2 Hz, an adherent proteinaceous layer was observed attached to the surface of the balls. The maximum temperature was also substantially higher in tests run with the cup mounted above the ball rather than below. Surprisingly, the tests running at higher frictional torque and temperature (i.e., those with the most protein precipitation and/or adherent layers) produced the least wear of. the polyethylene. This might have been due to the solid proteins that formed a protective layer between the ball and cup. Because patients with hip prostheses typically do not walk for hours without rest, the maximum temperatures irt vivo are likely to be much lower than those reached in the hip simulator, Therefore, the affects of protein precipitation on the resultant wear properties of the materials should be considered potential artifacts of the hip simulator tests. Increasing the volume of the lubricant bath reduced the maximum temperatures for tests running at 1.5 Nz but had little affect at 2 Hz. Reducing the cycling rate is an effective way to avoid overheating of the specimens, but this necessarily extends the time required to complete a test. (C) 1999 John Wiley & Sons, Inc.
引用
收藏
页码:458 / 464
页数:7
相关论文
共 50 条
  • [1] Hip joint simulators: State of the art
    Affatato, S.
    Leardini, W.
    Zovalloni, M.
    BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2006, : 171 - +
  • [2] Simultaneous measurement of friction and wear in hip simulators
    Haider, Hani
    Weisenburger, Joel N.
    Garvin, Kevin L.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (05) : 373 - 388
  • [3] Generation of physiological parameter sets for hip joint motions and loads during daily life activities for application in wear simulators of the artificial hip joint
    Fabry, Christian
    Herrmann, Sven
    Kaehler, Michael
    Klinkenberg, Ernst-Dieter
    Woernle, Christoph
    Bader, Rainer
    MEDICAL ENGINEERING & PHYSICS, 2013, 35 (01) : 131 - 139
  • [4] Metallographic evaluation of hip joint implants wear and electrochemical implants potential
    Kmiec, Krzysztof
    Sibinski, Marcin
    Synder, Marek
    Drobniewski, Marek
    Kozlowski, Piotr
    HIP INTERNATIONAL, 2014, 24 (06) : 604 - 609
  • [5] Differences of the mechanical setup of hip simulators and their consequences on the outcome of hip wear testing
    Institute of Sensor and Actuator Systems, Microsystems Technology, Vienna University of Technology, Floragasse 7, A 1040 Vienna, Austria
    不详
    不详
    不详
    不详
    J. ASTM Int., 2006, 8
  • [6] Differences of the mechanical setup of hip simulators and their consequences on the outcome of hip wear testing
    Reinisch, Georg
    Schoerg, Joachim
    Leder, Eva
    Judmann, Kurt P.
    Plitz, Wolfgang
    Franek, Friedrich
    WEAR OF ARTICULATING SURFACES: UNDERSTANDING JOINT SIMULATION, 2006, 1472 : 115 - +
  • [7] Parametric analysis for the design of hip joint replacement simulators
    Torabnia, Shams
    Mihcin, Senay
    Lazoglu, Ismail
    2021 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (IEEE MEMEA 2021), 2021,
  • [8] Some aspects of frictional measurements in hip joint simulators
    Unsworth, Anthony
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (05) : 359 - 365
  • [9] CHARACTERIZATION OF UHMWPE HIP CUPS RUN ON JOINT SIMULATORS
    EYERER, P
    KURTH, M
    MCKELLUP, HA
    MITTLMEIER, T
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1987, 21 (03): : 275 - 291
  • [10] Wear of artificial hip joint material
    Watters, EPJ
    Spedding, PL
    Grimshaw, J
    Duffy, JM
    Spedding, RL
    CHEMICAL ENGINEERING JOURNAL, 2005, 112 (1-3) : 137 - 144