Calibration of high-frequency hydrophone up to 40 MHz by heterodyne interferometer

被引:36
|
作者
Yang, Ping [1 ]
Xing, Guangzhen [1 ,2 ]
He, Longbiao [1 ]
机构
[1] Natl Inst Metrol, Div Mech & Acoust, Beijing 100013, Peoples R China
[2] Harbin Inst Technol, Ultraprecis Optoelect Instrument Engn Inst, Harbin 150001, Peoples R China
关键词
High-frequency hydrophone; Calibration; Heterodyne interferometer; Digital phase demodulation; OPTICAL INTERFEROMETRY; RANGE; RECIPROCITY;
D O I
10.1016/j.ultras.2013.07.013
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A calibration technique for high-frequency hydrophone utilizing a heterodyne interferometer is presented in this article. The calibration system is mainly composed of optical and signal processing modules. In the displacement measurement, a pellicle is mounted at the surface of water to avoid acoustooptical interaction. The phase modulated carrier signal is digitized and transferred to the computer, then processed by digital phase demodulation. A phase unwrapping algorithm is employed to remove ambiguity of the arctangent function and has proven effective in large displacement measurements. Pellicle displacement and voltage output of the hydrophone in focused ultrasonic field are processed by DFT to determine the amplitudes of the fundamental and harmonic components. Experiments show that the heterodyne technique can provide hydrophone calibration up to 40 MHz, with a slightly smaller sensitivity compared with the National Physical Laboratory (NPL) calibration results for most frequency ranges. Since the heterodyne technique is independent on assumptions about the geometry of the ultrasonic field and the performance of the transducer, it can be easily extended to high frequency and high power ultrasound measurement applications. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:402 / 407
页数:6
相关论文
共 50 条
  • [1] Comparison of hydrophone calibration by reciprocity and heterodyne interferometer in the frequency range 500 kHz to 15 MHz
    Xing, Guangzhen
    Yang, Ping
    He, Longbiao
    [J]. MEASUREMENT TECHNOLOGY AND ITS APPLICATION, PTS 1 AND 2, 2013, 239-240 : 10 - 15
  • [2] High-frequency angular vibration calibration using the mirror assembly diffraction grating heterodyne laser interferometer
    Xue, Jingfeng
    Zhang, Dazhi
    Li, Xinliang
    Zhao, Weiqian
    [J]. 7TH INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTICAL TEST AND MEASUREMENT TECHNOLOGY AND EQUIPMENT, 2014, 9282
  • [3] Phase calibration of interferometer arrays at high-frequency radars
    Burrell, Angeline G.
    Yeoman, Timothy K.
    Milan, Stephen E.
    Lester, Mark
    [J]. RADIO SCIENCE, 2016, 51 (09) : 1445 - 1456
  • [4] 40 MHz high-frequency ultrafast ultrasound imaging
    Huang, Chih-Chung
    Chen, Pei-Yu
    Peng, Po-Hsun
    Lee, Po-Yang
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2185 - 2195
  • [5] High-frequency membrane hydrophone
    Lum, P
    Greenstein, M
    Grossman, C
    Szabo, TL
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (04) : 536 - 544
  • [6] Broadband PVDF membrane hydrophone for comparisons of hydrophone calibration methods up to 140 MHz
    Wilkens, Volker
    Molkenstruck, Walter
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (09) : 1784 - 1791
  • [7] An Active Stabilization Technique Of A Homodyne Interferometer Based On PTDC For High Frequency Hydrophone Calibration
    Chang, Shuaishuai
    Zhu, Haijiang
    Yang, Ping
    [J]. PROCEEDINGS OF THE 3RD INTERNATIONAL CONFERENCE ON MECHANICAL ENGINEERING AND INTELLIGENT SYSTEMS (ICMEIS 2015), 2015, 26 : 317 - 324
  • [8] A passive hydrophone for high-frequency application
    Cuneo, Stefano
    Plotnikov, A.
    Repetto, Luca
    Anghinolfi, Marco
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 567 (02): : 518 - 520
  • [9] Absolute Hydrophone Calibration to 40 MHz Using Ultrasonic Far-Field
    Matsuda, Youichi
    Yoshioka, Masahiro
    Uchida, Takeyoshi
    [J]. MATERIALS TRANSACTIONS, 2014, 55 (07) : 1030 - 1033
  • [10] A novel method for determining calibration and behavior of PVDF ultrasonic hydrophone probes in the frequency range up to 100 MHz
    Bleeker, HJ
    Lewin, PA
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (06) : 1354 - 1362