High energy resolution five-crystal spectrometer for high quality fluorescence and absorption measurements on an x-ray absorption spectroscopy beamline

被引:55
|
作者
Llorens, Isabelle [1 ,2 ]
Lahera, Eric [3 ,4 ]
Delnet, William [3 ,4 ]
Proux, Olivier [3 ,4 ]
Braillard, Aurelien [4 ,5 ]
Hazemann, Jean-Louis [4 ,5 ]
Prat, Alain [4 ,5 ]
Testemale, Denis [4 ,5 ]
Dermigny, Quentin [6 ]
Gelebart, Frederic [6 ]
Morand, Marc [6 ]
Shukla, Abhay [6 ]
Bardou, Nathalie [7 ]
Ulrich, Olivier [1 ,8 ]
Arnaud, Stephan [5 ,9 ]
Berar, Jean-Francois [5 ,9 ]
Boudet, Nathalie [5 ,9 ]
Caillot, Bernard [5 ,9 ]
Chaurand, Perrine [10 ]
Rose, Jerome [10 ]
Doelsch, Emmanuel [11 ]
Martin, Philippe [12 ]
Solari, Pier Lorenzo [2 ]
机构
[1] CEA, DSM, INAC, SP2M,NRS, F-38054 Grenoble 9, France
[2] Synchrotron SOLEIL MARS Beamline, F-91192 Gif Sur Yvette, France
[3] Univ Grenoble 1, Observ Sci Univers Grenoble, CNRS, UMS 832, F-38041 Grenoble 9, France
[4] ESRF, FAME Beamline BM30B, F-38043 Grenoble 9, France
[5] CNRS, Inst Neel, UPR 2940, F-38042 Grenoble 9, France
[6] Inst Mineral & Phys Milieux Condenses, UMR 7590, F-75005 Paris, France
[7] CNRS, Lab Photon & Nanostruct, UPR 20, F-91460 Marcoussis, France
[8] ESRF, IF Beamline BM32, F-38043 Grenoble 9, France
[9] ESRF, Beamline BM02 D2AM, F-38043 Grenoble 9, France
[10] Ctr Europeen Rech & Enseignement Geosci Environm, UMR 7730, F-13545 Aix En Provence, France
[11] CIRAD, UPR Recyclage & Risque, Montpellier, France
[12] CEA, DEN, DEC, SESC,LLCC, F-13108 St Paul Les Durance, France
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2012年 / 83卷 / 06期
关键词
RAMAN-SCATTERING; DYNAMICAL THEORY; DIFFRACTION; ADSORPTION; EMISSION; ANALYZER; ELEMENTS; PERFORMANCE; ARSENITE; CRYSTALS;
D O I
10.1063/1.4728414
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Fluorescence detection is classically achieved with a solid state detector (SSD) on x-ray absorption spectroscopy (XAS) beamlines. This kind of detection however presents some limitations related to the limited energy resolution and saturation. Crystal analyzer spectrometers (CAS) based on a Johann-type geometry have been developed to overcome these limitations. We have tested and installed such a system on the BM30B/CRG-FAME XAS beamline at the ESRF dedicated to the structural investigation of very dilute systems in environmental, material and biological sciences. The spectrometer has been designed to be a mobile device for easy integration in multi-purpose hard x-ray synchrotron beamlines or even with a laboratory x-ray source. The CAS allows to collect x-ray photons from a large solid angle with five spherically bent crystals. It will cover a large energy range allowing to probe fluorescence lines characteristic of all the elements from Ca (Z = 20) to U (Z = 92). It provides an energy resolution of 1-2 eV. XAS spectroscopy is the main application of this device even if other spectroscopic techniques (RIXS, XES, XRS, etc.) can be also achieved with it. The performances of the CAS are illustrated by two experiments that are difficult or impossible to perform with SSD and the complementarity of the CAS vs SSD detectors is discussed. (C) 2012 American Institute of Physics. [http://dx.doi.org/10.1063/1.4728414]
引用
收藏
页数:9
相关论文
共 50 条
  • [21] A Photochemically Generated Selenyl Free Radical Observed by High Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy
    Nehzati, Susan
    Dolgova, Natalia V.
