A Realistic X-Ray Simulation for C-Arm Geometry Calibration

被引:0
|
作者
Thuerauf, Sabine [1 ]
Wolf, Markus [2 ]
Koerner, Mario [2 ]
Vogt, Florian [2 ]
Hornung, Oliver [2 ]
Nasseri, M. Ali [3 ]
Knoll, Alois [4 ]
机构
[1] Fortiss GmbH, D-80805 Munich, Germany
[2] Siemens Healthcare GmbH, D-91301 Forchheim, Germany
[3] Klinikum Rechts Der Isar, D-81675 Munich, Germany
[4] Tech Univ Munich, Fac Comp Sci, Chair Robot & Embedded Syst, D-85748 Garching, Germany
来源
2016 6TH IEEE INTERNATIONAL CONFERENCE ON BIOMEDICAL ROBOTICS AND BIOMECHATRONICS (BIOROB) | 2016年
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Applications like cone beam computed tomographies (CBCTs) or 2D-3D overlays need a high geometrical accuracy of the C-arm system. In order to achieve this, geometry calibrations are performed to increase the accuracy given by the kinematics of the system. Commonly X-ray images of a phantom with known geometry are taken for the calibration. The images, together with a 3D model of the phantom, serve as input for an optimizer, which estimates the pose of the C-arm relative to the phantom. Afterwards, the estimates are used to increase the geometrical accuracy of the system. Inaccuracies due to real world effects appear, e.g. manufacturing or assembly inaccuracies of the phantom, the position of the X-ray tube, or the pose of the detector. To evaluate these factors separately a simulation is helpful, which needs to be as realistic as possible. To achieve this we defined three requirements, which have to be fulfilled: realistic noise, realistic absolute errors and similar error distributions within the working volume. By means of these criteria we investigate if our simulation mirrors a real world C-arm pose measurement for C-arm geometry calibration sufficiently.
引用
收藏
页码:383 / 388
页数:6
相关论文
共 50 条
  • [21] C-arm X-ray image mosaic using home-made ruler
    Chen, Diansheng
    Fan, Xing
    Hu, Lei
    Gaojishu Tongxin/Chinese High Technology Letters, 2006, 16 (10): : 1030 - 1033
  • [22] The Dose Rate Measurements of C-arm Fluoroscopic X-ray Systems with and without Apron
    Tuncel, Nina
    1ST INTERNATIONAL CONFERENCE ON RADIATIONS AND APPLICATIONS (ICRA-2017), 2018, 1994
  • [23] C-arm calibration - Is it really necessary?
    Jain, A
    Kon, R
    Zhou, Y
    Fichtinger, G
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2005, PT 1, 2005, 3749 : 639 - 646
  • [24] Single-view geometric calibration for C-arm inverse geometry CT
    Slagowski J.M.
    Dunkerley D.A.P.
    Hatt C.R.
    Speidel M.A.
    Speidel, Michael A. (speidel@wisc.edu), 2017, SPIE (04)
  • [25] C-arm calibration - is it really necessary?
    Jain, Ameet Kumar
    An, Michael
    Chitphakdithai, Nicha
    Chintalapani, Gouthami
    Fichtinger, Gabor
    MEDICAL IMAGING 2007: VISUALIZATION AND IMAGE-GUIDED PROCEDURES, PTS 1 AND 2, 2007, 6509
  • [26] High-resolution volume imaging on conventional interventional C-arm X-ray systems
    Rasche, V
    Graeff, C
    Istel, T
    Koppe, R
    Schreiber, B
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 1250 - 1250
  • [27] Single-camera visual odometry to track a surgical X-ray C-arm base
    Esfandiari, Hooman
    Lichti, Derek
    Anglin, Carolyn
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2017, 231 (12) : 1140 - 1151
  • [28] Super-global distortion correction for a rotational C-arm x-ray image intensifier
    Liu, RR
    Rudin, S
    Bednarek, DR
    MEDICAL PHYSICS, 1999, 26 (09) : 1802 - 1810
  • [29] Super-global distortion correction for a rotational C-arm x-ray image intensifier
    Toshiba Stroke Research Center, Departments of Physics and Radiology, State University of New York at Buffalo, Buffalo, NY 14214, United States
    不详
    Med. Phys., 9 (1802-1810):
  • [30] Sequential intrinsic and extrinsic geometry calibration in fluoro CT imaging with a mobile C-arm
    Cheryauka, Arvi
    Breham, Sebastien
    Christensen, Wayne
    MEDICAL IMAGING 2006: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2006, 6141