The Effects of Cortical Bone Thickness and Trabecular Bone Strength on Noninvasive Measures of the Implant Primary Stability Using Synthetic Bone Models

被引:56
|
作者
Hsu, Jui-Ting [1 ]
Fuh, Lih-Jyh [1 ]
Tu, Ming-Gene [1 ]
Li, Yu-Fen [2 ,3 ]
Chen, Kuan-Ting [2 ]
Huang, Heng-Li [1 ]
机构
[1] China Med Univ, Sch Dent, Taichung 40402, Taiwan
[2] China Med Univ, Ctr Biostat, Taichung 40402, Taiwan
[3] China Med Univ, Grad Inst Biostat, Taichung 40402, Taiwan
关键词
cortical bone thickness; implant primary stability; insertion torque; periotest; resonance frequency analysis; strength of trabecular bone; RESONANCE FREQUENCY-ANALYSIS; DENTAL IMPLANTS; CANCELLOUS BONE; FINITE-ELEMENT; HUMAN CADAVER; OSSEOINTEGRATION; DENSITY; AUGMENTATION; OSTEOPOROSIS; SIMULATION;
D O I
10.1111/j.1708-8208.2011.00349.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study investigated how the primary stability of a dental implant as measured by the insertion torque value (ITV), Periotest value (PTV), and implant stability quotient (ISQ) is affected by varying thicknesses of cortical bone and strengths of trabecular bone using synthetic bone models. Materials and Methods: Four synthetic cortical shells (with thicknesses of 0, 1, 2, and 3mm) were attached to four cellular rigid polyurethane foams (with elastic moduli of 137, 47.5, 23, and 12.4MPa) and one open-cell rigid polyurethane foam which mimic the osteoporotic bone (with an elastic modulus 6.5MPa), to represent the jawbones with various cortical bone thicknesses and strengths of trabecular bone. A total of 60 bone specimens accompanied with implants was examined by a torque meter, Osstell resonance frequency analyzer, and Periotest electronic device. All data were statistically analyzed by two-way analysis of variance. In addition, second-order nonlinear regression was utilized to assess the correlations of the primary implant stability with the four cortex thicknesses and five strengths of trabecular bone. Results: ITV, ISQ, and PTV differed significantly (p<.05) and were strongly correlated with the thickness of cortical bone (R2>0.9) and the elastic modulus of trabecular bone (R2=0.740.99). Conclusions: The initial stability at the time of implant placement is influenced by both the cortical bone thickness and the strength of trabecular bone; however, these factors are mostly nonlinearly correlated with ITV, PTV, and ISQ. Using ITV and PTV seems more suitable for identifying the primary implant stability in osteoporotic bone with a thin cortex.
引用
收藏
页码:251 / 261
页数:11
相关论文
共 50 条
  • [21] Implant design and its effects on osseointegration over time within cortical and trabecular bone
    Beutel, Bryan G.
    Danna, Natalie R.
    Granato, Rodrigo
    Bonfante, Estevam A.
    Marin, Charles
    Tovar, Nick
    Suzuki, Marcelo
    Coelho, Paulo G.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2016, 104 (06) : 1091 - 1097
  • [22] Effects of Resorption Cavities on Strength of Trabecular Bone
    Easley, S. K.
    Shindich, D.
    Hernandez, C. J.
    Keaveny, T. M.
    JOURNAL OF BONE AND MINERAL RESEARCH, 2008, 23 : S230 - S230
  • [23] Effects of Elasticity and Structure of Trabecular Bone on the Primary Stability of Dental Implants
    Huang, Heng-Li
    Tu, Ming-Gene
    Fuh, Lih-Jyh
    Chen, Yuan-Chien
    Wu, Chu-Lung
    Chen, Shou-I
    Hsu, Jui-Ting
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2010, 30 (02) : 85 - 89
  • [24] Optimization of primary screw stability in Trabecular bone using neural network-based models
    Thou, Yijun
    Helgason, Benedikt
    Ferguson, Stephen J.
    Persson, Cecilia
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2025, 264
  • [25] CORTICAL AND TRABECULAR BONE CONTRIBUTE STRENGTH TO THE OSTEOPENIC DISTAL RADIUS
    SPADARO, JA
    WERNER, FW
    BRENNER, RA
    FORTINO, MD
    FAY, LA
    EDWARDS, WT
    JOURNAL OF ORTHOPAEDIC RESEARCH, 1994, 12 (02) : 211 - 218
  • [26] Prediction of low bone mass using a combinational approach of cortical and trabecular bone measures from dental panoramic radiographs
    Kathirvelu, D.
    Anburajan, M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (09) : 890 - 898
  • [27] In vitro evaluation of the influence of the cortical bone on the primary stability of two implant systems
    Andres-Garcia, Rodrigo
    Garcia Vives, Nuria
    Herrero Climent, Federico
    Fernandez Palacin, Ana
    Rios Santos, Vicente
    Herrero Climent, Mariano
    Bullon, Pedro
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (02): : E93 - E97
  • [28] Quantitative biomechanical analysis of the influence of the cortical bone and implant length on primary stability
    Hong, Jongrak
    Lim, Young-Jun
    Park, Sang-Oh
    CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (10) : 1193 - 1197
  • [29] In vivo monitoring of bone architecture and remodeling after implant insertion: The different responses of cortical and trabecular bone
    Li, Zihui
    Kuhn, Gisela
    von Salis-Soglio, Marcella
    Cooke, Stephen J.
    Schirmer, Michael
    Mueller, Ralph
    Ruffoni, Davide
    BONE, 2015, 81 : 468 - 477
  • [30] Impact of Bone Quality and Implant Type on the Primary Stability: An Experimental Study Using Bovine Bone
    Anil, Sukumaran
    Aldosari, Abdullah Alfarraj
    JOURNAL OF ORAL IMPLANTOLOGY, 2015, 41 (02) : 144 - 148