The Effects of Cortical Bone Thickness and Trabecular Bone Strength on Noninvasive Measures of the Implant Primary Stability Using Synthetic Bone Models

被引:56
|
作者
Hsu, Jui-Ting [1 ]
Fuh, Lih-Jyh [1 ]
Tu, Ming-Gene [1 ]
Li, Yu-Fen [2 ,3 ]
Chen, Kuan-Ting [2 ]
Huang, Heng-Li [1 ]
机构
[1] China Med Univ, Sch Dent, Taichung 40402, Taiwan
[2] China Med Univ, Ctr Biostat, Taichung 40402, Taiwan
[3] China Med Univ, Grad Inst Biostat, Taichung 40402, Taiwan
关键词
cortical bone thickness; implant primary stability; insertion torque; periotest; resonance frequency analysis; strength of trabecular bone; RESONANCE FREQUENCY-ANALYSIS; DENTAL IMPLANTS; CANCELLOUS BONE; FINITE-ELEMENT; HUMAN CADAVER; OSSEOINTEGRATION; DENSITY; AUGMENTATION; OSTEOPOROSIS; SIMULATION;
D O I
10.1111/j.1708-8208.2011.00349.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study investigated how the primary stability of a dental implant as measured by the insertion torque value (ITV), Periotest value (PTV), and implant stability quotient (ISQ) is affected by varying thicknesses of cortical bone and strengths of trabecular bone using synthetic bone models. Materials and Methods: Four synthetic cortical shells (with thicknesses of 0, 1, 2, and 3mm) were attached to four cellular rigid polyurethane foams (with elastic moduli of 137, 47.5, 23, and 12.4MPa) and one open-cell rigid polyurethane foam which mimic the osteoporotic bone (with an elastic modulus 6.5MPa), to represent the jawbones with various cortical bone thicknesses and strengths of trabecular bone. A total of 60 bone specimens accompanied with implants was examined by a torque meter, Osstell resonance frequency analyzer, and Periotest electronic device. All data were statistically analyzed by two-way analysis of variance. In addition, second-order nonlinear regression was utilized to assess the correlations of the primary implant stability with the four cortex thicknesses and five strengths of trabecular bone. Results: ITV, ISQ, and PTV differed significantly (p<.05) and were strongly correlated with the thickness of cortical bone (R2>0.9) and the elastic modulus of trabecular bone (R2=0.740.99). Conclusions: The initial stability at the time of implant placement is influenced by both the cortical bone thickness and the strength of trabecular bone; however, these factors are mostly nonlinearly correlated with ITV, PTV, and ISQ. Using ITV and PTV seems more suitable for identifying the primary implant stability in osteoporotic bone with a thin cortex.
引用
收藏
页码:251 / 261
页数:11
相关论文
共 50 条
  • [31] Effects of cortical bone thickness at different healing times on microscrew stability
    Wei, Xing
    Zhao, Lixing
    Xu, Zhenrui
    Tang, Tian
    Zhao, Zhihe
    ANGLE ORTHODONTIST, 2011, 81 (05) : 760 - 766
  • [32] The discrete nature of trabecular bone microarchitecture affects implant stability
    Wirth, Andreas J.
    Mueller, Ralph
    van Lenthe, G. Harry
    JOURNAL OF BIOMECHANICS, 2012, 45 (06) : 1060 - 1067
  • [33] The different contributions of cortical and trabecular bone to implant anchorage in a human vertebra
    Ruffoni, Davide
    Wirth, Andreas J.
    Steiner, Juri A.
    Parkinson, Ian H.
    Mueller, Ralph
    van Lenthe, G. Harry
    BONE, 2012, 50 (03) : 733 - 738
  • [34] Parathyroid hormone deficiency and excess: Similar effects on trabecular bone but differing effects on cortical bone
    Duan, YB
    De Luca, V
    Seeman, E
    JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1999, 84 (02): : 718 - 722
  • [35] Differing effects of denosumab and alendronate on cortical and trabecular bone
    Zebaze, Roger M.
    Libanati, Cesar
    Austin, Matthew
    Ghasem-Zadeh, Ali
    Hanley, David A.
    Zanchetta, Jose R.
    Thomas, Thierry
    Boutroy, Stephanie
    Bogado, Cesar E.
    Bilezikian, John P.
    Seeman, Ego
    BONE, 2014, 59 : 173 - 179
  • [36] Evaluation of Cortical Bone Microdamage and Primary Stability of Orthodontic Miniscrew Using a Human Bone Analogue
    Teekavanich, Chutimont
    Uezono, Masayoshi
    Takakuda, Kazuo
    Ogasawara, Takeshi
    Techalertpaisarn, Paiboon
    Moriyama, Keiji
    MATERIALS, 2021, 14 (08)
  • [37] Differential menopausal effects on trabecular and cortical bone density
    Lukacs, JL
    Kleerekoper, M
    Ansbacher, R
    Rolfes-Curl, A
    Padmanabhan, VE
    Reame, NE
    MENOPAUSE-THE JOURNAL OF THE NORTH AMERICAN MENOPAUSE SOCIETY, 2003, 10 (06): : 597 - 597
  • [38] The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis
    Sugiura, Tsutomu
    Yamamoto, Kazuhiko
    Horita, Satoshi
    Murakami, Kazuhiro
    Tsutsumi, Sadami
    Kirita, Tadaaki
    JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, 2016, 46 (03): : 152 - 165
  • [39] Bone density and primary stability in implant therapy
    Molly, Liene
    CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 : 124 - 135
  • [40] Cortical bone is more sensitive to alcohol dose effects than trabecular bone in the rat
    Maurel, Delphine B.
    Boisseau, Nathalie
    Benhamou, Claude-Laurent
    Jaffre, Christelle
    JOINT BONE SPINE, 2012, 79 (05) : 492 - 499