Extraction induced by emulsion breaking for simultaneous determination of Co, Fe and Ni in petroleum asphalt cement by high-resolution continuum source atomic absorption spectrometry

被引:17
|
作者
Adolfo, Franciele Rovasi [1 ]
do Nascimento, Paulo Cicero [1 ]
Bohrer, Denise [1 ]
Viana, Carine [1 ]
de Carvalho, Leandro Machado [1 ]
Coutinho Cravo, Margareth C. [2 ]
Nascimento, Luis [2 ]
机构
[1] Univ Fed Santa Maria, Dept Quim, Santa Maria, RS, Brazil
[2] Petrobras SA, Ctr Pesquisa & Desenvolvimento Leopoldo Amer Migu, Rio De Janeiro, RJ, Brazil
关键词
Simultaneous determination; High-resolution continuum source GF AAS; Extraction induced by emulsion breaking; Asphalt cement; Metals; OPTICAL-EMISSION SPECTROMETRY; PLASMA-MASS SPECTROMETRY; CRUDE-OIL; MICROEMULSION BREAKING; METALS DETERMINATION; LUBRICATING OILS; VANADIUM; NICKEL; CU; MN;
D O I
10.1016/j.fuel.2020.118098
中图分类号
TE [石油、天然气工业]; TK [能源与动力工程];
学科分类号
0807 ; 0820 ;
摘要
This paper reports, for the first time, the simultaneous determination of Co, Fe and Ni in petroleum asphalt cement by high-resolution continuum source atomic absorption spectrometry after extraction induced by the emulsion breaking method. The variables that could influence the sample preparation were evaluated, like sample dilution with toluene and mineral oil, extraction time, HNO3 and Triton X-100 concentration in the aqueous extractant solution. M optimized conditions, 0.08 g of the sample was dissolved with 1.5 mL of toluene and 3.5 mL of mineral oil and vigorous mixed with the extractant solution containing 40% (v/v) of HNO3 and 40% (v/v) of Triton-X-100. The emulsions were broken by heating just after their formation and the aqueous extracts were collected for analysis. Simultaneous determination was carried out at the secondary lines of Co (352.685 nm), Fe (352.604 nm) and Ni (352.454 nm). For this purpose, the best pyrolysis and atomization temperatures were 700 and 2400 degrees C, respectively. Calibration was performed with standard aqueous solutions and no matrix effects were observed. The limit of detection was 0.003 mg L-1 for Co, 0.018 mg L-1 for Fe and 0.027 mg L-1 for Ni. Method accuracy was confirmed by comparing the results obtained by the proposed extraction method with those obtained after samples acid digestion. Recoveries ranged from 100 to 107%, 103 to 109% and 95.5 to 98.9% were obtained for Co, Fe and Ni, respectively.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Three Advantages of High-Resolution Continuum Source Atomic Absorption Spectrometry
    Sun, Siqi
    SPECTROSCOPY, 2021, 36 (08) : 22 - 22
  • [22] High-Resolution Continuum Source Atomic and Molecular Absorption Spectrometry - A Review
    Welz, Bernhard
    Mores, Silvane
    Carasek, Eduardo
    Vale, Maria Goreti R.
    Okruss, Michael
    Becker-Ross, Helmut
    APPLIED SPECTROSCOPY REVIEWS, 2010, 45 (05) : 327 - 354
  • [23] High-Resolution Continuum Source Atomic Absorption Spectrometry with Microwave-Assisted Extraction for the Determination of Metals in Vegetable Sprouts
    Blicharska, E.
    Flieger, J.
    Oszust, K.
    Frac, M.
    Swieboda, R.
    Kocjan, R.
    ANALYTICAL LETTERS, 2015, 48 (14) : 2272 - 2287
  • [24] Determination of Gold by High-Resolution Continuum Source Atomic Absorption Spectrometry with Chemical Vapor Generation
    Krawczyk, Magdalena
    Matusiewicz, Henryk
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2013, 24 (05) : 749 - 757
  • [25] Direct Determination of Cd, Co, Cu, Fe, Mn, Na, Ni, Pb, and Zn in Ethanol Fuel by High-Resolution Continuum Source Flame Atomic Absorption Spectrometry
    Leite, Clarice C.
    de Jesus, Alexandre
    Potes, Mariana L.
    Vieira, Mariana A.
    Samios, Dimitrios
    Silva, Marcia M.
    ENERGY & FUELS, 2015, 29 (11) : 7358 - 7363
  • [26] Evaluation of solid sampling for determination of Mo, Ni, Co, and V in soil by high-resolution continuum source graphite furnace atomic absorption spectrometry
    Babos, Diego Victor
    Barros, Ariane Isis
    Ferreira, Edilene Cristina
    Gomes Neto, Jose Anchieta
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2017, 130 : 39 - 44
  • [27] Simultaneous Determination of Chromium and Iron in Powdered Milk Using High-Resolution Continuum Source Graphite Furnace Atomic Absorption Spectrometry
    Santos, Adilson S.
    de Souza, Cheilane T.
    Leao, Danilo J.
    Correia, Felipo O.
    Almeida, Tarcisio S.
    Ferreira, Sergio L. C.
    FOOD ANALYTICAL METHODS, 2020, 13 (01) : 284 - 290
  • [28] Simultaneous determination of iron and nickel in fluoropolymers by solid sampling high-resolution continuum source graphite furnace atomic absorption spectrometry
    Soares, Bruno M.
    Santos, Rafael F.
    Bolzan, Rodrigo C.
    Muller, Edson I.
    Primel, Ednei G.
    Duarte, Fabio A.
    TALANTA, 2016, 160 : 454 - 460
  • [29] Simultaneous determination of cadmium and iron in different kinds of cereal flakes using high-resolution continuum source atomic absorption spectrometry
    Zverina, Ondrej
    Kuta, Jan
    Coufalik, Pavel
    Koseckova, Pavlina
    Komarek, Josef
    FOOD CHEMISTRY, 2019, 298
  • [30] Simultaneous Determination of Chromium and Iron in Powdered Milk Using High-Resolution Continuum Source Graphite Furnace Atomic Absorption Spectrometry
    Adilson S. Santos
    Cheilane T. de Souza
    Danilo J. Leao
    Felipo O. Correia
    Tarcísio S. Almeida
    Sergio L. C. Ferreira
    Food Analytical Methods, 2020, 13 : 284 - 290