Pulse oximeters' reliability in detecting hypoxemia and bradycardia: Comparison between a conventional and two new generation oximeters

被引:70
|
作者
Bohnhorst, B [1 ]
Peter, CS [1 ]
Poets, CF [1 ]
机构
[1] Hannover Med Sch, Dept Neonatol & Pediat Pulmonol, D-3000 Hannover, Germany
关键词
arterial oxygen saturation; neonatal intensive care; preterm infants; pulse oximetry; sudden infant death syndrome;
D O I
10.1097/00003246-200005000-00050
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Pulse oximeters are increasingly used for patient monitoring; however, they are traditionally very prone to motion artifact. Newly developed instruments have lower false alarm rates. We wanted to know whether this is achieved at the expense of an increased proportion of false negative alarms such as missed or delayed identification of hypoxemia and/or bradycardia. Design: Observational study. Setting: Neonatal intensive care unit. Patients: A total of 17 unsedated preterm infants (median gestational age at birth, 25 wks; range, 24-30 wks). Intervention: Long-term recordings of transcutaneous partial pressure of oxygen (P-ToO2), heart rate, pulse oximeter saturation (Sp(O2)), and pulse rate from a conventional oximeter and two new generation oximeters. Measurements: Recordings were analyzed for episodes with P-TcO2 < 40 torr or with heart rate <80 beats/min for >5 sees. Hypoxemia was considered identified if Sp(O2) had fallen to <85% within 2 mins of P-TcO2 reaching 40 torr, and bradycardia was considered identified if pulse rate had fallen to <80 beats/min within 2 mins of the heart rate reaching this threshold. Main Results: A total of 202 falls in P-TcO2 to <40 torr occurred; 174 (86%) were identified by all three oximeters. Of the remaining episodes, manual analysis of red and infrared absorption signals confirmed that Sp(O2) had indeed been <85% for greater than or equal to 10 sees in 11 episodes; therefore, these episodes should have been identified by all three oximeters. None of these had been missed by the conventional oximeter, but 10 (5.4% of the total) were missed by one of the new generation instruments (Nellcor), and one (0.5%) was missed by the other (Masimo). Of 54 bradycardias, only 14 were identified by all three oximeters; 17 (32%) were missed by the conventional, 37 (69%) by the Nellcor, and 4 (7%) by the Masimo instrument. Conclusion: One of the two new generation instruments investigated in this study missed 5.4% of hypoxemic episodes and 69% of bradycardias. It thus appears that this instrument's reduced false alarm rate is achieved at the expense of an unreliable and/or delayed identification of hypoxemia and bradycardia. The other instrument identified both conditions equally as or more reliably than a conventional pulse oximeter.
引用
收藏
页码:1565 / 1568
页数:4
相关论文
共 47 条
  • [1] Assessment of 2 newer generation pulse oximeters in children with hypoxemia
    Torres, A
    Wohrley, J
    Raff, G
    CRITICAL CARE MEDICINE, 2002, 30 (12) : A71 - A71
  • [2] Comparison of two new generation pulse oximeters during emergency ambulance transportation
    Weber, Ulrike
    Tomschik, Elvira
    Resch, Irene
    Adelmann, Krista
    Hasun, Matthias
    Mora, Bruno
    Malzer, Reinhard
    Kober, Alexander
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2011, 28 (02) : 92 - 96
  • [3] A comparison of new generation pulse oximeters in critically ill patients
    Emberger, JS
    Fulda, GJ
    CRITICAL CARE MEDICINE, 2000, 28 (12) : A61 - A61
  • [4] Comparison of new generation motion-resistant pulse oximeters
    Townshend, Jennifer
    Taylor, Barry J.
    Galland, Barbara
    Williams, Sheila
    JOURNAL OF PAEDIATRICS AND CHILD HEALTH, 2006, 42 (06) : 359 - 365
  • [5] How reliable are the new generation pulse oximeters in the detection of hypoxemia in the intensive care unit?
    Appavu, S
    Haley, T
    Giuliano, KK
    Duller, B
    CRITICAL CARE MEDICINE, 2003, 31 (12) : A96 - A96
  • [6] Accuracy of Pulse Oximeters in Detecting Hypoxemia in Patients with Chronic Thromboembolic Pulmonary Hypertension
    Kohyama, Tomoki
    Moriyama, Kiyoshi
    Kanai, Riichiro
    Kotani, Mariko
    Uzawa, Kohji
    Satoh, Toru
    Yorozu, Tomoko
    PLOS ONE, 2015, 10 (05):
  • [7] Comparison of Two New Generation Pulse Oximeters with Arterial Oxygen Saturation in Critically Ill Children
    Jose, Bipin
    Lodha, Rakesh
    Kabra, S. K.
    INDIAN JOURNAL OF PEDIATRICS, 2014, 81 (12): : 1297 - 1301
  • [8] Comparison of Two New Generation Pulse Oximeters with Arterial Oxygen Saturation in Critically Ill Children
    Bipin Jose
    Rakesh Lodha
    S. K. Kabra
    The Indian Journal of Pediatrics, 2014, 81 : 1297 - 1301
  • [9] A Second Generation Technique for Evaluating Accuracy and Reliability of Second Generation Pulse Oximeters
    K. K. Tremper
    Journal of Clinical Monitoring and Computing, 2000, 16 : 473 - 474
  • [10] Comparison of Two New Generation Pulse Oximeters with Arterial Oxygen Saturation in Critically Ill Children: Correspondence
    Viswas Chhapola
    Sandeep Kumar Kanwal
    The Indian Journal of Pediatrics, 2015, 82 : 665 - 666