A method to provide rapid in situ determination of tip radius in dynamic atomic force microscopy

被引:77
|
作者
Santos, Sergio [1 ,2 ,3 ]
Guang, Li [3 ]
Souier, Tewfik [3 ]
Gadelrab, Karim [3 ]
Chiesa, Matteo [3 ]
Thomson, Neil H. [1 ,2 ]
机构
[1] Univ Leeds, Dept Oral Biol, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Sch Phys & Astron, Leeds LS2 9JT, W Yorkshire, England
[3] Masdar Inst Sci & Technol, Lab Energy & Nanosci, Abu Dhabi, U Arab Emirates
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2012年 / 83卷 / 04期
关键词
VIBRATING TIP; ENERGY-DISSIPATION; MICROLEVER SYSTEM; BEHAVIOR; CONTACT; CALIBRATION; CONTRAST;
D O I
10.1063/1.4704376
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We provide a method to characterize the tip radius of an atomic force microscopy in situ by monitoring the dynamics of the cantilever in ambient conditions. The key concept is that the value of free amplitude for which transitions from the attractive to repulsive force regimes are observed, strongly depends on the curvature of the tip. In practice, the smaller the value of free amplitude required to observe a transition, the sharper the tip. This general behavior is remarkably independent of the properties of the sample and cantilever characteristics and shows the strong dependence of the transitions on the tip radius. The main advantage of this method is rapid in situ characterization. Rapid in situ characterization enables one to continuously monitor the tip size during experiments. Further, we show how to reproducibly shape the tip from a given initial size to any chosen larger size. This approach combined with the in situ tip size monitoring enables quantitative comparison of materials measurements between samples. These methods are set to allow quantitative data acquisition and make direct data comparison readily available in the community. (C) 2012 American Institute of Physics. [http://dx.doi.org/10.1063/1.4704376]
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Rapid self-assembly of α-synuclein observed by in situ atomic force microscopy
    Hoyer, WG
    Cherny, D
    Subramaniam, V
    Jovin, TM
    JOURNAL OF MOLECULAR BIOLOGY, 2004, 340 (01) : 127 - 139
  • [22] Simulation of the influence of the tip radius on the sensitivity of an atomic force microscope
    Pokropivnyi, AV
    Pokropivnyi, VV
    Skorokhod, VV
    TECHNICAL PHYSICS, 1997, 42 (12) : 1435 - 1439
  • [23] Unstable amplitude and noisy image induced by tip contamination in dynamic force mode atomic force microscopy
    Nie, H. -Y.
    McIntyre, N. S.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (02):
  • [24] Simulation of the influence of the tip radius on the sensitivity of an atomic force microscope
    A. V. Pokropivnyi
    V. V. Pokropivnyi
    V. V. Skorokhod
    Technical Physics, 1997, 42 : 1435 - 1439
  • [25] Force feedback in dynamic atomic force microscopy
    Stark, R. W.
    PROCEEDINGS OF THE ASME DESIGN ENGINEERING DIVISION 2005, PTS A AND B, 2005, : 507 - 515
  • [26] Impact of the tip radius on the lateral resolution in piezoresponse force microscopy
    Jungk, T.
    Hoffmann, A.
    Soergel, E.
    NEW JOURNAL OF PHYSICS, 2008, 10
  • [27] DYNAMICS OF TIP SUBSTRATE INTERACTIONS IN ATOMIC FORCE MICROSCOPY
    LANDMAN, U
    LUEDTKE, WD
    NITZAN, A
    SURFACE SCIENCE, 1989, 210 (03) : L177 - L184
  • [28] EFFECT OF THE TIP STRUCTURE ON ATOMIC-FORCE MICROSCOPY
    SASAKI, N
    TSUKADA, M
    PHYSICAL REVIEW B, 1995, 52 (11): : 8471 - 8482
  • [29] ATOMIC FORCE MICROSCOPY USING ZNO WHISKER TIP
    KADO, H
    YOKOYAMA, K
    TOHDA, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (06): : 3330 - 3332
  • [30] Contrast artifacts in tapping tip atomic force microscopy
    Department of Physics, Bldg. 307, Technical University of Denmark, DK-2800 Lyngby, Denmark
    Appl Phys A, SUPPL. 1 (S329-S332):