Accuracy of radiotherapy dose calculations based on cone-beam CT: comparison of deformable registration and image correction based methods

被引:37
|
作者
Marchant, T. E. [1 ,2 ]
Joshi, K. D. [2 ]
Moore, C. J. [1 ]
机构
[1] Univ Manchester, Christie NHS Fdn Trust, Manchester Acad Hlth Sci Ctr, Manchester M20 4BX, Lancs, England
[2] Christie NHS Fdn Trust, Christie Med Phys & Engn, Manchester M20 4BX, Lancs, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 06期
基金
英国医学研究理事会;
关键词
cone-beam CT; image guided radiotherapy; deformable image registration; GUIDED RADIATION-THERAPY; SHADING CORRECTION; ADAPTIVE RADIOTHERAPY; COMPUTED-TOMOGRAPHY; CBCT; SCATTER;
D O I
10.1088/1361-6560/aab0f0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Radiotherapy dose calculations based on cone-beam CT (CBCT) images can be inaccurate due to unreliable Hounsfield units (HU) in the CBCT. Deformable image registration of planning CT images to CBCT, and direct correction of CBCT image values are two methods proposed to allow heterogeneity corrected dose calculations based on CBCT. In this paper we compare the accuracy and robustness of these two approaches. CBCT images for 44 patients were used including pelvis, lung and head & neck sites. CBCT HU were corrected using a 'shading correction' algorithm and via deformable registration of planning CT to CBCT using either Elastix or Niftyreg. Radiotherapy dose distributions were re-calculated with heterogeneity correction based on the corrected CBCT and several relevant dose metrics for target and OAR volumes were calculated. Accuracy of CBCT based dose metrics was determined using an 'override ratio' method where the ratio of the dose metric to that calculated on a bulk-density assigned version of the same image is assumed to be constant for each patient, allowing comparison to the patient's planning CT as a gold standard. Similar performance is achieved by shading corrected CBCT and both deformable registration algorithms, with mean and standard deviation of dose metric error less than 1% for all sites studied. For lung images, use of deformed CT leads to slightly larger standard deviation of dose metric error than shading corrected CBCT with more dose metric errors greater than 2% observed (7% versus 1%).
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Conventional CT to cone-beam CT image registration using combination of intensity- and landmark-based deformable image registration methods
    Gao, S.
    Wang, H.
    Zhang, L.
    Zhang, Y.
    Mohan, R.
    Dong, L.
    MEDICAL PHYSICS, 2007, 34 (06) : 2353 - 2353
  • [2] Deformable image registration in cone-beam CT images for image-guided adaptive radiotherapy
    Zhang, L.
    Dong, L.
    Zhu, X.
    Garden, A.
    Ahamad, A.
    Ang, K.
    Mohan, R.
    MEDICAL PHYSICS, 2006, 33 (06) : 1995 - 1995
  • [3] CT to Cone-Beam CT Deformable Registration With Simultaneous Intensity Correction
    Zhen, X.
    Yan, H.
    Gu, X.
    Zhou, L.
    Jia, X.
    Jiang, S.
    MEDICAL PHYSICS, 2012, 39 (06) : 3960 - 3960
  • [4] CT to cone-beam CT deformable registration with simultaneous intensity correction
    Zhen, Xin
    Gu, Xuejun
    Yan, Hao
    Zhou, Linghong
    Jia, Xun
    Jiang, Steve B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (21): : 6807 - 6826
  • [5] Effect of dose reduction on image registration and image quality for cone-beam CT in radiotherapy
    Loutfi-Krauss, B.
    Koehn, J.
    Bluemer, N.
    Freundl, K.
    Koch, T.
    Kara, E.
    Scherf, C.
    Roedel, C.
    Ramm, U.
    Licher, J.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (02) : 192 - 200
  • [6] Deformable registration of CT and cone-beam CT by local CBCT intensity correction
    Park, Seyoun
    Plishker, William
    Shekhar, Raj
    Quon, Harry
    Wong, John
    Lee, Junghoon
    MEDICAL IMAGING 2015: IMAGE PROCESSING, 2015, 9413
  • [7] The Use of Deformable Image Registration with Cone-Beam CT for a Better Evaluation of Cumulative Dose to Organs
    Fillion, O.
    Gingras, L.
    Archambault, L.
    MEDICAL PHYSICS, 2015, 42 (06) : 3288 - 3289
  • [8] Accumulation of the delivered dose based on cone-beam CT and deformable image registration for non-small cell lung cancer treated with hypofractionated radiotherapy
    Wang, Bin
    Wang, Da Quan
    Lin, Mao Sheng
    Lu, Shi Pei
    Zhang, Jun
    Chen, Li
    Li, Qi Wen
    Cheng, Zhang Kai
    Liu, Fang Jie
    Guo, Jin Yu
    Liu, Hui
    Qiu, Bo
    BMC CANCER, 2020, 20 (01)
  • [9] Streaming architectures for cone-beam CT image reconstruction and deformable registration
    Singh, H.
    Kandasamy, N.
    Folkert, M.
    Sharp, G.
    MEDICAL PHYSICS, 2007, 34 (06) : 2634 - 2634
  • [10] Accumulation of the delivered dose based on cone-beam CT and deformable image registration for non-small cell lung cancer treated with hypofractionated radiotherapy
    Bin Wang
    Da Quan Wang
    Mao Sheng Lin
    Shi Pei Lu
    Jun Zhang
    Li Chen
    Qi Wen Li
    Zhang Kai Cheng
    Fang Jie Liu
    Jin Yu Guo
    Hui Liu
    Bo Qiu
    BMC Cancer, 20