On timing response improvement of an NE213 scintillator attached to two PMTs

被引:2
|
作者
Zare, S. [1 ]
Ghal-Eh, N. [1 ]
Bayat, E. [2 ]
机构
[1] Damghan Univ, Sch Phys, Damghan, Iran
[2] Univ Birjand, Fac Sci, Dept Phys, Bidand, Iran
关键词
Neutron-gamma discrimination; Scintillator; PMT; Timing response; PHOTRACK; DISCRIMINATION; DETECTORS; CODE;
D O I
10.1016/j.radphyschem.2013.04.036
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
A 5 cm diameter by 6 cm height NE213 scintillator attached to two XP2282 PHOTONIS photomultiplier tubes (PMTs) exposed to (241)Americium-Berylium (Am-Be) neutron-gamma source has been used for timing response studies. The neutron-gamma discrimination (NGD) measurements based on a modified zero-crossing (ZC) method show that the discrimination quality, usually expressed in figure-of-merit (FoM) and peak-to-valley (P/V) values, has been improved. The timing response evaluated with Monte Carlo light transport code, PHOTRACK, also verifies this improvement. (c) 2013 Elsevier Ltd. All rights reserved.