    Sokaras, Dimosthenis
    Kroll, Thomas
    Cotelesage, Julien J. H.
    Pickering, Ingrid J.
    George, Graham N.
    INORGANIC CHEMISTRY, 2018, 57 (17) : 10867 - 10872
  • [22] Exploring Platinum Speciation with X-ray Absorption Spectroscopy under High-Energy Resolution Fluorescence Detection Mode
    Laskar, Clement
    Bazarkina, Elena F. F.
    Kokh, Maria A. A.
    Hazemann, Jean-Louis
    Foulon, Stephane
    Leynaud, Olivier
    Desmaele, Elsa
    Pokrovski, Gleb S. S.
    MINERALS, 2022, 12 (12)
  • [23] High energy resolution fluorescence detection x-ray absorption spectroscopy: Detection of adsorption sites in supported metal catalysts
    Tromp, Moniek
    van Bokhoven, Jeroen A.
    Safonova, Olga V.
    De Groot, Frank M. F.
    Evans, John
    Glatzel, Pieter
    X-RAY ABSORPTION FINE STRUCTURE-XAFS13, 2007, 882 : 651 - +
  • [24] The Dispersive X-ray Absorption Spectroscopy beamline at LNLS
    Tolentino, Helio C. N.
    Cezar, Julio C.
    Watanabe, Noemia
    Piamonteze, Cinthia
    Souza-Neto, Narcizo M.
    Tamura, Edilson
    Ramos, AlineY.
    Neueschwander, Regis
    PHYSICA SCRIPTA, 2005, T115 : 977 - 979
  • [25] X-ray absorption spectroscopy and high pressure
    Itie, J.
    Baudelet, F.
    Dartyge, E.
    Fontaine, A.
    Tolentino H.
    San-Miguel, A.
    High Pressure Research, 1992, 8 (5-6) : 697 - 702
  • [26] High-resolution XEOL spectroscopy setup at the X-ray absorption spectroscopy beamline P65 of PETRA III
    Levcenko, S.
    Biller, R.
    Pfeiffelmann, T.
    Ritter, K.
    Falk, H. H.
    Wang, T.
    Siebentritt, S.
    Welter, E.
    Schnohr, C. S.
    JOURNAL OF SYNCHROTRON RADIATION, 2022, 29 (Pt 5) : 1209 - 1215
  • [27] Johann spectrometer for high resolution X-ray spectroscopy
    Machek, Pavel
    Welter, Edmund
    Caliebe, Wolfgang
    Brueggmann, Ulf
    Draeger, Guenter
    Froeba, Michael
    SYNCHROTRON RADIATION INSTRUMENTATION, PTS 1 AND 2, 2007, 879 : 1755 - +
  • [28] High resolution spectrometer for extended x-ray absorption fine structure measurements in the 6 keV to 15 keV energy range
    Seely, J. F.
    Hudson, L. T.
    Henins, Albert
    Feldman, U.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (11):
  • [29] THE ENERGY-DISPERSIVE X-RAY ABSORPTION SPECTROMETER DEXAFS AT THE HASYLAB BEAMLINE-X
    HAGELSTEIN, M
    CUNIS, S
    RABE, P
    FRAHM, R
    NIEMANN, W
    ZEITSCHRIFT FUR KRISTALLOGRAPHIE, 1988, 182 (1-4): : 113 - 113
  • [30] High Energy Resolution X-ray Absorption Spectroscopy of Environmentally Relevant Lead(II) Compounds
    Swarbrick, Janine C.
    Skyllberg, Ulf
    Karlsson, Torbjorn
    Glatzel, Pieter
    INORGANIC CHEMISTRY, 2009, 48 (22) : 10748 - 10756