摘要
In this work we present the experimental and Monte Carlo calibration of the Compton-suppressed spectrometer of the IAEA's Environment Laboratories, Monaco. For this purpose the GESPECOR code was extended to include the specific geometry and to implement the veto logic, integrated with the coincidence summing module of the code. The simulation results are in good accordance with experimental calibrations. The code is fast and user-friendly, able to evaluate the efficiency and the correction factors for nuclides with arbitrary complex decay schemes. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:109 / 113
页数:5
相关论文
共 50 条
  • [31] Improvements of a low-level gamma-ray spectrometry system at the underground laboratory "UDO"
    Neumaier, S.
    Wojcik, M.
    Dombrowski, H.
    Arnold, D.
    APPLIED RADIATION AND ISOTOPES, 2009, 67 (05) : 726 - 730
  • [32] Variations of gamma-ray background in the Belgrade shallow underground low-level laboratory
    Banjanac, Radomir
    Dragic, Aleksandar
    Udovicic, Vladimir
    Jokovic, Dejan
    Maetic, Dimitrije
    Veselinovic, Nikola
    Savic, Mihailo
    APPLIED RADIATION AND ISOTOPES, 2014, 87 : 70 - 72
  • [33] COMPTON-SUPPRESSED LOW LEVEL GAMMA-RAY SCINTILLATION SPECTROMETER FOR SAMPLES OF EXTENDED GEOMETRY
    SETHI, SK
    NUCLEAR INSTRUMENTS & METHODS, 1971, 91 (03): : 397 - &
  • [34] A comprehensive calibration method of Ge detectors for low-level gamma-spectrometry measurements
    Pérez-Moreno, JP
    San Miguel, EG
    Bolivar, JP
    Aguado, JL
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 491 (1-2): : 152 - 162
  • [35] Determination of 137Cs in Soil Samples by Low-Level Compton Suppression Gamma-Counting
    F.Y. Iskander
    S. Landsberger
    S.D. Warren
    Journal of Radioanalytical and Nuclear Chemistry, 2000, 244 : 159 - 163
  • [36] Determination of 137Cs in soil samples by low-level Compton suppression gamma-counting
    Iskander, FY
    Landsberger, S
    Warren, SD
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2000, 244 (01) : 159 - 163
  • [37] COMPARISON OF A GE(LI) DETECTOR WITH A GE(LI) - NAL(TL) COMPTON SUPPRESSION SPECTROMETER FOR LOW-LEVEL ACTIVITY MEASUREMENTS
    SITAR, B
    JADERNA ENERGIE, 1977, 23 (09): : 342 - 344
  • [38] Calibration of low-level beta-gamma coincidence detector systems for xenon isotope detection
    Khrustalev, K.
    Wieslander, J. S. E.
    Auer, M.
    Gheddou, A.
    APPLIED RADIATION AND ISOTOPES, 2016, 109 : 418 - 424
  • [39] Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
    Muraishi, Hiroshi
    Enomoto, Ryoji
    Katagiri, Hideaki
    Kagaya, Mika
    Watanabe, Takara
    Narita, Naofumi
    Kano, Daisuke
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2020, (155):
  • [40] Measurement of 137Cs in the soil in Korea by low-level background gamma-ray spectrometer
    Sang-Han Lee
    Sun Ah Lee
    Jong Man Lee
    Tae Soon Park
    K. B. Lee
    Journal of Radioanalytical and Nuclear Chemistry, 2013, 296 : 721 - 725