A Realistic X-Ray Simulation for C-Arm Geometry Calibration

被引:0
|
作者
Thuerauf, Sabine [1 ]
Wolf, Markus [2 ]
Koerner, Mario [2 ]
Vogt, Florian [2 ]
Hornung, Oliver [2 ]
Nasseri, M. Ali [3 ]
Knoll, Alois [4 ]
机构
[1] Fortiss GmbH, D-80805 Munich, Germany
[2] Siemens Healthcare GmbH, D-91301 Forchheim, Germany
[3] Klinikum Rechts Der Isar, D-81675 Munich, Germany
[4] Tech Univ Munich, Fac Comp Sci, Chair Robot & Embedded Syst, D-85748 Garching, Germany
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Applications like cone beam computed tomographies (CBCTs) or 2D-3D overlays need a high geometrical accuracy of the C-arm system. In order to achieve this, geometry calibrations are performed to increase the accuracy given by the kinematics of the system. Commonly X-ray images of a phantom with known geometry are taken for the calibration. The images, together with a 3D model of the phantom, serve as input for an optimizer, which estimates the pose of the C-arm relative to the phantom. Afterwards, the estimates are used to increase the geometrical accuracy of the system. Inaccuracies due to real world effects appear, e.g. manufacturing or assembly inaccuracies of the phantom, the position of the X-ray tube, or the pose of the detector. To evaluate these factors separately a simulation is helpful, which needs to be as realistic as possible. To achieve this we defined three requirements, which have to be fulfilled: realistic noise, realistic absolute errors and similar error distributions within the working volume. By means of these criteria we investigate if our simulation mirrors a real world C-arm pose measurement for C-arm geometry calibration sufficiently.
引用
收藏
页码:383 / 388
页数:6
相关论文
共 50 条
  • [1] Optimal configuration for dynamic calibration of projection geometry of X-ray C-arm systems
    Mitschke, MM
    Navab, N
    [J]. IEEE WORKSHOP ON MATHEMATICAL METHODS IN BIOMEDICAL IMAGE ANALYSIS, PROCEEDINGS, 2000, : 204 - 209
  • [2] Calibration Method of Projectional Geometry for X-ray C-arm Fluoroscopy Using Sinogram Data
    Lee, Donggi
    Kim, Sungmin
    [J]. APPLIED SCIENCES-BASEL, 2022, 12 (15):
  • [3] Distortion Correction, Geometric Calibration, and Volume Reconstruction for an Isocentric C-Arm X-Ray System
    Mennessier, C.
    Spencer, B.
    Clackdoyle, R.
    Conneau, A-C.
    Xu, T.
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2943 - 2947
  • [4] Photogrammetric calibration of a C-arm X-ray system as a verification tool for orthopaedic navigation systems
    Broers, H
    Hemken, H
    Luhmann, T
    Ritschl, P
    [J]. ISPRS JOURNAL OF PHOTOGRAMMETRY AND REMOTE SENSING, 2002, 56 (5-6) : 338 - 346
  • [5] Error Analysis of the X-ray Projection Geometry of Camera-Augmented Mobile C-arm
    Chen, Xin
    Wang, Lejing
    Fallavollita, Pascal
    Navab, Nassir
    [J]. MEDICAL IMAGING 2012: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2012, 8316
  • [6] ECG-Gated CT Reconstruction for a C-Arm Inverse Geometry X-Ray System
    Slagowski, J. M.
    Dunkerley, D. A. P.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3701 - 3701
  • [7] An X-Ray C-Arm Guided Automatic Targeting System for Histotripsy
    Wagner, Martin G.
    Periyasamy, Sarvesh
    Kutlu, Ayca Z.
    Pieper, Alexander A.
    Swietlik, John F.
    Ziemlewicz, Tim J.
    Hall, Timothy L.
    Xu, Zhen
    Speidel, Michael A.
    Lee, Fred T., Jr.
    Laeseke, Paul F.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2023, 70 (02) : 592 - 602
  • [8] Prospects for the use of the "C-arm" mobile X-ray diagnostic apparatus
    Blinov N.N.
    Mazurov A.I.
    [J]. Biomedical Engineering, 2000, 34 (5) : 250 - 253
  • [9] Modeling and optimization of rotational C-arm stereoscopic X-ray angiography
    Talukdar, AS
    Wilson, DL
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (07) : 604 - 616
  • [10] X-ray beam tracking system for a fluoroscopic C-arm unit
    Dinu, PM
    Bednarek, DR
    Wobschall, D
    Peterson, R
    Ionita, C
    Rudin, S
    Zeng, M
    Chugh, K
    Hoffmann, KR
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1424 - 1424