Artifact analysis of approximate helical cone-beam CT reconstruction algorithms

被引:34
|
作者
Köhler, T
Proksa, R
Bontus, C
Grass, M
Timmer, J
机构
[1] Philips Res Labs, Sector Tech Syst, D-22335 Hamburg, Germany
[2] Philips Med Syst, NL-5680 DA Best, Netherlands
关键词
cone-beam CT; cone-beam reconstruction; helical CT; PI; PI-SLANT; n-PI; ASSR;
D O I
10.1118/1.1413518
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper, four approximate cone-beam CT reconstruction algorithms are compared: Advanced single slice rebinning (ASSR) as a representative of algorithms employing a two dimensional approximation, PI, PI-SLANT, and 3-PI which all use a proper three dimensional back-projection. A detailed analysis of the image artifacts produced by these techniques shows that aliasing in the z-direction is the predominant source of artifacts for a 16-row scanner with 1.25 mm nominal slice thickness. For a detector with isotropic resolution of 0.5 mm, we found that ASSR and PI produce different kinds of artifacts which are almost at the same level, while PI-SLANT produces none of these artifacts. It is shown that the used of redundant data in the 3-PI method suppresses aliasing artifacts efficiently for both scanners. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:51 / 64
页数:14
相关论文
共 50 条
  • [1] Exact and approximate algorithms for helical cone-beam CT
    Kudo, H
    Rodet, T
    Noo, F
    Defrise, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13): : 2913 - 2931
  • [2] Evaluation of helical cone-beam CT reconstruction algorithms
    Köhler, T
    Bontus, C
    Brown, K
    Heuscher, D
    Grass, M
    Shechter, G
    Proksa, R
    [J]. 2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1217 - 1220
  • [3] Sparse sampling in helical cone-beam CT perfect reconstruction algorithms
    Bendory, Tamir
    Feuer, Arie
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (03) : 389 - 405
  • [4] Artifact analysis and reconstruction improvement in helical cardiac cone beam CT
    Manzke, R
    Grass, M
    Hawkes, D
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2004, 23 (09) : 1150 - 1164
  • [5] Comparison among approximate cone-beam reconstruction algorithms
    Zeng, Kai
    Chen, Zhi-Qiang
    Zhang, Li
    Zhao, Zi-Ran
    [J]. Hedianzixue Yu Tance Jishu/Nuclear Electronics and Detection Technology, 2004, 24 (05):
  • [6] Comparing Short Scan CT Reconstruction Algorithms Regarding Cone-Beam Artifact Performance
    Maass, Clemens
    Dennerlein, Frank
    Noo, Frederic
    Kachelriess, Marc
    [J]. 2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2188 - 2193
  • [7] An Improved Helical Cone-beam CT Image Reconstruction
    Gao, Yang
    Ma, Jianhua
    Chen, Wufan
    [J]. 2009 3RD INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING, VOLS 1-11, 2009, : 2178 - 2182
  • [8] Generalized quasi-exact algorithms for image reconstruction in helical cone-beam CT
    Zou, Y
    Sidky, EY
    Pan, XC
    [J]. MEDICAL IMAGING 2003: IMAGE PROCESSING, PTS 1-3, 2003, 5032 : 696 - 702
  • [9] Oblique surface reconstruction to approximate cone-beam helical data in multi-slice CT
    Chen, LG
    Heuscher, DJ
    Liang, Y
    [J]. IMAGE RECONSTRUCTION FROM INCOMPLETE DATA, 2000, 4123 : 279 - 284
  • [10] Rebinning-based algorithms for helical cone-beam CT
    Defrise, M
    Noo, F
    Kudo, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11): : 2911 - 2937