EchoSeed Model 6733 Iodine-125 brachytherapy source: Improved dosimetric characterization using the MCNP5 Monte Carlo code

被引:10
|
作者
Mosleh-Shirazi, M. A. [1 ,2 ]
Hadad, K. [3 ,4 ]
Faghihi, R. [3 ,4 ]
Baradaran-Ghahfarokhi, M. [3 ,4 ]
Naghshnezhad, Z. [3 ,4 ]
Meigooni, A. S. [5 ]
机构
[1] Shiraz Univ Med Sci, Dept Radiotherapy, Ctr Res Med Phys & Biomed Engn, Shiraz 7193613311, Iran
[2] Shiraz Univ Med Sci, Dept Radiotherapy, Phys Unit, Shiraz 7193613311, Iran
[3] Shiraz Univ, Sch Engn, Ctr Radiat Res, Shiraz 7193613311, Iran
[4] Shiraz Univ, Sch Engn, Dept Radiat Med, Shiraz 7193613311, Iran
[5] Comprehens Canc Ctr Nevada, Las Vegas, NV 89169 USA
关键词
EchoSeed Model 6733 Iodine-125 source; Monte Carlo simulation; MCNP; AAPM TG-43U1; brachytherapy; PERMANENT PROSTATE IMPLANTS; PARAMETERS; SEED; UPDATE; CANCER;
D O I
10.1118/1.4736418
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study primarily aimed to obtain the dosimetric characteristics of the Model 6733 I-125 seed (EchoSeed) with improved precision and accuracy using a more up-to-date Monte-Carlo code and data (MCNP5) compared to previously published results, including an uncertainty analysis. Its secondary aim was to compare the results obtained using the MCNP5, MCNP4c2, and PTRAN codes for simulation of this low-energy photon-emitting source. The EchoSeed geometry and chemical compositions together with a published I-125 spectrum were used to perform dosimetric characterization of this source as per the updated AAPM TG-43 protocol. These simulations were performed in liquid water material in order to obtain the clinically applicable dosimetric parameters for this source model. Dose rate constants in liquid water, derived from MCNP4c2 and MCNP5 simulations, were found to be 0.993 cGyh(-1) U-1 (+/- 1.73%) and 0.965 cGyh(-1) U-1 (+/- 1.68%), respectively. Overall, the MCNP5 derived radial dose and 2D anisotropy functions results were generally closer to the measured data (within +/- 4%) than MCNP4c and the published data for PTRAN code (Version 7.43), while the opposite was seen for dose rate constant. The generally improved MCNP5 Monte Carlo simulation may be attributed to a more recent and accurate cross-section library. However, some of the data points in the results obtained from the above-mentioned Monte Carlo codes showed no statistically significant differences. Derived dosimetric characteristics in liquid water are provided for clinical applications of this source model. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4736418]
引用
收藏
页码:4653 / 4659
页数:7
相关论文
共 37 条
  • [1] Monte Carlo study of an electronic brachytherapy source using MCNP5 and EGSnrc
    Davis, S.
    DeWerd, L.
    MEDICAL PHYSICS, 2007, 34 (06) : 2472 - 2473
  • [2] Experimental determination of the TG-43 dosimetric characteristics of EchoSeed™ model 6733 125I brachytherapy source
    Meigooni, AS
    Dini, SA
    Sowards, K
    Hayes, JL
    Al-Otoom, A
    MEDICAL PHYSICS, 2002, 29 (06) : 939 - 942
  • [3] Dosimetric characterization of a brachytherapy applicator using MCNP5 modelisation and in-phantom measurements
    Gerardy, I.
    Rodenas, J.
    van Dycke, M.
    Gallardo, S.
    Ceccolini, Elisa
    APPLIED RADIATION AND ISOTOPES, 2010, 68 (4-5) : 735 - 737
  • [4] Uncertainty analysis in environmental radioactivity measurements using the Monte Carlo code MCNP5
    Gallardo, S.
    Querol, A.
    Ortiz, J.
    Rodenas, J.
    Verdu, G.
    Villanueva, J. F.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 116 : 214 - 218
  • [5] Dosimetric capabilities of the Iview GT portal imager using MCNP5 Monte Carlo simulations
    Juste, B.
    Miro, R.
    Diez, S.
    Campayo, J. M.
    Verdu, G.
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 3743 - +
  • [6] Validation of a Clinical PET Scanner Using Monte Carlo Simulation Code: MCNP5
    Musarudin, Marianie
    Saripan, M. Iqbal
    Mashohor, Syamsiah
    Saad, Wira Hidayat Mohd
    Hashim, Suhairul
    Nordin, Abdul Jalil
    2013 8TH EUROSIM CONGRESS ON MODELLING AND SIMULATION (EUROSIM), 2013, : 36 - 41
  • [7] Dosimetric characterization of model Cs-1 Rev2 cesium-131 brachytherapy source in water phantoms and human tissues with MCNP5 Monte Carlo simulation
    Wang, Jianhua
    Zhang, Hualin
    MEDICAL PHYSICS, 2008, 35 (04) : 1571 - 1579
  • [8] Dosimetric characterization of an 192Ir brachytherapy source with the Monte Carlo code PENELOPE
    Javier Casado, Francisco
    Garcia-Pareja, Salvador
    Cenizo, Elena
    Mateo, Beatriz
    Bodineau, Coral
    Galan, Pedro
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2010, 26 (03): : 132 - 139
  • [9] Dosimetric characterization of a newly designed encapsulated interstitial brachytherapy source of iodine-125 -: model LS-1 Brachyseed™
    Nath, R
    Yue, N
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (06) : 813 - 821
  • [10] Evaluation of an extrapolation chamber for dosimetry in computed tomography beams using Monte Carlo code (MCNP5)
    Castro, Maysa C.
    Silva, Natalia F.
    Santos, Lucas R.
    Cintra, Felipe B.
    Caldas, Linda V. E.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 167